Calculation of Photoneutrons from Varian Clinac Accelerators and Their Transmissions in Materials
Monte Carlo calculations of the giant-dipole-resonance photoneutrons (GRN) around the Varian Clinac 2100C/2300C medical accelerator heads (10-20 MV modes) were made using the coupled EGS4-MORSE code. The actual head materials and geometries were simulated in great detail using the Combinatorial Geometry facility of MORSE. The neutron production (i.e., sites and yields) was calculated with EGS4 and, then, the neutron transport in the accelerator head was done with MORSE. Both the evaporation and direct neutron components of the GRN were considered by incorporating the EVAP4 code and an empirical algorithm, respectively, into MORSE. With the calculated neutron spectra around the head as source terms, MCNP4a was used to estimate the corresponding dose equivalent transmission (considering both the neutron attenuation and the build-up of captured gamma rays) in several different types of concrete. The calculated results of the absolute neutron fluence and spectra around the heads, as well as the transmission curves, are presented and discussed.
- Research Organization:
- SLAC National Accelerator Lab., Menlo Park, CA (United States)
- Sponsoring Organization:
- USDOE
- DOE Contract Number:
- AC02-76SF00515
- OSTI ID:
- 895272
- Report Number(s):
- SLAC-PUB-7404; TRN: US0700425
- Resource Relation:
- Conference: Presented at International Conference on Radiation Dosimetry and Safety, Taipei, Taiwan,, 31 Mar - 2 Apr 1997
- Country of Publication:
- United States
- Language:
- English
Similar Records
The neutron dose equivalent evaluation and shielding at the maze entrance of a Varian Clinac 23EX treatment room
SU-E-T-403: Measurement of the Neutron Ambient Dose Equivalent From the TrueBeam Linac Head and Varian 2100 Clinac