skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: A submillimeter polychromator for plasma spectroscopy

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.1147436· OSTI ID:389513
; ; ; ;  [1]
  1. Associazione EURATOM-ENEA sulla Fusione Centro Ricerche Frascati, CP 65, 00044-Frascati (Rome) (Italy)

A 12-channel grating polychromator for the measurement of electron cyclotron emission on the Frascati tokamak upgrade (FTU) has been built. Three gratings can be exchanged remotely, to allow the instrument to cover three different spectral ranges (between 200 and 600 GHz) which correspond to the second harmonic of electron cyclotron emission for tokamak magnetic fields of 4, 6, and 8 T. Three sets of low-pass filters, whose cut-off frequencies are optimized for each grating, are used to cut the high diffraction orders. In order to exchange the filters together with the grating it was necessary to integrate the groups of filters with the polychromator itself. This has been done reducing the grating sizes and assembling them all closely on a moveable structure. The exchange of gratings and filters can be performed remotely via computer in few seconds without any lost of the alignment. InSb cryogenic hot-electron bolometers are used as detectors. The cryostat hold-time is more than 30 days. Spectral resolution {lambda}/{delta}{lambda} is about 40 corresponding to a spatial resolution of {delta}{ital R}{approx_equal}2.3 cm. The minimum detectable temperature is about 10 eV at 20 kHz bandwidth. The spatial structure and temporal evolution of several kinds of fast temperature perturbations (as the ones due to magnetohydrodynamic activity, temperature fall during pellet injection and plasma disruption) have been measured on FTU by means of this diagnostic system. {copyright} {ital 1996 American Institute of Physics.}

OSTI ID:
389513
Journal Information:
Review of Scientific Instruments, Vol. 67, Issue 9; Other Information: PBD: Sep 1996
Country of Publication:
United States
Language:
English

Similar Records

Ten-channel grating polychromator for electron cyclotron emission plasma diagnostics
Journal Article · Thu Sep 01 00:00:00 EDT 1983 · Rev. Sci. Instrum.; (United States) · OSTI ID:389513

Twenty-channel grating polychromator diagnostic system for electron cyclotron emission measurement in JT-60
Journal Article · Mon Oct 01 00:00:00 EDT 1990 · Review of Scientific Instruments; (USA) · OSTI ID:389513

A 19-channel fast grating polychromator for ECE measurements in the MTX tokamak
Journal Article · Thu Nov 01 00:00:00 EST 1990 · Review of Scientific Instruments; (USA) · OSTI ID:389513