Actinic EUV mask inspection beyond 0.25 NA
The SEMATECH Berkeley Actinic Inspection Tool (AIT) is an EUV-wavelength mask inspection microscope designed for direct aerial image measurements, and pre-commercial EUV mask research. Operating on a synchrotron bending magnet beamline, the AIT uses an off-axis Fresnel zoneplate lens to project a high-magnification EUV image directly onto a CCD camera. We present the results of recent system upgrades that have improved the imaging resolution, illumination uniformity, and partial coherence. Benchmarking tests show image contrast above 75% for 100-nm mask features, and significant improvements and across the full range of measured sizes. The zoneplate lens has been replaced by an array of user-selectable zoneplates with higher magnification and NA values up to 0.0875, emulating the spatial resolution of a 0.35-NA 4x EUV stepper. Illumination uniformity is above 90% for mask areas 2-{micro}m-wide and smaller. An angle-scanning mirror reduces the high coherence of the synchrotron beamline light source giving measured {sigma} values of approximately 0.125 at 0.0875 NA.
- Research Organization:
- Lawrence Berkeley National Lab. (LBNL), Berkeley, CA (United States)
- Sponsoring Organization:
- Materials Sciences Division
- DOE Contract Number:
- DE-AC02-05CH11231
- OSTI ID:
- 941172
- Report Number(s):
- LBNL-991E; JVTBD9; TRN: US0807299
- Journal Information:
- JVST B, Related Information: Journal Publication Date: Nov/Dec; ISSN 0734-211X
- Country of Publication:
- United States