skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: A mechanical resonance detector for multiwire proportional chambers

Journal Article · · Nuclear Instruments and Methods

The mechanical tension in the anode wires of a multiwire proportional chamber must be sufficient to prevent wire displacement due to electrostatic forces between wires but must not exceed the breaking strength of the wires. A technique is described for inferring the tension by measuring the mechanical resonart frequency of the wires. It employs a servomechanism which uses the small capacitance change of the selected wire to tune a variable frequency oscillator to the mechanical resonance. Since it is not necessary to observe the wires visually, the technique is applicable to closed chambers in inaccessible locations.

Research Organization:
Paris Univ., Orsay, France
Sponsoring Organization:
USDOE
NSA Number:
NSA-29-010182
OSTI ID:
4406695
Journal Information:
Nuclear Instruments and Methods, Vol. 113, Issue 3; Other Information: Orig. Receipt Date: 30-JUN-74; ISSN 0029-554X
Publisher:
Elsevier
Country of Publication:
Country unknown/Code not available
Language:
English