Title: Signal-to-noise ratio in neuro activation PET studies

Abstract

It has become commonplace to compare scanner sensitivity characteristics by comparing noise equivalent count rate curves. However, because a 20-cm diameter uniform phantom is drastically difference from a human brain, these curves give misleading information when planning a neuro activation PET experiment. Signal-to-noise ratio (SNR) calculations have been performed using measured data (Siemens 921 scanner) from the three-dimensional (3-D) Hoffman brain phantom for the purpose of determining the optimal injection and scanning protocol for [{sup 15}O] labeled activation experiments. Region of interest (ROI) values along with the variance due to prompt (trues plus randoms) and random events were determined for various regions and radioactivity concentrations. Calculated attenuation correction was used throughout. Scatter correction was not used when calculating the SNR in activation studies because the number of scattered events is almost identical in each data acquisition and hence cancels. The results indicate that randoms correction should not be performed and that rather than being limited by the scanner capabilities, neuro activation experiments are limited by the amount of radioactivity that can be injected and the length of time the patient can stay in the scanner.

Authors:
 [1]
  1. Emory Center for Positron Emission Tomography, Atlanta, GA (United States). Div. of Nuclear Medicine
Publication Date:
OSTI Identifier:
227873
Resource Type:
Journal Article
Resource Relation:
Journal Name: IEEE Transactions on Medical Imaging; Journal Volume: 15; Journal Issue: 2; Other Information: PBD: Apr 1996
Country of Publication:
United States
Language:
English
Subject:
55 BIOLOGY AND MEDICINE, BASIC STUDIES; POSITRON COMPUTED TOMOGRAPHY; SIGNAL-TO-NOISE RATIO; BRAIN; PHANTOMS; THREE-DIMENSIONAL CALCULATIONS; IMAGE PROCESSING; OXYGEN 15; LABELLED COMPOUNDS

Citation Formats

Votaw, J.R. Signal-to-noise ratio in neuro activation PET studies. United States: N. p., 1996. Web. doi:10.1109/42.491421.
Votaw, J.R. Signal-to-noise ratio in neuro activation PET studies. United States. doi:10.1109/42.491421.
Votaw, J.R. 1996. "Signal-to-noise ratio in neuro activation PET studies". United States. doi:10.1109/42.491421.
@article{osti_227873,
title = {Signal-to-noise ratio in neuro activation PET studies},
author = {Votaw, J.R.},
abstractNote = {It has become commonplace to compare scanner sensitivity characteristics by comparing noise equivalent count rate curves. However, because a 20-cm diameter uniform phantom is drastically difference from a human brain, these curves give misleading information when planning a neuro activation PET experiment. Signal-to-noise ratio (SNR) calculations have been performed using measured data (Siemens 921 scanner) from the three-dimensional (3-D) Hoffman brain phantom for the purpose of determining the optimal injection and scanning protocol for [{sup 15}O] labeled activation experiments. Region of interest (ROI) values along with the variance due to prompt (trues plus randoms) and random events were determined for various regions and radioactivity concentrations. Calculated attenuation correction was used throughout. Scatter correction was not used when calculating the SNR in activation studies because the number of scattered events is almost identical in each data acquisition and hence cancels. The results indicate that randoms correction should not be performed and that rather than being limited by the scanner capabilities, neuro activation experiments are limited by the amount of radioactivity that can be injected and the length of time the patient can stay in the scanner.},
doi = {10.1109/42.491421},
journal = {IEEE Transactions on Medical Imaging},
number = 2,
volume = 15,
place = {United States},
year = 1996,
month = 4
}
  • PET activation studies employing the autoradiographic technique and {sup 15}O-water or {sup 15}O-butanol use the difference between images acquired under baseline conditions and during activation to detect focal changes in cerebral blood flow which occur upon stimulus presentation. Typically, the activating task or baseline conditions are maintained throughout the entire imaging period. Simulations of the kinetics of these freely diffusible tracers suggest there may be an advantage to switching between activation and baseline conditions during the course of the study which results in images which maximize the difference signal rather than seeking to quantitate blood flow. The authors examine themore » potential of these switched protocols to increase signal-to-noise (S/N) in PET activation studies. They examined S/N in activation studies using both standard and switched paradigms with a simple switched protocol and dynamic three-dimensional PET data from human subjects. With tracer kinetic simulations, they investigated the sensitivity of the S/N gain to factors such as the shape of the input function, the time at which the conditions are switched and the magnitude of the activation. In human studies of activation sites in the visual cortex, primary motor and premotor areas, S/N improvements of 20%-30% were detected using the switched paradigms. Simulations show that this gain is virtually independent of activation magnitude and that there is a broad time window of 20 sec for making the switch between conditions. Switched paradigms have the potential to significantly increase S/N in PET activation studies. In human studies, the S/N increase averaged 25% which is equivalent to increasing the number of counts collected by 50%. Switched paradigms can be used to maximize the difference signal in many activation studies, and do not preclude the absolute quantitation of blood flow using the standard autoradiographic-technique. 29 refs., 10 figs.« less
  • An attempt is made here to illustrate the relevance of several parameters in conventional flame ionization experiments. The difference between current measurement and charge measurement and the role of the stray circuit capacitance are stressed. Some optimized examples of the minimum number of atoms detectable are presented.
  • The multi-pass Thomson scattering (TS) scheme enables obtaining many photons by accumulating multiple TS signals. The signal-to-noise ratio (SNR) depends on the accumulation number. In this study, we performed multi-pass TS measurements for ohmically heated plasmas, and the relationship between SNR and the accumulation number was investigated. As a result, improvement of SNR in this experiment indicated similar tendency to that calculated for the background noise dominant situation.
  • Noise formula describing the 1/f noise in InSb films is pointed out. The noise-intensity coefficient K is introduced as a variable depending on film-preparation methods. Experimentally, K/sup 2/ changes by two orders of magnitude. However, existing noise calculations show that K/sup 2/ should remain constant except for a dimension factor. Therefore, the experimental results cannot be explained by the existing theory.