skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Radon-immune air monitor for plutonium

Technical Report ·
DOI:https://doi.org/10.2172/5215341· OSTI ID:5215341

The gross alpha activity in particulate-associated daughters of radon in the work environments may be many times that of one MPC/sub a/ of /sup 239/Pu or /sup 238/Pu. The daughters of radon emit both alpha and beta particles. The ratio of the alpha-to-beta activity is variable and a function of several factors. In spite of this variability, the ratio changes slowly with time and has upper and lower bounds in buildings that have controlled ventilation. This provides the basis for the development of an air monitor in which the radon interference is virtually eliminated. The radon-immune air monitor has three detectors: one that detects the alpha activity on the filter, one that detects the beta activity on the filter plus gamma background, and one that observes gamma background. The counts from these detectors are fed to a computer that is programmed with an algorithm for computing the non-radon-associated alpha activity.

Research Organization:
Oak Ridge National Lab. (ORNL), Oak Ridge, TN (United States)
DOE Contract Number:
W-7405-ENG-26
OSTI ID:
5215341
Report Number(s):
ORNL/TM-8389; ON: DE82019892
Country of Publication:
United States
Language:
English