Title: WE-AB-206-01: Diagnostic Ultrasound Imaging Quality Assurance

Abstract

The involvement of medical physicists in diagnostic ultrasound imaging service is increasing due to QC and accreditation requirements. The goal of this ultrasound hands-on workshop is to demonstrate quality control (QC) testing in diagnostic ultrasound and to provide updates in ACR ultrasound accreditation requirements. The first half of this workshop will include two presentations reviewing diagnostic ultrasound QA/QC and ACR ultrasound accreditation requirements. The second half of the workshop will include live demonstrations of basic QC tests. An array of ultrasound testing phantoms and ultrasound scanners will be available for attendees to learn diagnostic ultrasound QC in a hands-on environment with live demonstrations and on-site instructors. The targeted attendees are medical physicists in diagnostic imaging. Learning Objectives: Gain familiarity with common elements of a QA/QC program for diagnostic ultrasound imaging dentify QC tools available for testing diagnostic ultrasound systems and learn how to use these tools Learn ACR ultrasound accreditation requirements Jennifer Walter is an employee of American College of Radiology on Ultrasound Accreditation.

Authors:
 [1]
  1. University of Wisconsin (United States)
Publication Date:
OSTI Identifier:
22654128
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 43; Journal Issue: 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
61 RADIATION PROTECTION AND DOSIMETRY; BIOMEDICAL RADIOGRAPHY; EDUCATIONAL FACILITIES; LEARNING; QUALITY ASSURANCE; QUALITY CONTROL; TESTING

