skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Performance of an imaging gas scintillation proportional counter with microchannel plate read-out

Conference · · IEEE Trans. Nucl. Sci.; (United States)
OSTI ID:5029399

Energy and position resolution measurements have been carried out on an imaging gas scintillation proportional counter with microchannelplate read-out. Despite the degradation of the CsI photocathode due to exposure to air, an energy and position resolution of 14.8% and 1.8 mm at 5.9 keV and 5.4 keV, respectively, has been obtained. The measured values are in very good agreement with the predictions based on the measured quantum efficiency of the CsI photocathode. On the basis of this agreement one predicts that energy and position resolutions of better than 10% and 1 mm, respectively, can be achieved.

Research Organization:
Lab. for Space Research Leiden, Leiden
OSTI ID:
5029399
Report Number(s):
CONF-841007-; TRN: 85-026107
Journal Information:
IEEE Trans. Nucl. Sci.; (United States), Vol. NS-32:1; Conference: Nuclear science symposium, Orlando, FL, USA, 31 Oct 1984
Country of Publication:
United States
Language:
English