Title: WE-AB-BRA-09: Registration of Preoperative MRI to Intraoperative Radiographs for Automatic Vertebral Target Localization

Abstract

Purpose: Accurate localization of target vertebrae is essential to safe, effective spine surgery, but wrong-level surgery occurs with surprisingly high frequency. Recent research yielded the “LevelCheck” method for 3D-2D registration of preoperative CT to intraoperative radiographs, providing decision support for level localization. We report a new method (MR-LevelCheck) to perform 3D-2D registration based on preoperative MRI, presenting a solution for the increasingly common scenario in which MRI (not CT) is used for preoperative planning. Methods: Direct extension of LevelCheck is confounded by large mismatch in image intensity between MRI and radiographs. The proposed method overcomes such challenges with a simple vertebrae segmentation. Using seed points at centroids, vertebrae are segmented using continuous max-flow method and dilated by 1.8 mm to include surrounding cortical bone (inconspicuous in T2w-MRI). MRI projections are computed (analogous to DRR) using segmentation and registered to intraoperative radiographs. The method was tested in a retrospective IRB-approved study involving 11 patients undergoing cervical, thoracic, or lumbar spine surgery following preoperative MRI. Registration accuracy was evaluated in terms of projection-distance-error (PDE) between the true and estimated location of vertebrae in each radiograph. Results: The method successfully registered each preoperative MRI to intraoperative radiographs and maintained desirable properties of robustnessmore » against image content mismatch, and large capture range. Segmentation achieved Dice coefficient = 89.2 ± 2.3 and mean-absolute-distance (MAD) = 1.5 ± 0.3 mm. Registration demonstrated robust performance under realistic patient variations, with PDE = 4.0 ± 1.9 mm (median ± iqr) and converged with run-time = 23.3 ± 1.7 s. Conclusion: The MR-LevelCheck algorithm provides an important extension to a previously validated decision support tool in spine surgery by extending its utility to preoperative MRI. With initial studies demonstrating PDE <5 mm and 0% failure rate, the method is now in translation to larger scale prospective clinical studies. S. Vogt and G. Kleinszig are employees of Siemens Healthcare.« less

Authors:
; ; ; ;  [1]; ;  [2];  [3];
  1. Johns Hopkins University, Baltimore, MD (United States)
  2. Siemens Healthcare, Erlangen (Germany)
  3. The Johns Hopkins Hospital, Baltimore, MD (United States)
Publication Date:
OSTI Identifier:
22654099
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 43; Journal Issue: 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; 61 RADIATION PROTECTION AND DOSIMETRY; IMAGES; MATHEMATICAL SOLUTIONS; NMR IMAGING; SURGERY; VERTEBRAE

Citation Formats

De Silva, T, Uneri, A, Ketcha, M, Reaungamornrat, S, Goerres, J, Vogt, S, Kleinszig, G, Wolinsky, J, and Siewerdsen, JH. WE-AB-BRA-09: Registration of Preoperative MRI to Intraoperative Radiographs for Automatic Vertebral Target Localization. United States: N. p., 2016. Web. doi:10.1118/1.4957738.
De Silva, T, Uneri, A, Ketcha, M, Reaungamornrat, S, Goerres, J, Vogt, S, Kleinszig, G, Wolinsky, J, & Siewerdsen, JH. WE-AB-BRA-09: Registration of Preoperative MRI to Intraoperative Radiographs for Automatic Vertebral Target Localization. United States. doi:10.1118/1.4957738.
De Silva, T, Uneri, A, Ketcha, M, Reaungamornrat, S, Goerres, J, Vogt, S, Kleinszig, G, Wolinsky, J, and Siewerdsen, JH. Wed . "WE-AB-BRA-09: Registration of Preoperative MRI to Intraoperative Radiographs for Automatic Vertebral Target Localization". United States. doi:10.1118/1.4957738.