Citation Formats

Zagzebski, J. WE-AB-206-01: Diagnostic Ultrasound Imaging Quality Assurance. United States: N. p., 2016. Web. doi:10.1118/1.4957767.
Zagzebski, J. WE-AB-206-01: Diagnostic Ultrasound Imaging Quality Assurance. United States. doi:10.1118/1.4957767.
Zagzebski, J. 2016. "WE-AB-206-01: Diagnostic Ultrasound Imaging Quality Assurance". United States. doi:10.1118/1.4957767.
@article{osti_22654128,
title = {WE-AB-206-01: Diagnostic Ultrasound Imaging Quality Assurance},
author = {Zagzebski, J.},
abstractNote = {The involvement of medical physicists in diagnostic ultrasound imaging service is increasing due to QC and accreditation requirements. The goal of this ultrasound hands-on workshop is to demonstrate quality control (QC) testing in diagnostic ultrasound and to provide updates in ACR ultrasound accreditation requirements. The first half of this workshop will include two presentations reviewing diagnostic ultrasound QA/QC and ACR ultrasound accreditation requirements. The second half of the workshop will include live demonstrations of basic QC tests. An array of ultrasound testing phantoms and ultrasound scanners will be available for attendees to learn diagnostic ultrasound QC in a hands-on environment with live demonstrations and on-site instructors. The targeted attendees are medical physicists in diagnostic imaging. Learning Objectives: Gain familiarity with common elements of a QA/QC program for diagnostic ultrasound imaging dentify QC tools available for testing diagnostic ultrasound systems and learn how to use these tools Learn ACR ultrasound accreditation requirements Jennifer Walter is an employee of American College of Radiology on Ultrasound Accreditation.},
doi = {10.1118/1.4957767},
journal = {Medical Physics},
number = 6,
volume = 43,
place = {United States},
year = 2016,
month = 6
}
  • The goal of this ultrasound hands-on workshop is to demonstrate advancements in high intensity focused ultrasound (HIFU) and to demonstrate quality control (QC) testing in diagnostic ultrasound. HIFU is a therapeutic modality that uses ultrasound waves as carriers of energy. HIFU is used to focus a beam of ultrasound energy into a small volume at specific target locations within the body. The focused beam causes localized high temperatures and produces a well-defined regions of necrosis. This completely non-invasive technology has great potential for tumor ablation and targeted drug delivery. At the workshop, attendees will see configurations, applications, and hands-on demonstrationsmore » with on-site instructors at separate stations. The involvement of medical physicists in diagnostic ultrasound imaging service is increasing due to QC and accreditation requirements. At the workshop, an array of ultrasound testing phantoms and ultrasound scanners will be provided for attendees to learn diagnostic ultrasound QC in a hands-on environment with live demonstrations of the techniques. Target audience: Medical physicists and other medical professionals in diagnostic imaging and radiation oncology with interest in high-intensity focused ultrasound and in diagnostic ultrasound QC. Learning Objectives: Learn ultrasound physics and safety for HIFU applications through live demonstrations Get an overview of the state-of-the art in HIFU technologies and equipment Gain familiarity with common elements of a quality control program for diagnostic ultrasound imaging Identify QC tools available for testing diagnostic ultrasound systems and learn how to use these tools List of supporting vendors for HIFU and diagnostic ultrasound QC hands-on workshop: Philips Healthcare Alpinion Medical Systems Verasonics, Inc Zonare Medical Systems, Inc Computerized Imaging Reference Systems (CIRS), Inc. GAMMEX, Inc., Cablon Medical BV Steffen Sammet: NIH/NCI grant 5R25CA132822, NIH/NINDS grant 5R25NS080949 and Philips Healthcare research grant C32.« less
  • We describe quality assurance paradigms for ultrasound imaging systems for target localization (UISTL). To determine the absolute localization accuracy of a UISTL, an absolute coordinate system can be established in the treatment room and spherical targets at various depths can be localized. To test the ability of such a system to determine the magnitude of internal organ motion, a phantom that mimics the human male pelvic anatomy can be used to simulate different organ motion ranges. To assess the interuser variability of ultrasound (US) guidance, different experienced users can independently determine the daily organ shifts for the same patients formore » a number of consecutive fractions. The average accuracy for a UISTL for the localization of spherical targets at various depths has been found to be 0.57 {+-} 0.47 mm in each spatial dimension for various focal depths. For the phantom organ motion test it was found that the true organ motion could be determined to within 1.0 mm along each axis. The variability between different experienced users who localized the same 5 patients for five consecutive fractions was small in comparison to the indicated shifts. In addition to the quality assurance tests that address the ability of a UISTL to accurately localize a target, a thorough quality assurance program should also incorporate the following two aspects to ensure consistent and accurate localization in daily clinical use: (1) adequate training and performance monitoring of users of the US target localization system, and (2) prescreening of patients who may not be good candidates for US localization.« less
  • The requirements of quality assurance (QA) for both brachytherapy and imaging devices are well-defined, but image-guided brachytherapy has raised new issues. Image guidance in brachytherapy involves the transition from reference point dosimetry using films to volumetric imaging such as computed tomography, ultrasonography, and magnetic resonance imaging for treatment planning and guidance of applicator, needle, or seed placement. The QA of these devices might not reflect the conditions of use in brachytherapy or the requirements of brachytherapy treatment planning. Image interpretation becomes much more important with image-guided brachytherapy. The success of a procedure could depend on the interpretation of a singlemore » image in a calibration phase done under the time pressures of the operative setting. This change has implications at the level of treatment, the process, and the field of brachytherapy as a whole. The QA concerns arising from brachytherapy procedures using ultrasound, computed tomography, and magnetic resonance imaging guidance are discussed, as are the problems associated with using imaging in an interventional setting. This report was intended to indicate the QA concerns arising from the convergence of brachytherapy and imaging-highlighting areas in which technical improvements are needed.« less
  • While ultrasound guided prostate brachytherapy has gained wide acceptance as a primary treatment tool for prostate cancer, quality assurance of the ultrasound guidance system has received very little attention. Task Group 128 of the American Association of Physicists in Medicine was created to address quality assurance requirements specific to transrectal ultrasound used for guidance of prostate brachytherapy. Accurate imaging guidance and dosimetry calculation depend upon the quality and accuracy of the ultrasound image. Therefore, a robust quality assurance program for the ultrasound system is essential. A brief review of prostate brachytherapy and ultrasound physics is provided, followed by a recommendationmore » for elements to be included in a comprehensive test phantom. Specific test recommendations are presented, covering grayscale visibility, depth of penetration, axial and lateral resolution, distance measurement, area measurement, volume measurement, needle template/electronic grid alignment, and geometric consistency with the treatment planning computer.« less
  • Purpose: The purpose of this work is to develop and verify our quality assurance (QA) procedures to ensure the safety and efficacy of MR-guided focused ultrasound (MRgFUS) treatment of bone metastases. Methods: A practical QA program was developed. Monthly and daily QA (DQA) procedures were performed. The major QA items included the checks of the machine hardware, software and patient safety features. Briefly, these checks/tests include: 1) the cooling system reservoir and treatment table; 2) power to the treatment table; 3) the MR coil; 4) the transducer position with MRI; 5) image display on the treatment work station; 6) themore » effective focal spot in 3 directions using MR thermometry; and 7) all the safety devices including a sonication lamp, and the emergency stop-sonication switches. In order to avoid patient skin burn, it is important to remove gas bubbles in the interfaces between the treatment table and the gel pad, and the gel pad and patients skin during the patient setup. Our QA procedures have been verified and evaluated through patient treatments. Seven patients with scapula, humeral head, sacrum, ilium, pubic ramus and acetabular bone metastases were treated using MRgFUS. Results: Our study showed that all seven patients tolerated the MRgFUS treatment well. No skin toxicity or other complications were observed. The pain score (0–10) using the visual analog scale (VAS) was significantly reduced from 8.0 ± 1.1 before treatment to 4.7 ± 3.0, 3.0 ± 1.5, 3.2 ± 2.8 and 3.4 ± 1.5 at one day, one month, two months and three months after the MRgFUS treatment, respectively. Conclusion: We demonstrated that with the appropriate QA procedures, MRgFUS is a safe, effective and noninvasive treatment modality for palliation of bone metastases.« less