Study on the large position sensitive detector

Abstract

A large position sensitive detector, in which the charge division technique is used, with a single wire anode proportional counter is developed. The sensitive area of the detector is 1200 x 60 mm{sup 2}. The anode of the detector is NiCr wire with 10 {mu}m in diameter and 12 {Omega}/mm of its resistance. The cathodes of the detector are single side aluminized Mylar films and the thickness 12.5 {mu}m. The detector works under the condition of flushing the mixed gas of 70% Ar + 30% CH{sub 4} with one atmosphere. In order to test the detector a {sup 241}Am alpha source (about 5 MeV) is put on the collimator. The collimator can freely and accurately move along the detector window. Along the direction of anode wire, measurement is performed in the distance of every 6 cm. The total is 19 measuring points. The average intrinsic position resolution (FWHM) is about 0.6 mm, and the integral nonlinearity is better than 0.2%. This detector has been installed on the detector system of Q3D magnetic spectrometer and will be tested on the tandem accelerator with ion beam.
Authors:
Yude, Wan; Zhongzhen, Li [1] 
  1. Academia Sinica, Beijing, BJ (China). Inst. of Atomic Energy
Publication Date:
Oct 01, 1993
Product Type:
Technical Report
Report Number:
CNIC-00786; IAE-0125.
Reference Number:
SCA: 440104; PA: AIX-26:011689; EDB-95:031014; SN: 95001324324
Resource Relation:
Other Information: PBD: Oct 1993
Subject:
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; POSITION SENSITIVE DETECTORS; MAGNETIC SPECTROMETERS; BEAM MONITORING; ELECTRONIC CIRCUITS; FABRICATION; PERFORMANCE TESTING; PROPORTIONAL COUNTERS; SPATIAL RESOLUTION; 440104; HIGH ENERGY PHYSICS INSTRUMENTATION
OSTI ID:
10111709
Research Organizations:
China Nuclear Information Centre, Beijing, BJ (China)
Country of Origin:
China
Language:
Chinese
Other Identifying Numbers:
Other: ON: DE95613069; ISBN 7-5022-1052-0; TRN: CN9402342011689
Availability:
OSTI; NTIS (US Sales Only); INIS
Submitting Site:
INIS
Size:
8 p.
Announcement Date:
Jun 30, 2005

Citation Formats

Yude, Wan, and Zhongzhen, Li. Study on the large position sensitive detector. China: N. p., 1993. Web.
Yude, Wan, & Zhongzhen, Li. Study on the large position sensitive detector. China.
Yude, Wan, and Zhongzhen, Li. 1993. "Study on the large position sensitive detector." China.
@misc{etde_10111709,
title = {Study on the large position sensitive detector}
author = {Yude, Wan, and Zhongzhen, Li}
abstractNote = {A large position sensitive detector, in which the charge division technique is used, with a single wire anode proportional counter is developed. The sensitive area of the detector is 1200 x 60 mm{sup 2}. The anode of the detector is NiCr wire with 10 {mu}m in diameter and 12 {Omega}/mm of its resistance. The cathodes of the detector are single side aluminized Mylar films and the thickness 12.5 {mu}m. The detector works under the condition of flushing the mixed gas of 70% Ar + 30% CH{sub 4} with one atmosphere. In order to test the detector a {sup 241}Am alpha source (about 5 MeV) is put on the collimator. The collimator can freely and accurately move along the detector window. Along the direction of anode wire, measurement is performed in the distance of every 6 cm. The total is 19 measuring points. The average intrinsic position resolution (FWHM) is about 0.6 mm, and the integral nonlinearity is better than 0.2%. This detector has been installed on the detector system of Q3D magnetic spectrometer and will be tested on the tandem accelerator with ion beam.}
place = {China}
year = {1993}
month = {Oct}
}