You need JavaScript to view this

Equipment for Measurement ICE and e{gamma}-Coincidence Spectra; Ustanovka dlya izmereniya spektrov EhVK i e{gamma}-sovpadenij

Technical Report:

Abstract

The mini-orange magnetic spectrometer system for the measurement of conversion electron spectra and e{gamma}-coincidences has been described. Mini-orange spectrometer efficiencies for two types of magnets are determined at distances 40 and 80 mm between the source and detector. The efficiency increase and the effect of elimination of the continuous background connected with the positron radiation have been demonstrated using the mini-orange on the example of {sup 147} Tb(T{sub 1/2}=1.7 h) conversion electron spectrum in the energy range 1-2 MeV. (author). 7 refs., 2 figs.
Publication Date:
Dec 31, 1994
Product Type:
Technical Report
Report Number:
JINR-R-13-94-267
Reference Number:
SCA: 440103; PA: AIX-26:011618; EDB-95:030976; SN: 95001324262
Resource Relation:
Other Information: PBD: 1994
Subject:
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; MAGNETIC SPECTROMETERS; COINCIDENCE SPECTROMETRY; CALIBRATION; EFFICIENCY; ELECTRON SPECTRA; GAMMA SPECTRA; INTERNAL CONVERSION; SI SEMICONDUCTOR DETECTORS; 440103; NUCLEAR SPECTROSCOPIC INSTRUMENTATION
OSTI ID:
10111688
Research Organizations:
Joint Inst. for Nuclear Research, Dubna (Russian Federation). Lab. of Nuclear Problems
Country of Origin:
JINR
Language:
Russian
Other Identifying Numbers:
Other: ON: DE95613007; TRN: XJ9406578011618
Availability:
OSTI; NTIS (US Sales Only); INIS
Submitting Site:
INIS
Size:
6 p.
Announcement Date:
Jun 30, 2005

Technical Report:

Citation Formats

Sehrehehtehr, Zh, Yuldashev, M B, Gorozhankin, V M, Gromov, K Ya, Kalinnikov, V G, Kuznetsov, V V, Osipenko, B P, and Fominykh, V I. Equipment for Measurement ICE and e{gamma}-Coincidence Spectra; Ustanovka dlya izmereniya spektrov EhVK i e{gamma}-sovpadenij. JINR: N. p., 1994. Web.
Sehrehehtehr, Zh, Yuldashev, M B, Gorozhankin, V M, Gromov, K Ya, Kalinnikov, V G, Kuznetsov, V V, Osipenko, B P, & Fominykh, V I. Equipment for Measurement ICE and e{gamma}-Coincidence Spectra; Ustanovka dlya izmereniya spektrov EhVK i e{gamma}-sovpadenij. JINR.
Sehrehehtehr, Zh, Yuldashev, M B, Gorozhankin, V M, Gromov, K Ya, Kalinnikov, V G, Kuznetsov, V V, Osipenko, B P, and Fominykh, V I. 1994. "Equipment for Measurement ICE and e{gamma}-Coincidence Spectra; Ustanovka dlya izmereniya spektrov EhVK i e{gamma}-sovpadenij." JINR.
@misc{etde_10111688,
title = {Equipment for Measurement ICE and e{gamma}-Coincidence Spectra; Ustanovka dlya izmereniya spektrov EhVK i e{gamma}-sovpadenij}
author = {Sehrehehtehr, Zh, Yuldashev, M B, Gorozhankin, V M, Gromov, K Ya, Kalinnikov, V G, Kuznetsov, V V, Osipenko, B P, and Fominykh, V I}
abstractNote = {The mini-orange magnetic spectrometer system for the measurement of conversion electron spectra and e{gamma}-coincidences has been described. Mini-orange spectrometer efficiencies for two types of magnets are determined at distances 40 and 80 mm between the source and detector. The efficiency increase and the effect of elimination of the continuous background connected with the positron radiation have been demonstrated using the mini-orange on the example of {sup 147} Tb(T{sub 1/2}=1.7 h) conversion electron spectrum in the energy range 1-2 MeV. (author). 7 refs., 2 figs.}
place = {JINR}
year = {1994}
month = {Dec}
}