Title: Pulse Rise Time Characterization of a High Pressure Xenon Gamma Detector for use in Resolution Enhancement

Conference ·
OSTI ID:804476

High pressure xenon ionization chamber detectors are possible alternatives to traditional thallium doped sodium iodide (NaI(Tl)) and hyperpure germanium as gamma spectrometers in certain applications. Xenon detectors incorporating a Frisch grid exhibit energy resolutions comparable to cadmium/zinc/telluride (CZT) (e.g. 2% {at} 662keV) but with far greater sensitive volumes. The Frisch grid reduces the position dependence of the anode pulse risetimes, but it also increases the detector vibration sensitivity, anode capacitance, voltage requirements and mechanical complexity. We have been investigating the possibility of eliminating the grid electrode in high-pressure xenon detectors and preserving the high energy resolution using electronic risetime compensation methods. A two-electrode cylindrical high pressure xenon gamma detector coupled to time-to-amplitude conversion electronics was used to characterize the pulse rise time of deposited gamma photons. Time discrimination was used to characterize the pulse rise time versus photo peak position and resolution. These data were collected to investigate the effect of pulse rise time compensation on resolution and efficiency.

Research Organization:
Fluor Hanford, Richland, WA (United States)
Sponsoring Organization:
USDOE Office of Environmental Management (EM) (US)
DOE Contract Number:
AC06-96RL13200
OSTI ID:
804476
Report Number(s):
HNF-6815-FP, Rev.0; TRN: US0300960
Resource Relation:
Conference: Conference title not supplied, Conference location not supplied, Conference dates not supplied; Other Information: PBD: 25 Aug 2000
Country of Publication:
United States
Language:
English

Similar Records

Multi-channel front-end ASIC for a 3D position-sensitive detector
Journal Article · Mon Feb 07 00:00:00 EST 2022 · Journal of Instrumentation · OSTI ID:804476

Front-end ASIC for spectroscopic readout of virtual Frisch-grid CZT bar sensors
Journal Article · Fri May 24 00:00:00 EDT 2019 · Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment · OSTI ID:804476

High Pressure XENON Gamma-Ray Spectrometers for Field Use
Technical Report · Mon Feb 16 00:00:00 EST 2004 · OSTI ID:804476