Title: ACCU Core Sampling/Storage Device for VOC Analysis

Abstract

The Accu Core sampler system consists of alternating cylindrical clear acrylic sections and one-inch cylindrical stainless steel sections arranged in clear shrink wrap. The set of alternating acrylic and stainless steel sections in the shrink wrap are designed to fit in a Geoprobe dual-tube penetrometer for collection of continuous soil cores. The clear acrylic sections can have 1/2-inch access holes for easy soil headspace screening without violating the integrity of the adjacent stainless steel sections. The Accu Core sampler system can be used to store a soil sample collected in the stainless steel section by capping the ends of the section so it becomes a sample storage container. The sampler system can also be used to collect a subsurface soil sample in one of the sections that can be directly extruded from the section into a container for storage during shipment to the laboratory. In addition, the soil in a sampler section can be quickly sub-sampled using a coring tool and extruded into a storage container so the integrity of the soil is not disrupted and the potential for VOC loss during sub-sampling is greatly reduced. A field validation study was conducted to evaluate the performance of the Accu Coremore » sampler to store VOC soil samples during transportation to the laboratory for analysis and to compare the performance of the Accu Core with current sampling and storage techniques, all of which require sub-sampling when the soil sample is brought to the surface. During some of the validation testing, the acrylic sections having access holes for headspace screening were included in the Accu Core sampler configuration and soil in these sections was screened to show the usefulness of the sample screening capability provided by the Accu Core system. This report presents the results of the field validation study as well as recommendations for the Accu Core sampler system.« less

Authors:
; ;
Publication Date:
Research Org.:
Western Research Institute
Sponsoring Org.:
USDOE
OSTI Identifier:
908781
DOE Contract Number:
FC26-98FT40323
Resource Type:
Technical Report
Country of Publication:
United States
Language:
English
Subject:
36 MATERIALS SCIENCE; CONFIGURATION; CONTAINERS; ORGANIC COMPOUNDS; PENETROMETERS; PERFORMANCE; RECOMMENDATIONS; SAMPLERS; SAMPLING; SOILS; STAINLESS STEELS; STORAGE; TESTING; VALIDATION; VOLATILE MATTER

Citation Formats

Susan S. Sorini, John F. Schabron, and Mark M. Sanderson. ACCU Core Sampling/Storage Device for VOC Analysis. United States: N. p., 2007. Web. doi:10.2172/908781.
Susan S. Sorini, John F. Schabron, & Mark M. Sanderson. ACCU Core Sampling/Storage Device for VOC Analysis. United States. doi:10.2172/908781.
Susan S. Sorini, John F. Schabron, and Mark M. Sanderson. Mon . "ACCU Core Sampling/Storage Device for VOC Analysis". United States. doi:10.2172/908781. https://www.osti.gov/servlets/purl/908781.
@article{osti_908781,
title = {ACCU Core Sampling/Storage Device for VOC Analysis},
author = {Susan S. Sorini and John F. Schabron and Mark M. Sanderson},
abstractNote = {The Accu Core sampler system consists of alternating cylindrical clear acrylic sections and one-inch cylindrical stainless steel sections arranged in clear shrink wrap. The set of alternating acrylic and stainless steel sections in the shrink wrap are designed to fit in a Geoprobe dual-tube penetrometer for collection of continuous soil cores. The clear acrylic sections can have 1/2-inch access holes for easy soil headspace screening without violating the integrity of the adjacent stainless steel sections. The Accu Core sampler system can be used to store a soil sample collected in the stainless steel section by capping the ends of the section so it becomes a sample storage container. The sampler system can also be used to collect a subsurface soil sample in one of the sections that can be directly extruded from the section into a container for storage during shipment to the laboratory. In addition, the soil in a sampler section can be quickly sub-sampled using a coring tool and extruded into a storage container so the integrity of the soil is not disrupted and the potential for VOC loss during sub-sampling is greatly reduced. A field validation study was conducted to evaluate the performance of the Accu Core sampler to store VOC soil samples during transportation to the laboratory for analysis and to compare the performance of the Accu Core with current sampling and storage techniques, all of which require sub-sampling when the soil sample is brought to the surface. During some of the validation testing, the acrylic sections having access holes for headspace screening were included in the Accu Core sampler configuration and soil in these sections was screened to show the usefulness of the sample screening capability provided by the Accu Core system. This report presents the results of the field validation study as well as recommendations for the Accu Core sampler system.},
doi = {10.2172/908781},
journal = {},
number = ,
volume = ,
place = {United States},
year = {Mon Apr 30 00:00:00 EDT 2007},
month = {Mon Apr 30 00:00:00 EDT 2007}
}

