Title: Passive colorimetric dosimeter tubes for ammonia, carbon monoxide, carbon dioxide, hydrogen sulfide, nitrogen dioxide and sulfur dioxide

Journal Article · · Am. Ind. Hyg. Assoc. J.; (United States)

Colorimetric, stain length, personal dosimeters operating by gas diffusion have been developed to determine worker exposure for up to an eight-hour period for several inorganic airborne contaminants in the range of their Threshold Limit Values. Length of stain, colorimetric dosimeters have been made for the detection of ammonia (NH/sub 3/), carbon monoxide (CO), carbon dioxide (CO/sub 2/), hydrogen sulfide (H/sub 2/S), nitrogen dioxide (NO/sub 2/), and sulfur dioxide (SO/sub 2/) in air. For each gas detection system, the sampler depends on the transfer of the gas by diffusion into a glass tube containing a colorimetric length of stain indicator. The stain length developed in a given period of time is compared to a calibration chart to determine, on the spot, the average gas concentration to which the dosimeter has been exposed. These dosimeters are known by the trade name Vapor Gard.

Research Organization:
Mine Safety Appliances Co., Pittsburgh, PA
OSTI ID:
5100854
Journal Information:
Am. Ind. Hyg. Assoc. J.; (United States), Vol. 46:7
Country of Publication:
United States
Language:
English