We have designed a high-efficiency neutron detector for passive neutron coincidence and multiplicity counting of dirty scrap and bulk samples of plutonium. The counter will be used for the measurement of impure plutonium samples at the JNC MOX fabrication facility in Japan. The counter can also be used to create working standards from bulk process MOX. The detector uses advanced design 3He tubes to increase the efficiency and to shorten the neutron die-away time. The efficiency is 64% and the die-away time is 19.1 μs. The Epithermal Neutron Multiplicity Counter (ENMC) is designed for high-precision measurements of bulk plutonium samples with diameters of less than 200 mm. The average neutron energy from the sample can be measured using the ratio of the inner ring of He-3 tubes to the outer ring. This report describes the hardware, performance, and calibration for the ENMC.