DOE Data Explorer title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Experimental calibration procedures for rotating Lorentz-force flowmeters

Abstract

Rotating Lorentz-force flowmeters are a novel and useful technology with a range ofapplications in a variety of different industries. However, calibrating these flowmeters canbe challenging, time-consuming, and expensive. In this paper, simple calibration proceduresfor rotating Lorentz-force flowmeters are presented. These procedures eliminate the need forexpensive equipment, numerical modeling, redundant flowmeters, and system down-time.The calibration processes are explained in a step-by-step manner and compared to experimental results.

Authors:
; ; ;
Publication Date:
DOE Contract Number:  
AC02-09CH11466
Research Org.:
Princeton Plasma Physics Lab. (PPPL), Princeton, NJ (United States)
Sponsoring Org.:
U. S. Department of Energy
Keywords:
Lorentz force
OSTI Identifier:
1562012
DOI:
https://doi.org/10.11578/1562012

Citation Formats

Hvasta, M G, Slighton, N T, Kolemen, E, and Fisher, A E. Experimental calibration procedures for rotating Lorentz-force flowmeters. United States: N. p., 2017. Web. doi:10.11578/1562012.
Hvasta, M G, Slighton, N T, Kolemen, E, & Fisher, A E. Experimental calibration procedures for rotating Lorentz-force flowmeters. United States. doi:https://doi.org/10.11578/1562012
Hvasta, M G, Slighton, N T, Kolemen, E, and Fisher, A E. 2017. "Experimental calibration procedures for rotating Lorentz-force flowmeters". United States. doi:https://doi.org/10.11578/1562012. https://www.osti.gov/servlets/purl/1562012. Pub date:Sun Jan 01 00:00:00 EST 2017
@article{osti_1562012,
title = {Experimental calibration procedures for rotating Lorentz-force flowmeters},
author = {Hvasta, M G and Slighton, N T and Kolemen, E and Fisher, A E},
abstractNote = {Rotating Lorentz-force flowmeters are a novel and useful technology with a range ofapplications in a variety of different industries. However, calibrating these flowmeters canbe challenging, time-consuming, and expensive. In this paper, simple calibration proceduresfor rotating Lorentz-force flowmeters are presented. These procedures eliminate the need forexpensive equipment, numerical modeling, redundant flowmeters, and system down-time.The calibration processes are explained in a step-by-step manner and compared to experimental results.},
doi = {10.11578/1562012},
journal = {},
number = ,
volume = ,
place = {United States},
year = {Sun Jan 01 00:00:00 EST 2017},
month = {Sun Jan 01 00:00:00 EST 2017}
}

Works referenced in this record:

Experimental calibration procedures for rotating Lorentz-force flowmeters
journal, July 2017


    Works referencing / citing this record:

    Experimental calibration procedures for rotating Lorentz-force flowmeters
    journal, July 2017