Title: Burn Depth Estimation Using Thermal Excitation and Imaging

Conference ·
OSTI ID:2455

Accurate estimation of the depth of partial-thickness burns and the early prediction of a need for surgical intervention are difficult. A non-invasive technique utilizing the difference in thermal relaxation time between burned and normal skin may be useful in this regard. In practice, a thermal camera would record the skin's response to heating or cooling by a small amount-roughly 5{degrees} Celsius for a short duration. The thermal stimulus would be provided by a heat lamp, hot or cold air, or other means. Processing of the thermal transients would reveal areas that returned to equilibrium at different rates, which should correspond to different burn depths. In deeper thickness burns, the outside layer of skin is further removed from the constant-temperature region maintained through blood flow. Deeper thickness areas should thus return to equilibrium more slowly than other areas. Since the technique only records changes in the skin's temperature, it is not sensitive to room temperature, the burn's location, or the state of the patient. Preliminary results are presented for analysis of a simulated burn, formed by applying a patch of biosynthetic wound dressing on top of normal skin tissue.

Research Organization:
Sandia National Laboratories (SNL), Albuquerque, NM, and Livermore, CA (United States)
Sponsoring Organization:
USDOE
DOE Contract Number:
AC04-94AL85000
OSTI ID:
2455
Report Number(s):
SAND98-1454C; ON: DE00002455
Resource Relation:
Conference: Photonics West - Biomedical Optics '99; San Jose, CA; 01/24-29/1999
Country of Publication:
United States
Language:
English