Technical Report:

Save / Share:
  • Soil sampling and storage practices for volatile organic analysis must be designed to minimize loss of volatile organic compounds (VOCs) from samples. The En Core{reg_sign} sampler is designed to collect and store soil samples in a manner that minimizes loss of contaminants due to volatilization and/or biodegradation. An American Society for Testing and Materials (ASTM) standard practice, D 6418, Standard Practice for Using the Disposable En Core Sampler for Sampling and Storing Soil for Volatile Organic Analysis, describes use of the En Core sampler to collect and store a soil sample of approximately 5 grams or 25 grams for volatilemore » organic analysis. To support the ASTM practice, a study was performed to estimate the precision of the performance of the 5-gram and 25-gram En Core samplers to store soil samples spiked with low concentrations of VOCs. This report discusses revision of ASTM Practice D 6418 to include information on the precision of the En Core devices and to reference an ASTM research report on the precision study. This report also discusses revision of the ASTM practice to list storage at -12 {+-} 2 C for up to 14 days and at 4 {+-} 2 C for up to 48 hours followed by storage at -12 {+-} 2C for up to 5 days as acceptable conditions for samples stored in the En Core devices. Data supporting use of these storage conditions are given in an appendix to the practice and are presented in the research report referenced for the precision study. Prior to this revision, storage in the device was specified at 4 {+-} 2 C for up to 48 hours. The En Core sampler is designed to collect soil samples for VOC analysis at the soil surface. To date, a sampling tool for collecting and storing subsurface soil samples for VOC analysis does not exist. Development of a subsurface VOC sampling/storage device was initiated in 1999. This device, which is called the Accu Core sampler, is designed so that a soil sample can be collected below the surface using a penetrometer and transported to the laboratory for analysis in the same container. During the past year, prototype devices have been tested for their performance in storing soil samples containing low concentrations of VOCs. The Accu Core sampler testing is also described in this report.« less
  • Soil sampling and storage practices for volatile organic analysis must be designed to minimize loss of volatile organic compounds (VOCs) from samples. The En Core{reg_sign} sampler is designed to collect and store soil samples in a manner that minimizes loss of contaminants due to volatilization and/or biodegradation. An ASTM International (ASTM) standard practice, D 6418, Standard Practice for Using the Disposable En Core Sampler for Sampling and Storing Soil for Volatile Organic Analysis, describes use of the En Core sampler to collect and store a soil sample of approximately five grams or 25 grams for volatile organic analysis. Prior tomore » the study described in this report, D 6418 specified sample storage in the En Core sampler at 4 {+-} 2 C for up to 48 hours; at -12 {+-} 2 C for up to 14 days; or at 4 {+-} 2 C for up to 48 hours followed by storage at -12 {+-} 2 C for up to five days to minimize loss of volatile compounds due to volatilization and/or biodegradation. The study described in this report was conducted to evaluate the performance of the disposable En Core sampler to store low concentrations of VOCs in soil at -7 {+-} 1 C and -21 {+-} 2 C. In the study, data on the performance of the En Core sampler to store soils spiked with low-level concentrations of VOCs at 4 {+-} 2 C for 48 hours followed by storage at -7 {+-} 1 C for five days, at -7 {+-} 1 C for 14 days, at 4 {+-} 2 C for 48 hours followed by storage at -21 {+-} 2 C for five days, and at -21 {+-} 2 C for 14 days were generated. Based on these data, a new revision of D 6418 was prepared and balloted in ASTM. The new revision, which was approved on February 1, 2004, now specifies sample storage in the En Core sampler at 4 {+-} 2 C for up to 48 hours; -7 to -21 C for up to 14 days; or 4 {+-} 