skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Charge cloud diffusion in a sealed position sensitive proportional counter filled with xenon-methane and xenon-carbon dioxide

Journal Article · · IEEE Trans. Nucl. Sci.; (United States)

The influence of diffusion in a xenon filled imaging proportional counter has been studied in two gas mixtures, xenon-methane and xenon-carbon dioxide. As expected in xenon-methane, by increasing the charge cloud diameter from 280 to 1130 microns, interpolation between the 2mm pitch anode wires is greatly improved. The use of a long diffusion region also delays secondary avalanches and consequently improves the resolution. In xenon-carbon dioxide the charge clouds are smaller, and the improvement in interpolation is less. The resolution in this case is not limited by secondary avalanches and diffusion causes a very small degradation.

Research Organization:
X-Ray Astronomy Group, Physics Department, Leicester University Leicester
OSTI ID:
5766053
Journal Information:
IEEE Trans. Nucl. Sci.; (United States), Vol. 30:1
Country of Publication:
United States
Language:
English