Title: X-ray crystal imagers for inertial confinement fusion experiments (invited)

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.1149352· OSTI ID:295626
;  [1]; ;  [2]; ;  [3]
  1. Science Applications International Corp., McLean , Virginia 22102 (United States)
  2. Plasma Physics Division, Naval Research Laboratory, Washington, DC 20375 (United States)
  3. Space Science Division, Naval Research Laboratory, Washington, DC 20375 (United States)

We report on our continued development of high resolution monochromatic x-ray imaging system based on spherically curved crystals. This system can be extensively used in the relevant experiments of the inertial confinement fusion (ICF) program. The system is currently used, but not limited to diagnostics of the targets ablatively accelerated by the Nike KrF laser. A spherically curved quartz crystal (2d=6.687thinsp03thinsp{Angstrom}, R=200thinspmm) has been used to produce monochromatic backlit images with the He-like Si resonance line (1865 eV) as the source of radiation. Another quartz crystal (2d=8.5099thinsp{Angstrom}, R=200thinspmm) with the H-like Mg resonance line (1473 eV) has been used for backlit imaging with higher contrast. The spatial resolution of the x-ray optical system is 1.7 {mu}m in selected places and 2{endash}3 {mu}m over a larger area. A second crystal with a separate backlighter was added to the imaging system. This makes it possible to make use of all four strips of the framing camera. Time resolved, 20{times} magnified, backlit monochromatic images of CH planar targets driven by the Nike facility have been obtained with spatial resolution of 2.5 {mu}m in selected places and 5 {mu}m over the focal spot of the Nike laser. We are exploring the enhancement of this technique to the higher and lower backlighter energies. {copyright} {ital 1999 American Institute of Physics.} thinsp

OSTI ID:
295626
Report Number(s):
CONF-980605-; ISSN 0034-6748; TRN: 99:001812
Journal Information:
Review of Scientific Instruments, Vol. 70, Issue 1; Conference: 12. topical conference on high-temperature plasma diagnostics, Princeton, NJ (United States), 7-11 Jun 1998; Other Information: PBD: Jan 1999
Country of Publication:
United States
Language:
English

Similar Records

High resolution monochromatic X-ray imaging system based on spherically bent crystals
Journal Article · Thu May 01 00:00:00 EDT 1997 · AIP Conference Proceedings · OSTI ID:295626

High-resolution monochromatic x-ray imaging system based on spherically bent crystals
Journal Article · Sat Aug 01 00:00:00 EDT 1998 · Applied Optics · OSTI ID:295626

High resolution monochromatic X-ray imaging system based on spherically bent crystals
Journal Article · Mon May 05 00:00:00 EDT 1997 · AIP Conference Proceedings · OSTI ID:295626