Title: High-resolution digital dosimetric system for spatial characterization of radiation fields using a thermoluminescent CaF/sub 2/:Dy crystal

Journal Article · · Med. Phys.; (United States)
OSTI ID:5879433

A high-resolution digital dosimetric system has been developed for the spatial characterization of radiation fields. The system comprises the following: 0.5-mm-thick, 25-mm-diam CaF/sub 2/:Dy thermoluminescent crystal; intensified charge coupled device video camera; video cassette recorder; and a computerized image processing subsystem. The optically flat single crystal is used as a radiation imaging device and the subsequent thermally stimulated phosphorescence is viewed by the intensified camera for further processing and analysis. Parameters governing the performance characteristics of the system were measured. A spatial resolution limit of 31 +- 2 ..mu..m (1sigma) corresponding to 16 +- 1 line pair/mm measured at the 4% level of the modulation transfer function has been achieved. The full width at half maximum of the line spread function measured independently by the slit method or derived from the edge response function was found to be 69 +- 4 ..mu..m (1sigma). The high resolving power, speed of readout, good precision, wide dynamic range, and the large image storage capacity make the system suitable for the digital mapping of the relative distribution of absorbed doses for various small radiation fields and the edges of larger fields.

Research Organization:
Physics Department, Kuwait University, P.O. Box 5969, Kuwait
OSTI ID:
5879433
Journal Information:
Med. Phys.; (United States), Vol. 13:3
Country of Publication:
United States
Language:
English