@article{osti_22654099,
title = {WE-AB-BRA-09: Registration of Preoperative MRI to Intraoperative Radiographs for Automatic Vertebral Target Localization},
author = {De Silva, T and Uneri, A and Ketcha, M and Reaungamornrat, S and Goerres, J and Vogt, S and Kleinszig, G and Wolinsky, J and Siewerdsen, JH},
abstractNote = {Purpose: Accurate localization of target vertebrae is essential to safe, effective spine surgery, but wrong-level surgery occurs with surprisingly high frequency. Recent research yielded the “LevelCheck” method for 3D-2D registration of preoperative CT to intraoperative radiographs, providing decision support for level localization. We report a new method (MR-LevelCheck) to perform 3D-2D registration based on preoperative MRI, presenting a solution for the increasingly common scenario in which MRI (not CT) is used for preoperative planning. Methods: Direct extension of LevelCheck is confounded by large mismatch in image intensity between MRI and radiographs. The proposed method overcomes such challenges with a simple vertebrae segmentation. Using seed points at centroids, vertebrae are segmented using continuous max-flow method and dilated by 1.8 mm to include surrounding cortical bone (inconspicuous in T2w-MRI). MRI projections are computed (analogous to DRR) using segmentation and registered to intraoperative radiographs. The method was tested in a retrospective IRB-approved study involving 11 patients undergoing cervical, thoracic, or lumbar spine surgery following preoperative MRI. Registration accuracy was evaluated in terms of projection-distance-error (PDE) between the true and estimated location of vertebrae in each radiograph. Results: The method successfully registered each preoperative MRI to intraoperative radiographs and maintained desirable properties of robustness against image content mismatch, and large capture range. Segmentation achieved Dice coefficient = 89.2 ± 2.3 and mean-absolute-distance (MAD) = 1.5 ± 0.3 mm. Registration demonstrated robust performance under realistic patient variations, with PDE = 4.0 ± 1.9 mm (median ± iqr) and converged with run-time = 23.3 ± 1.7 s. Conclusion: The MR-LevelCheck algorithm provides an important extension to a previously validated decision support tool in spine surgery by extending its utility to preoperative MRI. With initial studies demonstrating PDE <5 mm and 0% failure rate, the method is now in translation to larger scale prospective clinical studies. S. Vogt and G. Kleinszig are employees of Siemens Healthcare.},
doi = {10.1118/1.4957738},
journal = {Medical Physics},
number = 6,
volume = 43,
place = {United States},
year = {Wed Jun 15 00:00:00 EDT 2016},
month = {Wed Jun 15 00:00:00 EDT 2016}
}
  • Purpose: To develop a technique generating ultrafast on-board VC-MRI using prior 4D-MRI and on-board phase-skipped encoding k-space acquisition for real-time 3D target tracking of liver and lung radiotherapy. Methods: The end-of-expiration (EOE) volume in 4D-MRI acquired during the simulation was selected as the prior volume. 3 major respiratory deformation patterns were extracted through the principal component analysis of the deformation field maps (DFMs) generated between EOE and all other phases. The on-board VC-MRI at each instant was considered as a deformation of the prior volume, and the deformation was modeled as a linear combination of the extracted 3 major deformationmore » patterns. To solve the weighting coefficients of the 3 major patterns, a 2D slice was extracted from VC-MRI volume to match with the 2D on-board sampling data, which was generated by 8-fold phase skipped-encoding k-space acquisition (i.e., sample 1 phase-encoding line out of every 8 lines) to achieve an ultrafast 16–24 volumes/s frame rate. The method was evaluated using XCAT digital phantom to simulate lung cancer patients. The 3D volume of end-ofinhalation (EOI) phase at the treatment day was used as ground-truth onboard VC-MRI with simulated changes in 1) breathing amplitude and 2) breathing amplitude/phase change from the simulation day. A liver cancer patient case was evaluated for in-vivo feasibility demonstration. Results: The comparison between ground truth and estimated on-board VC-MRI shows good agreements. In XCAT study with changed breathing amplitude, the volume-percent-difference(VPD) between ground-truth and estimated tumor volumes at EOI was 6.28% and the Center-of-Mass-Shift(COMS) was 0.82mm; with changed breathing amplitude and phase, the VPD was 8.50% and the COMS was 0.54mm. The study of liver patient case also demonstrated a promising in vivo feasibility of the proposed method Conclusion: Preliminary results suggest the feasibility to estimate ultrafast VC-MRI for on-board target localization with phase skipped-encoding k-space acquisition. Research grant from NIH R01-184173.« less
