skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Precision metrology of NSTX surfaces using coherent laser radar ranging

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.1310586· OSTI ID:40204832

A frequency modulated Coherent Laser Radar ranging diagnostic is being used on the National Spherical Torus Experiment (NSTX) for precision metrology. The distance (range) between the 1.5 {mu}m laser source and the target is measured by the shift in frequency of the linearly modulated beam reflected off the target. The range can be measured to a precision of <100 {mu}m at distances of up to 22 m. A description is given of the geometry and procedure for measuring NSTX interior and exterior surfaces during open vessel conditions, and the results of measurements are elaborated.

Sponsoring Organization:
(US)
OSTI ID:
40204832
Journal Information:
Review of Scientific Instruments, Vol. 72, Issue 1; Other Information: Othernumber: RSINAK000072000001000533000001; 526101CON; PBD: Jan 2001; ISSN 0034-6748
Publisher:
The American Physical Society
Country of Publication:
United States
Language:
English