W-026, acceptance test report gamma energy assay (GEA) system A(submittal {number_sign}39.8) C3
This test procedure is designed to test all of the functions of the GEA system at the WRAP facility at Hanford, Washington. The procedure is designed to be executed in the sequence given, starting with: * verification of components and labeling * verification or entry of all of the setup parameters * performing the calibrations (energy, efficiency, transmission, and reference peak) * performing the measurements in local mode * performing the measurements in remote mode * confirming the message formats * utility operations (LN2 fill, archiving, password maintenance) The system is designed to be operated in a remote mode in which all commands are received over a network, and the results are returned over the same network. A manual mode is provided for the setup and calibration operations, and for maintenance functions. The system is designed to start up in remote mode. The user can go to local mode when at the keyboard by selecting Exit on the screen shown during remote mode. A user logon screen, requiring 2 username and password, limits access to local mode.
- Research Organization:
- Westinghouse Hanford Co., Richland, WA (United States)
- Sponsoring Organization:
- USDOE Office of Environmental Restoration and Waste Management, Washington, DC (United States)
- DOE Contract Number:
- AC06-96RL13200
- OSTI ID:
- 289825
- Report Number(s):
- HNF-SD-W026-ATR-024; ON: DE98059533; BR: 39EW31300; TRN: 97000027
- Resource Relation:
- Other Information: PBD: 28 Jan 1997
- Country of Publication:
- United States
- Language:
- English
Similar Records
Department 1824 Job Card System: A new web-based business tool
W-026 acceptance test report plant control system hardware (submittal {number_sign} 220.C)