Title: Accuracy of a dose-area product compared to an absorbed dose to water at a point in a 2 cm diameter field

Abstract

Purpose: Graphite calorimeters with a core diameter larger than the beam can be used to establish dosimetric references in small fields. The dose-area product (DAP) measured can theoretically be linked to an absorbed dose at a point by the determination of a profile correction. This study aims at comparing the DAP-based protocol to the usual absorbed dose at a point protocol in a 2 cm diameter field for which both references exist. Methods: Two calorimeters were used, respectively, with a sensitive volume of 0.6 cm (for the absorbed dose at a point measurement) and 3 cm diameter (for the DAP measurement). Profile correction was calculated from a 2D dose mapping using three detectors: a PinPoint chamber, a synthetic diamond, and EBT3 films. A specific protocol to read EBT3 films was implemented and the dose-rate and energy dependences were studied to assure a precise measurement, especially in the penumbra and out-of-field regions. Results: EBT3 films were found independent on dose rates over the range studied but showed a strong under-response (18%) at low energies. Depending on the dosimeter used for calculating the profile correction, a deviation of 0.8% (PinPoint chamber), 0.9% (diamond), or 1.9% (EBT3 films) was observed between the calibrationmore » coefficient derived from DAP measurements and the one directly established in terms of absorbed dose to water at a point. Conclusions: The DAP method can currently be linked to the classical dosimetric reference system based in an absorbed dose at a point only with a confidence interval of 95% (k = 2). None of the detectors studied can be used to determine an absorbed dose to water at a point from a DAP measurement with an uncertainty smaller than 1.2%.« less

Authors:
; ; ;
Publication Date:
OSTI Identifier:
22689460
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 43; Journal Issue: 7; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; 61 RADIATION PROTECTION AND DOSIMETRY; ABSORBED RADIATION DOSES; ACCURACY; CORRECTIONS; DOSE RATES; ENERGY DEPENDENCE; WATER

Citation Formats

Dufreneix, S., Ostrowsky, A., Rapp, B., and Daures, J. Accuracy of a dose-area product compared to an absorbed dose to water at a point in a 2 cm diameter field. United States: N. p., 2016. Web. doi:10.1118/1.4953207.
Dufreneix, S., Ostrowsky, A., Rapp, B., & Daures, J. Accuracy of a dose-area product compared to an absorbed dose to water at a point in a 2 cm diameter field. United States. doi:10.1118/1.4953207.
Dufreneix, S., Ostrowsky, A., Rapp, B., and Daures, J. Fri . "Accuracy of a dose-area product compared to an absorbed dose to water at a point in a 2 cm diameter field". United States. doi:10.1118/1.4953207.
@article{osti_22689460,
title = {Accuracy of a dose-area product compared to an absorbed dose to water at a point in a 2 cm diameter field},
author = {Dufreneix, S. and Ostrowsky, A. and Rapp, B. and Daures, J.},
abstractNote = {Purpose: Graphite calorimeters with a core diameter larger than the beam can be used to establish dosimetric references in small fields. The dose-area product (DAP) measured can theoretically be linked to an absorbed dose at a point by the determination of a profile correction. This study aims at comparing the DAP-based protocol to the usual absorbed dose at a point protocol in a 2 cm diameter field for which both references exist. Methods: Two calorimeters were used, respectively, with a sensitive volume of 0.6 cm (for the absorbed dose at a point measurement) and 3 cm diameter (for the DAP measurement). Profile correction was calculated from a 2D dose mapping using three detectors: a PinPoint chamber, a synthetic diamond, and EBT3 films. A specific protocol to read EBT3 films was implemented and the dose-rate and energy dependences were studied to assure a precise measurement, especially in the penumbra and out-of-field regions. Results: EBT3 films were found independent on dose rates over the range studied but showed a strong under-response (18%) at low energies. Depending on the dosimeter used for calculating the profile correction, a deviation of 0.8% (PinPoint chamber), 0.9% (diamond), or 1.9% (EBT3 films) was observed between the calibration coefficient derived from DAP measurements and the one directly established in terms of absorbed dose to water at a point. Conclusions: The DAP method can currently be linked to the classical dosimetric reference system based in an absorbed dose at a point only with a confidence interval of 95% (k = 2). None of the detectors studied can be used to determine an absorbed dose to water at a point from a DAP measurement with an uncertainty smaller than 1.2%.},
doi = {10.1118/1.4953207},
journal = {Medical Physics},
number = 7,
volume = 43,
place = {United States},
year = {Fri Jul 15 00:00:00 EDT 2016},
month = {Fri Jul 15 00:00:00 EDT 2016}
}