  • Purpose: In image-guided spine surgery, mapping 3D preoperative images to 2D intraoperative images via 3D-2D registration can provide valuable assistance in target localization. However, the presence of surgical instrumentation, hardware implants, and soft-tissue resection/displacement causes mismatches in image content, confounding existing registration methods. Manual/semi-automatic methods to mask such extraneous content is time consuming, user-dependent, error prone, and disruptive to clinical workflow. We developed and evaluated 2 novel similarity metrics within a robust registration framework to overcome such challenges in target localization. Methods: An IRB-approved retrospective study in 19 spine surgery patients included 19 preoperative 3D CT images and 50 intraoperativemore » mobile radiographs in cervical, thoracic, and lumbar spine regions. A neuroradiologist provided truth definition of vertebral positions in CT and radiography. 3D-2D registration was performed using the CMA-ES optimizer with 4 gradient-based image similarity metrics: (1) gradient information (GI); (2) gradient correlation (GC); (3) a novel variant referred to as gradient orientation (GO); and (4) a second variant referred to as truncated gradient correlation (TGC). Registration accuracy was evaluated in terms of the projection distance error (PDE) of the vertebral levels. Results: Conventional similarity metrics were susceptible to gross registration error and failure modes associated with the presence of surgical instrumentation: for GI, the median PDE and interquartile range was 33.0±43.6 mm; similarly for GC, PDE = 23.0±92.6 mm respectively. The robust metrics GO and TGC, on the other hand, demonstrated major improvement in PDE (7.6 ±9.4 mm and 8.1± 18.1 mm, respectively) and elimination of gross failure modes. Conclusion: The proposed GO and TGC similarity measures improve registration accuracy and robustness to gross failure in the presence of strong image content mismatch. Such registration capability could offer valuable assistance in target localization without disruption of clinical workflow. G. Kleinszig and S. Vogt are employees of Siemens Healthcare.« less
  • Purpose: Online adaptive therapy (ART) relies on auto-contouring using deformable image registration (DIR). DIR’s inherent uncertainties require user intervention and manual edits while the patient is on the table. We investigated the dosimetric impact of DIR errors on the quality of re-optimized plans, and used the findings to establish regions for focusing manual edits to where DIR errors can Result in clinically relevant dose differences. Methods: Our clinical implementation of online adaptive MR-IGRT involves using DIR to transfer contours from CT to daily MR, followed by a physicians’ edits. The plan is then re-optimized to meet the organs at riskmore » (OARs) constraints. Re-optimized abdomen and pelvis plans generated based on physician edited OARs were selected as the baseline for evaluation. Plans were then re-optimized on auto-deformed contours with manual edits limited to pre-defined uniform rings (0 to 5cm) around the PTV. A 0cm ring indicates that the auto-deformed OARs were used without editing. The magnitude of the variations caused by the non-deterministic optimizer was quantified by repeat re-optimizations on the same geometry to determine the mean and standard deviation (STD). For each re-optimized plan, various volumetric parameters for the PTV, the OARs were extracted along with DVH and isodose evaluation. A plan was deemed acceptable if the variation from the baseline plan was within one STD. Results: Initial results show that for abdomen and pancreas cases, a minimum of 5cm margin around the PTV is required for contour corrections, while for pelvic and liver cases a 2–3 cm margin is sufficient. Conclusion: Focusing manual contour edits to regions of dosimetric relevance can reduce contouring time in the online ART process while maintaining a clinically comparable plan. Future work will further refine the contouring region by evaluating the path along the beams, dose gradients near the target and OAR dose metrics.« less
