Title: HOT-CATHODE MAGNETRON IONIZATION GAUGE WITH AN ELECTRON MULTIPLIER ION DETECTOR FOR THE MEASUREMENT OF EXTREME ULTRA-HIGH VACUA

Journal Article · · Rev. Sci. Instr.
DOI:https://doi.org/10.1063/1.1718411· OSTI ID:4485173

A hot-cathode magnetron ionization gage capable of measuring pressures as low as 10/sup -14/ Torr has been described in previous publications. The low- pressure limit of this gage is extended by the addition of an electron multiplier ion detector. The ions generated in the magnetron gage are focused onto the first dynode of the multiplier by an electrostatic lens system. The multiplier gain is high enough to permit the use of an ion counting technique for the measurement of very low pressures. The lens system focuses the ions from the gage into a beam that passes through two limiting apertures before striking the first dynode. This arrangement prevents the soft x rays generated at the magnetron gage anode from falling directly on the first dynode. The x-ray photoemission from the first dynode is reduced by a factor of 50,000 below that observed for the magnetron gage alone. Measurements of sensitivity and x-ray photocurrent indicate that this device should have a linear response down to pressures of 2 x 10/sup -17/ Torr. Ability to measure low pressures with the gage appears to be limited by the inherent dark current of the multiplier. The gage may be used to advantage for the measurement of higher pressures by operating at reduced emission current levels to minimize pumping and degassing effects. Photoemission from the first dynode caused by incident light from the magnetron gage filament has been reduced to vaiues below the x-ray photoemission by the use of a lanthanum hexaboride cathode which operates at only 675 c- C. This low operating temperature reduces the problem of gas reactions with the filament. Radiation from gas molecules excited by electron impact in the magnetron gage was observed in the 10/sup -9/ Torr pressure range. (auth)

Research Organization:
General Electric Research Lab., Schenectady, N.Y.
Sponsoring Organization:
USDOE
NSA Number:
NSA-17-025574
OSTI ID:
4485173
Journal Information:
Rev. Sci. Instr., Vol. Vol: 34; Other Information: Orig. Receipt Date: 31-DEC-63
Country of Publication:
Country unknown/Code not available
Language:
English

Similar Records

Electron emission characteristics of sputtered lanthanum hexaboride
Journal Article · Wed May 01 00:00:00 EDT 1991 · Journal of Vacuum Science and Technology, A (Vacuum, Surfaces and Films); (USA) · OSTI ID:4485173

New deflected-beam gauge for pressures below 10$sup -12$ Torr
Journal Article · Mon Sep 01 00:00:00 EDT 1975 · J. Vac. Sci. Technol., v. 12, no. 5, pp. 1072-1078 · OSTI ID:4485173

Pressure Measurements Below 10-10 Torr with Bayard-Alpert and Magnetron Gauges
Journal Article · Tue Jan 01 00:00:00 EST 1963 · Review of Scientific Instruments · OSTI ID:4485173