Unexpectedly large dose rate dependent output from a linear accelerator
Journal Article
·
· Med. Phys.; (United States)
During our routine calibration of a Varian Clinac-20 linear accelerator, the absorbed dose for a fixed monitor unit (mu) was found to decrease with increasing dose rate. Between dose rates of 100 and 500 mu/min, there was up to 20% difference in absorbed dose for a 20-MeV electron beam. The cause of this problem was a failure in the electronics circuit of an integrating board. This paper presents our analysis of the problem and suggests a possible means of isolating such a failure to warn technologists, physicists, and engineers.
- Research Organization:
- Department of Radiation Oncology, Albany Medical College and Veterans Administration Hospital, Albany, New York 12208
- OSTI ID:
- 6841196
- Journal Information:
- Med. Phys.; (United States), Vol. 15:5
- Country of Publication:
- United States
- Language:
- English
Similar Records
Development of an accurate EPID-based output measurement and dosimetric verification tool for electron beam therapy
Improvements in dose accuracy delivered with static-MLC IMRT on an integrated linear accelerator control system
Ultra-high dose rate FLASH irradiator at the radiological research accelerator facility
Journal Article
·
Wed Jul 15 00:00:00 EDT 2015
· Medical Physics
·
OSTI ID:6841196
Improvements in dose accuracy delivered with static-MLC IMRT on an integrated linear accelerator control system
Journal Article
·
Tue May 15 00:00:00 EDT 2012
· Medical Physics
·
OSTI ID:6841196
+2 more
Ultra-high dose rate FLASH irradiator at the radiological research accelerator facility
Journal Article
·
Thu Dec 22 00:00:00 EST 2022
· Scientific Reports
·
OSTI ID:6841196
+4 more
Related Subjects
43 PARTICLE ACCELERATORS
ELECTRON DOSIMETRY
DOSE RATES
LINEAR ACCELERATORS
ELECTRONIC CIRCUITS
MONITORING
RADIOTHERAPY
CALIBRATION
ELECTRON BEAMS
FAILURES
MEV RANGE 10-100
ACCELERATORS
BEAMS
DOSIMETRY
ENERGY RANGE
LEPTON BEAMS
MEDICINE
MEV RANGE
NUCLEAR MEDICINE
PARTICLE BEAMS
RADIOLOGY
THERAPY
430100* - Particle Accelerators- Design
Development
& Operation
ELECTRON DOSIMETRY
DOSE RATES
LINEAR ACCELERATORS
ELECTRONIC CIRCUITS
MONITORING
RADIOTHERAPY
CALIBRATION
ELECTRON BEAMS
FAILURES
MEV RANGE 10-100
ACCELERATORS
BEAMS
DOSIMETRY
ENERGY RANGE
LEPTON BEAMS
MEDICINE
MEV RANGE
NUCLEAR MEDICINE
PARTICLE BEAMS
RADIOLOGY
THERAPY
430100* - Particle Accelerators- Design
Development
& Operation