Phantom feet on digital radionuclide images and other scary computer tales
Journal Article
·
· Journal of Nuclear Medicine; (USA)
OSTI ID:5470687
- William Beaumont Hospital, Royal Oak, MI (USA)
Malfunction of a computer-assisted digital gamma camera is reported. Despite what appeared to be adequate acceptance testing, an error in the system gave rise to switching of images and identification text. A suggestion is made for using a hot marker, which would avoid the potential error of misinterpretation of patient images.