Title: WE-AB-BRB-01: Memorial Introduction; Storage Phosphor Panels for Radiation Therapy Dosimetry

Abstract

Despite widespread IMRT treatments at modern radiation therapy clinics, precise dosimetric commissioning of an IMRT system remains a challenge. In the most recent report from the Radiological Physics Center (RPC), nearly 20% of institutions failed an end-to-end test with an anthropomorphic head and neck phantom, a test that has rather lenient dose difference and distance-to-agreement criteria of 7% and 4 mm. The RPC report provides strong evidence that IMRT implementation is prone to error and that improved quality assurance tools are required. At the heart of radiation therapy dosimetry is the multidimensional dosimeter. However, due to the limited availability of water-equivalent dosimetry materials, research and development in this important field is challenging. In this session, we will review a few dosimeter developments that are either in the laboratory phase or in the pre-commercialization phase. 1) Radiochromic plastic. Novel formulations exhibit light absorbing optical contrast with very little scatter, enabling faster, broad beam optical CT design. 2) Storage phosphor. After irradiation, the dosimetry panels will be read out using a dedicated 2D scanning apparatus in a non-invasive, electro-optic manner and immediately restored for further use. 3) Liquid scintillator. Scintillators convert the energy from x-rays and proton beams into visible light, whichmore » can be recorded with a scientific camera (CCD or CMOS) from multiple angles. The 3D shape of the dose distribution can then be reconstructed. 4) Cherenkov emission imaging. Gated intensified imaging allows video-rate passive detection of Cherenkov emission during radiation therapy with the room lights on. Learning Objectives: To understand the physics of a variety of dosimetry techniques based upon optical imaging To investigate the strategies to overcome respective challenges and limitations To explore novel ideas of dosimeter design Supported in part by NIH Grants R01CA148853, R01CA182450, R01CA109558. Brian Pogue is founder and president of the company DoseOptics LLC, dedicated to developing and commercializing the first dedicated Cerenkov imaging camera and system for radiation dose imaging. Work reported in this talk does not involve the use of DoseOptics technology.; H. Li, this work was supported in part by NIH Grant No. R01CA148853; S. Beddar, NIH funding R01-CA182450.« less

Authors:
 [1]
  1. Washington University School of Medicine (United States)
Publication Date:
OSTI Identifier:
22654090
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 43; Journal Issue: 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; 61 RADIATION PROTECTION AND DOSIMETRY; BIOMEDICAL RADIOGRAPHY; DOSEMETERS; DOSIMETRY; PROTON BEAMS; QUALITY ASSURANCE; RADIATION DOSE DISTRIBUTIONS; RADIOTHERAPY; VISIBLE RADIATION

