Title: Fully automated four-column capillary LC-MS system for maximizing throughput in proteomic analyses

Journal Article · · Analytical Chemistry, 80(1):294-302
DOI:https://doi.org/10.1021/ac701727r· OSTI ID:921809

We describe a 4-column, high-pressure capillary liquid chromatography (LC) system for robust, high-throughput LC-MS(/MS) analyses. This system performs multiple LC separations in parallel, but staggers each of them such that the data-rich region of each separation is sampled sequentially. By allowing nearly continuous data acquisition, this design maximizes the use of the mass spectrometer. Each analytical column is connected to a corresponding ESI emitter in order to avoid the use of post-column switching and associated dead volume issues. Encoding translation stages are employed to sequentially position the emitters at the MS inlet. The high reproducibility of this system is demonstrated using consecutive analyses of global tryptic digest of the microbe Shewanella oneidensis.

Research Organization:
Pacific Northwest National Lab. (PNNL), Richland, WA (United States)
Sponsoring Organization:
USDOE
DOE Contract Number:
AC05-76RL01830
OSTI ID:
921809
Report Number(s):
PNNL-SA-56532; 400412000; TRN: US200802%%913
Journal Information:
Analytical Chemistry, 80(1):294-302, Vol. 80, Issue 1
Country of Publication:
United States
Language:
English