You need JavaScript to view this

Determination of the detection probability of neutron counters; Bestimmung der Nachweiswahrscheinlichkeit von Neutronenzaehlern

Abstract

At the PHOENICS experiment neutron detection is done by scintillation counters consisting of 100x20x20 cm{sup 3} NE110 or Nuplex3 bars with photomultiplier tubes mounted on both ends. Calibration measurements to determine the neutron detection efficiency have been done for 4 counters at the LNS (Laboratoire National Saturne, Saclay, France). The method to obtain neutrons of well-defined energy and angle was neutron-tagging in elastic np-scattering off a CH{sub 2}-target. The efficiency was measured at different impact positions and for up to five neutron energies between 20 and 175 MeV. The efficiency shows a strong energy dependence for low energies and becomes rather constant (ca. 20%) at higher energies. The detection thresholds were varied between 1.1 and 15 MeV. Especially for low energies, the detection efficiency depends strongly on the threshold. The results of Monte Carlo simulation are in good agreement with the measurements. (orig.). [Deutsch] Die am PHOENICS-Experiment verwendeten Neutronenzaehler bestehen aus einem 20x20x100 cm{sup 3} Szintillatorblock aus Nuplex3 oder NE110, der beidseitig ueber Lichtleiter von einem Photomultiplier ausgelesen wird. Ihre Nachweiswahrscheinlichkeit wurde auf zwei Arten in Abhaengigkeit von Energie, Auftreffort und Schwelle bestimmt: Zunaechst wurde am Saturne (Saclay/F) eine Messung mit elastischer Neutron-Streuung und Neutronenmarkierung durch Nachweis von Rueckstossprotonen an  More>>
Authors:
Publication Date:
Jul 22, 1991
Product Type:
Thesis/Dissertation
Report Number:
BONN-IR-91-50
Reference Number:
SCA: 440103; PA: DEN-92:000696; SN: 92000645575
Resource Relation:
Other Information: TH: Diploma Thesis; PBD: 22 Jul 1991
Subject:
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; NEUTRON DETECTORS; PLASTIC SCINTILLATION DETECTORS; NEUTRON DETECTION; EFFICIENCY; CALIBRATION; MEV RANGE 10-100; MEV RANGE 100-1000; ENERGY DEPENDENCE; MONTE CARLO METHOD; COMPUTERIZED SIMULATION; 440103; NUCLEAR SPECTROSCOPIC INSTRUMENTATION
OSTI ID:
10112686
Research Organizations:
Bonn Univ. (Germany). Physikalisches Inst.; Bonn Univ. (Germany). Mathematisch-Naturwissenschaftliche Fakultaet
Country of Origin:
Germany
Language:
German
Other Identifying Numbers:
Other: ON: DE92759277; TRN: DE9200696
Availability:
OSTI; NTIS (US Sales Only); INIS
Submitting Site:
DEN
Size:
56 p.
Announcement Date:
Jun 30, 2005

Citation Formats

Harpes, N. Determination of the detection probability of neutron counters; Bestimmung der Nachweiswahrscheinlichkeit von Neutronenzaehlern. Germany: N. p., 1991. Web.
Harpes, N. Determination of the detection probability of neutron counters; Bestimmung der Nachweiswahrscheinlichkeit von Neutronenzaehlern. Germany.
Harpes, N. 1991. "Determination of the detection probability of neutron counters; Bestimmung der Nachweiswahrscheinlichkeit von Neutronenzaehlern." Germany.
@misc{etde_10112686,
title = {Determination of the detection probability of neutron counters; Bestimmung der Nachweiswahrscheinlichkeit von Neutronenzaehlern}
author = {Harpes, N}
abstractNote = {At the PHOENICS experiment neutron detection is done by scintillation counters consisting of 100x20x20 cm{sup 3} NE110 or Nuplex3 bars with photomultiplier tubes mounted on both ends. Calibration measurements to determine the neutron detection efficiency have been done for 4 counters at the LNS (Laboratoire National Saturne, Saclay, France). The method to obtain neutrons of well-defined energy and angle was neutron-tagging in elastic np-scattering off a CH{sub 2}-target. The efficiency was measured at different impact positions and for up to five neutron energies between 20 and 175 MeV. The efficiency shows a strong energy dependence for low energies and becomes rather constant (ca. 20%) at higher energies. The detection thresholds were varied between 1.1 and 15 MeV. Especially for low energies, the detection efficiency depends strongly on the threshold. The results of Monte Carlo simulation are in good agreement with the measurements. (orig.). [Deutsch] Die am PHOENICS-Experiment verwendeten Neutronenzaehler bestehen aus einem 20x20x100 cm{sup 3} Szintillatorblock aus Nuplex3 oder NE110, der beidseitig ueber Lichtleiter von einem Photomultiplier ausgelesen wird. Ihre Nachweiswahrscheinlichkeit wurde auf zwei Arten in Abhaengigkeit von Energie, Auftreffort und Schwelle bestimmt: Zunaechst wurde am Saturne (Saclay/F) eine Messung mit elastischer Neutron-Streuung und Neutronenmarkierung durch Nachweis von Rueckstossprotonen an CH{sub 2}- und C-Target durchgefuehrt. Dabei wurde die Nachweiswahrscheinlichkeit von vier Zaehlern bis zu fuenf Energien zwischen 20 und 175 MeV und drei verschiedenen Auftrefforten bestimmt. Bei niedrigen Energien ist die Nachweiswahrscheinlichkeit stark schwellenabhaengig, sie erreicht bei hohen Energien Werte von ca. 20%. Mit dem von Cecil et al. stammenden und auf die PHOENICS-Zaehler angepassten Monte-Carlo-Efficiencyprogramms wurde die Nachweiswahrscheinlichkeit bei allen gemessenen Energien und Auftrefforten berechnet. Die Ergebnisse sind in guter Uebereinstimmung mit den experimentell ermittelten Werten. Zusaetzlich wurde die Nachweiswahrscheinlichkeit auch fuer Energien und Auftrefforte berechnet, an denen nicht gemessen wurde. (orig.).}
place = {Germany}
year = {1991}
month = {Jul}
}