Advanced research in instrumentation and control technology: Capacitance flowmeter
The purpose of this task is to develop and apply instruments to measure the solid feedrate in coal gasification plant gas-solid feedlines, interstage lines, and recycling lines. To date, one of the most promising techniques for mass flow measurements in a solid/gas flow stream is the use of the nonintrusive capacitance sensor. The performance of the capacitance flowmeter (CF) with an improved electrode geometry has been measured experimentally in a gas-entrained 1000 ..mu..m glass bead flow stream at the Argonne National Laboratory (ANL) Solid/Gas Flow Test Facility (SGFTF). The CF measured the solid volume fraction (SVF) and particle velocity and the results were compared with time-of-flight measurements of short-lived radioactive particles. 21 refs., 9 figs.
- Research Organization:
- Argonne National Lab., IL (USA)
- DOE Contract Number:
- W-31109-ENG-38
- OSTI ID:
- 6956267
- Report Number(s):
- CONF-860961-7; ON: DE87004654
- Resource Relation:
- Conference: Instrumentation, components and materials contractors' meeting, Morgantown, WV, USA, 16 Sep 1986; Other Information: Portions of this document are illegible in microfiche products
- Country of Publication:
- United States
- Language:
- English
Similar Records
Performance characteristics of a capacitive mass flowmeter in a coal-oil slurry at the ANL Solid/Liquid Test Facility
Preliminary report on development of nonintrusive flow instrumentation at the ANL Solid-Liquid Test Facility
Related Subjects
47 OTHER INSTRUMENTATION
COAL GASIFICATION PLANTS
MEASURING INSTRUMENTS
FLOWMETERS
DESIGN
PERFORMANCE TESTING
PNEUMATIC TRANSPORT
TEST FACILITIES
SOLIDS
ELECTRODES
FLOW RATE
GLASS
TRACER TECHNIQUES
INDUSTRIAL PLANTS
ISOTOPE APPLICATIONS
METERS
TESTING
TRANSPORT
010404* - Coal
Lignite
& Peat- Gasification
440300 - Miscellaneous Instruments- (-1989)