2 C for up to 48 hours followed by storage at -7 to -21 C for up to five days. The En Core sampler is designed to collect soil samples for VOC analysis at the soil surface. To date, a sampling tool for collecting and storing subsurface soil samples for VOC analysis does not exist. Development of a subsurface VOC sampling/storage device was initiated in 1999. This device, which is called the Accu Core sampler, is designed so that a soil sample can be collected below the surface using a dual-tube penetrometer and transported to the laboratory for analysis in the same container. During the past year, prototype devices have been tested for their performance in storing soil samples containing low concentrations of VOCs. Evaluation of the various Accu Core prototypes and the design selected for additional validation testing are described in this report.« less
  • Soil sampling and storage practices for volatile organic analysis must be designed to minimize loss of volatile organic compounds (VOCs) from samples. The En Core{reg_sign} sampler is designed to collect and store soil samples in a manner that minimizes loss of contaminants due to volatilization and/or biodegradation. An ASTM International (ASTM) standard practice, D 6418, Standard Practice for Using the Disposable En Core Sampler for Sampling and Storing Soil for Volatile Organic Analysis, describes use of the En Core sampler to collect and store a soil sample of approximately 5 grams or 25 grams for volatile organic analysis and specifiesmore » sample storage in the En Core sampler at 4 {+-} 2 C for up to 48 hours; -7 to -21 C for up to 14 days; or 4 {+-} 2 C for up to 48 hours followed by storage at -7 to -21 C for up to five days. This report discusses activities performed during the past year to promote and continue acceptance of the En Core samplers based on their performance to store soil samples for VOC analysis. The En Core sampler is designed to collect soil samples for VOC analysis at the soil surface. To date, a sampling tool for collecting and storing subsurface soil samples for VOC analysis is not available. Development of a subsurface VOC sampling/storage device was initiated in 1999. This device, which is called the Accu Core{trademark} sampler, is designed so that a soil sample can be collected below the surface using a dual-tube penetrometer and transported to the laboratory for analysis in the same container. Laboratory testing of the current Accu Core design shows that the device holds low-level concentrations of VOCs in soil samples during 48-hour storage at 4 {+-} 2 C and that the device is ready for field evaluation to generate additional performance data. This report discusses a field validation exercise that was attempted in Pennsylvania in 2004 and activities being performed to plan and conduct a field validation study in 2006. A draft ASTM practice describing use of the Accu Core sampler is being prepared. An update on the status of the ASTM practice is given in this report.« less
  • This Sampling and Analysis Plan (SAP) details the sampling and analyses to be performed on fuel canisters transferred to the Weasel Pit of the 105-KW fuel storage basin. The radionuclide content of the liquid and gas in the canisters must be evaluated to support the shipment of fuel elements to the 300 Area in support of the fuel characterization studies (Abrefah, et al. 1994, Trimble 1995). The following sections provide background information and a description of the facility under investigation, discuss the existing site conditions, present the constituents of concern, outline the purpose and scope of the investigation, outline themore » data quality objectives (DQO), provide analytical detection limit, precision, and accuracy requirements, and address other quality assurance (QA) issues.« less
  • This Sampling and Analysis Plan describes the equipment,procedures and techniques for obtaining gas and liquid samples from sealed K West fuel canisters. The analytical procedures and quality assurance requirements for the subsequent laboratory analysis of the samples are also discussed.