skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Technical Performance of the Luxel Al2O3:C Optically Stimulated Luminescence Dosemeter Element at Radiation Oncology and Nuclear Accident Dose Levels

Journal Article · · Radiation Protection Dosimetry, 8 pages
DOI:https://doi.org/10.1093/rpd/ncl500· OSTI ID:898104

The dose ranges typical for radiation oncology and nuclear accident dosimetry are on the order of 2?70 Gy and 0.1?5 Gy, respectively. In terms of solid-state passive dosimetry; thermoluminescent (TL) materials historically have been used extensively for these two applications, with silver-halide, leuco-dye, and BaFBr:Eu-based films being used on a more limited basis than TL for radiation oncology. This present work provides results on the performance of a film based on an aluminum oxide, Al2O3:C, for these dosimetry applications, using the optically-stimulated luminescence (OSL) readout method. There have been few investigations of Al2O3:C performance at radiation oncology and nuclear accident dose levels, and these have included minimal dosimetric and environmental effects information. Based on investigations already published, the authors of this present study determined that overall improvements over film and TLDs for this Al2O3:C OSL technology at radiation oncology and nuclear accident dose levels may include (1) a more tissue-equivalent response to photons compared to X-ray film, (2) higher sensitivity, (3) ability to reread dosemeters, and (4) diagnostic capability using small-area imaging. The results of the present investigation indicate that additional favorable performance characteristics for the Al2O3:C dosemeter are a wide dynamic range(0.001 to 100 Gy), a response insensitive to temperature and moisture over a wide range, negligible dose rate dependence, and minimal change in post-irradiation response. As a radiation detection medium, this OSL phosphor offers an assortment of dosimetry properties that will permit it to compete with current radiation detection technologies such as silver-halide, leuco-dye, and photostimulable-phosphor based films, as well as TLDs.

Research Organization:
Pacific Northwest National Lab. (PNNL), Richland, WA (United States)
Sponsoring Organization:
USDOE
DOE Contract Number:
AC05-76RL01830
OSTI ID:
898104
Report Number(s):
PNWD-SA-7505; TRN: US0701572
Journal Information:
Radiation Protection Dosimetry, 8 pages, Journal Name: Radiation Protection Dosimetry, 8 pages
Country of Publication:
United States
Language:
English