Abstract
The liquid scintillation counting standardization of organic and inorganic samples of ``125 I by the CIEMAT/NIST method using five different scintillators is described. The discrepancies between experimental and computed efficiencies are lower than 1.4% and 1.7%, for inorganic and organic samples, respectively, in the interval 421-226 of quenching parameter. Both organic and inorganic solutions have been standardized in terms of activity concentration to an overall uncertainty of 0.76%.
Citation Formats
Rodriguez Barquera, L, Grau Malonda, A, Los Arcos Merino, J M, and Grau Carles, A.
Liquid scintillation counting standardization of ``125 I in organic and inorganic samples by the CIEMAT/NIST method; Calibracin por centelleo lquido del ``125 I en muestras orgnicas e inorgnicas, mediante el mtodo CIEMAT/NIST.
Spain: N. p.,
1994.
Web.
Rodriguez Barquera, L, Grau Malonda, A, Los Arcos Merino, J M, & Grau Carles, A.
Liquid scintillation counting standardization of ``125 I in organic and inorganic samples by the CIEMAT/NIST method; Calibracin por centelleo lquido del ``125 I en muestras orgnicas e inorgnicas, mediante el mtodo CIEMAT/NIST.
Spain.
Rodriguez Barquera, L, Grau Malonda, A, Los Arcos Merino, J M, and Grau Carles, A.
1994.
"Liquid scintillation counting standardization of ``125 I in organic and inorganic samples by the CIEMAT/NIST method; Calibracin por centelleo lquido del ``125 I en muestras orgnicas e inorgnicas, mediante el mtodo CIEMAT/NIST."
Spain.
@misc{etde_10104725,
title = {Liquid scintillation counting standardization of ``125 I in organic and inorganic samples by the CIEMAT/NIST method; Calibracin por centelleo lquido del ``125 I en muestras orgnicas e inorgnicas, mediante el mtodo CIEMAT/NIST}
author = {Rodriguez Barquera, L, Grau Malonda, A, Los Arcos Merino, J M, and Grau Carles, A}
abstractNote = {The liquid scintillation counting standardization of organic and inorganic samples of ``125 I by the CIEMAT/NIST method using five different scintillators is described. The discrepancies between experimental and computed efficiencies are lower than 1.4% and 1.7%, for inorganic and organic samples, respectively, in the interval 421-226 of quenching parameter. Both organic and inorganic solutions have been standardized in terms of activity concentration to an overall uncertainty of 0.76%.}
place = {Spain}
year = {1994}
month = {Dec}
}
title = {Liquid scintillation counting standardization of ``125 I in organic and inorganic samples by the CIEMAT/NIST method; Calibracin por centelleo lquido del ``125 I en muestras orgnicas e inorgnicas, mediante el mtodo CIEMAT/NIST}
author = {Rodriguez Barquera, L, Grau Malonda, A, Los Arcos Merino, J M, and Grau Carles, A}
abstractNote = {The liquid scintillation counting standardization of organic and inorganic samples of ``125 I by the CIEMAT/NIST method using five different scintillators is described. The discrepancies between experimental and computed efficiencies are lower than 1.4% and 1.7%, for inorganic and organic samples, respectively, in the interval 421-226 of quenching parameter. Both organic and inorganic solutions have been standardized in terms of activity concentration to an overall uncertainty of 0.76%.}
place = {Spain}
year = {1994}
month = {Dec}
}