  • Purpose: To produce multi-modality compatible, realistic datasets for the joint evaluation of segmentation and registration with a reliable ground truth using a 4D biomechanical lung phantom. The further development of a computer controlled air flow system for recreation of real patient breathing patterns is incorporated for additional evaluation of motion prediction algorithms. Methods: A pair of preserved porcine lungs was pneumatically manipulated using an in-house computer controlled respirator. The respirator consisted of a set of bellows actuated by a 186 W computer controlled industrial motor. Patient breathing traces were recorded using a respiratory bellows belt during CT simulation and inputmore » into a control program incorporating a proportional-integral-derivative (PID) feedback controller in LabVIEW. Mock tumors were created using dual compartment vacuum sealed sea sponges. 65% iohexol,a gadolinium-based contrast agent and 18F-FDG were used to produce contrast and thus determine a segmentation ground truth. The intensity distributions of the compartments were then digitally matched for the final dataset. A bifurcation tracking pipeline provided a registration ground truth using the bronchi of the lung. The lungs were scanned using a GE Discovery-ST PET/CT scanner and a Phillips Panorama 0.23T MRI using a T1 weighted 3D fast field echo (FFE) protocol. Results: The standard deviation of the error between the patient breathing trace and the encoder feedback from the respirator was found to be ±4.2%. Bifurcation tracking error using CT (0.97×0.97×3.27 mm{sup 3} resolution) was found to be sub-voxel up to 7.8 cm displacement for human lungs and less than 1.32 voxel widths in any axis up to 2.3 cm for the porcine lungs. Conclusion: An MRI/PET/CT compatible anatomically and temporally realistic swine lung phantom was developed for the evaluation of simultaneous registration and segmentation algorithms. With the addition of custom software and mock tumors, the entire package offers ground truths for benchmarking performance with high fidelity.« less
  • Purpose: To determine the iterative deterministic solution stability of the Linear Boltzmann Transport Equation (LBTE) in the presence of magnetic fields. Methods: The LBTE with magnetic fields under investigation is derived using a discrete ordinates approach. The stability analysis is performed using analytical and numerical methods. Analytically, the spectral Fourier analysis is used to obtain the convergence rate of the source iteration procedures based on finding the largest eigenvalue of the iterative operator. This eigenvalue is a function of relevant physical parameters, such as magnetic field strength and material properties, and provides essential information about the domain of applicability requiredmore » for clinically optimal parameter selection and maximum speed of convergence. The analytical results are reinforced by numerical simulations performed using the same discrete ordinates method in angle, and a discontinuous finite element spatial approach. Results: The spectral radius for the source iteration technique of the time independent transport equation with isotropic and anisotropic scattering centers inside infinite 3D medium is equal to the ratio of differential and total cross sections. The result is confirmed numerically by solving LBTE and is in full agreement with previously published results. The addition of magnetic field reveals that the convergence becomes dependent on the strength of magnetic field, the energy group discretization, and the order of anisotropic expansion. Conclusion: The source iteration technique for solving the LBTE with magnetic fields with the discrete ordinates method leads to divergent solutions in the limiting cases of small energy discretizations and high magnetic field strengths. Future investigations into non-stationary Krylov subspace techniques as an iterative solver will be performed as this has been shown to produce greater stability than source iteration. Furthermore, a stability analysis of a discontinuous finite element space-angle approach (which has been shown to provide the greatest stability) will also be investigated. Dr. B Gino Fallone is a co-founder and CEO of MagnetTx Oncology Solutions (under discussions to license Alberta bi-planar linac MR for commercialization)« less