Title: A SYSTEM FOR CONTINUOUS MEASUREMENT OF RADIOACTIVITY IN FLOWING STREAMS

Journal Article · · Nature
DOI:https://doi.org/10.1038/194034a0· OSTI ID:4783902

An apparatus for the determination of alpha or BETA radioactivity in either circulating liquid or gas streams was developed. Solid anthracene crystals are used. The detector consists of a Lucite light pipe coated with titanium dioxide and coupled to two photomultipliers which are in turn fed to appropriate coincidence type circuitry. The detection cell, which consists of a 9-mm OD glass tube with appropriate fittings on either end, was packed with anthracene crystals. A glass frit, or glass wool, was incorporated in the cell on the downstream side to contain the anthracene and a pledget of glass wool was placed above the anthracene on the upstream side. Carbon-14 counting efficiency was found to be of the order of 50% with a coincident background from 100 divisions to infinity of less than 40 cpm at 900 v. Tritium counting efficiency was in the range of 7% and the integral background from 100 divisions to infinity was about 90 cpm at 1130 v. Discussion is also given on the electronics of the detector and the performance in closed flowing systems and gas analysis. (P.C.H.)