Title: Accurate wavelength meter for cw lasers

Journal Article · · Appl. Opt.; (United States)
DOI:https://doi.org/10.1364/AO.25.000639· OSTI ID:6063514

An accurate wavelength meter (wavemeter) for cw lasers has been developed. The wavemeter is a Michelson-type fringe-counting two-beam interferometer operated in vacuum. A multicoincidence fringe counting technique has also been developed by which fringe fractions are determined with an uncertainty of 1/500 of a fringe. The result of wavelength comparisons between 612- and 633-nm I/sub 2/ stabilized He-Ne lasers showed good agreement with the value recommended by CIPM 1983. The discrepancy was <4 x 10/sup -10/ and the standard deviation of the comparison was 4 x 10/sup -10/.

Research Organization:
National Research Laboratory of Metrology, 1-1-4 Umezono, Sakura-mura, Niiharigun, Ibaraki 305, Japan
OSTI ID:
6063514
Journal Information:
Appl. Opt.; (United States), Vol. 25:5
Country of Publication:
United States
Language:
English