Citation Formats

Li, H. WE-AB-BRB-01: Memorial Introduction; Storage Phosphor Panels for Radiation Therapy Dosimetry. United States: N. p., 2016. Web. doi:10.1118/1.4957726.
Li, H. WE-AB-BRB-01: Memorial Introduction; Storage Phosphor Panels for Radiation Therapy Dosimetry. United States. doi:10.1118/1.4957726.
Li, H. 2016. "WE-AB-BRB-01: Memorial Introduction; Storage Phosphor Panels for Radiation Therapy Dosimetry". United States. doi:10.1118/1.4957726.
@article{osti_22654090,
title = {WE-AB-BRB-01: Memorial Introduction; Storage Phosphor Panels for Radiation Therapy Dosimetry},
author = {Li, H.},
abstractNote = {Despite widespread IMRT treatments at modern radiation therapy clinics, precise dosimetric commissioning of an IMRT system remains a challenge. In the most recent report from the Radiological Physics Center (RPC), nearly 20% of institutions failed an end-to-end test with an anthropomorphic head and neck phantom, a test that has rather lenient dose difference and distance-to-agreement criteria of 7% and 4 mm. The RPC report provides strong evidence that IMRT implementation is prone to error and that improved quality assurance tools are required. At the heart of radiation therapy dosimetry is the multidimensional dosimeter. However, due to the limited availability of water-equivalent dosimetry materials, research and development in this important field is challenging. In this session, we will review a few dosimeter developments that are either in the laboratory phase or in the pre-commercialization phase. 1) Radiochromic plastic. Novel formulations exhibit light absorbing optical contrast with very little scatter, enabling faster, broad beam optical CT design. 2) Storage phosphor. After irradiation, the dosimetry panels will be read out using a dedicated 2D scanning apparatus in a non-invasive, electro-optic manner and immediately restored for further use. 3) Liquid scintillator. Scintillators convert the energy from x-rays and proton beams into visible light, which can be recorded with a scientific camera (CCD or CMOS) from multiple angles. The 3D shape of the dose distribution can then be reconstructed. 4) Cherenkov emission imaging. Gated intensified imaging allows video-rate passive detection of Cherenkov emission during radiation therapy with the room lights on. Learning Objectives: To understand the physics of a variety of dosimetry techniques based upon optical imaging To investigate the strategies to overcome respective challenges and limitations To explore novel ideas of dosimeter design Supported in part by NIH Grants R01CA148853, R01CA182450, R01CA109558. Brian Pogue is founder and president of the company DoseOptics LLC, dedicated to developing and commercializing the first dedicated Cerenkov imaging camera and system for radiation dose imaging. Work reported in this talk does not involve the use of DoseOptics technology.; H. Li, this work was supported in part by NIH Grant No. R01CA148853; S. Beddar, NIH funding R01-CA182450.},
doi = {10.1118/1.4957726},
journal = {Medical Physics},
number = 6,
volume = 43,
place = {United States},
year = 2016,
month = 6
}
  • Purpose: An important property of a reusable dosimeter is its radiation hardness, that is, its ability to retain its dosimetric merits after irradiation. The radiation hardness of europium doped potassium chloride (KCl:Eu{sup 2+}), a storage phosphor material recently proposed for radiation therapy dosimetry, is examined in this study. Methods: Pellet-style KCl:Eu{sup 2+} dosimeters, 6 mm in diameter, and 1 mm thick, were fabricated in-house for this study. The pellets were exposed by a 6 MV photon beam or in a high dose rate {sup 137}Cs irradiator. Macroscopic properties, such as radiation sensitivity, dose response linearity, and signal stability, were studiedmore » with a laboratory photostimulated luminescence (PSL) readout system. Since phosphor performance is related to the state of the storage centers and the activator, Eu{sup 2+}, in the host lattice, spectroscopic and temporal measurements were carried out in order to explore radiation-induced changes at the microscopic level. Results: KCl:Eu{sup 2+} dosimeters retained approximately 90% of their initial signal strength after a 5000 Gy dose history. Dose response was initially supralinear over the dose range of 100-700 cGy but became linear after 60 Gy. Linearity did not change significantly in the 0-5000 Gy dose history spanned in this study. Annealing high dose history chips resulted in a return of supralinearity and a recovery of sensitivity. There were no significant changes in the PSL stimulation spectra, PSL emission spectra, photoluminescence spectra, or luminescence lifetime, indicating that the PSL signal process remains intact after irradiation but at a reduced efficiency due to reparable radiation-induced perturbations in the crystal lattice. Conclusions: Systematic studies of KCl:Eu{sup 2+} material are important for understanding how the material can be optimized for radiation therapy dosimetry purposes. The data presented here indicate that KCl:Eu{sup 2+} exhibits strong radiation hardness and lends support for further investigations of this novel material.« less
  • Purpose: We previously demonstrated that intensity-modulated radiation therapy (IMRT) significantly improves radiation dose distribution over three-dimensional planning for nasopharynx cancer and reported positive early clinical results. We now evaluate whether IMRT has resulted in improved outcomes for a larger cohort of patients with longer follow-up. Methods and Materials: Since 1998, all 74 patients with newly diagnosed, nonmetastatic nasopharynx cancer were treated with IMRT using accelerated fractionation to 70 Gy; 59 received a hyperfractionated concomitant boost, and more recently 15 received once-daily treatment with dose painting. With the exception of Stage I disease (n = 5) and patient preference (n =more » 1), 69 patients received concurrent and adjuvant platinum-based chemotherapy similar to that in the Intergroup 0099 trial. Results: Patient characteristics: median age 45; 32% Asian; 72% male; 65% World Health Organization III; 6% Stage I, 16% Stage II, 30% Stage III, 47% Stage IV. Median follow-up is 35 months. The 3-year actuarial rate of local control is 91%, and regional control is 93%; freedom from distant metastases, progression-free survival, and overall survival at 3 years are 78%, 67%, and 83%, respectively. There was 100% local control for Stage T1/T2 disease, compared to 83% for T3/T4 disease (p = 0.01). Six patients failed at the primary site, with median time to local tumor progression 16 months; 5 were exclusively within the 70 Gy volume, and 1 was both within and outside the target volume. There is a trend for improved local control with IMRT when compared to local control of 79% for 35 patients treated before 1998 with three-dimensional planning and chemotherapy (p 0.11). Six months posttherapy, 21%, 13%, 15%, and 0% of patients with follow-up audiograms (n = 24 patients) had Grade 1, 2, 3, and 4 sensorineural hearing loss, respectively. For patients with >1 year follow-up (n = 59), rates of long-term xerostomia were as follows: 26% none, 42% Grade 1, 32% Grade 2, and zero Grade 3. Conclusions: The pattern of primary site failure within the target volume suggests locally advanced T stage disease may require a higher biologic dose to gross tumor. Rates of severe (Grade 3-4) ototoxicity and xerostomia are low with IMRT as a result of normal-tissue protection. Distant metastases are now the dominant form of failure, emphasizing the need for improved systemic therapy.« less
  • Purpose: To review the Memorial Sloan-Kettering Cancer Center's experience in using intensity-modulated radiation therapy (IMRT) for the treatment of oropharyngeal cancer. Methods and Materials: Between September 1998 and June 2004, 50 patients with histologically confirmed cancer of the oropharynx underwent IMRT at our institution. There were 40 men and 10 women with a median age of 56 years (range, 28-78 years). The disease was Stage I in 1 patient (2%), Stage II in 3 patients (6%), Stage III in 7 (14%), and Stage IV in 39 (78%). Forty-eight patients (96%) received definitive treatment, and 2 (4%) were treated in themore » postoperative adjuvant setting. Concurrent chemotherapy was used in 43 patients (86%). Patients were treated using three different IMRT approaches: 76% dose painting, 18% concomitant boost with IMRT in both am and pm deliveries, and 6% concomitant boost with IMRT only in pm delivery. Regardless of the approach, the average prescription dose to the gross tumor planning target volume was 70 Gy, while the average dose delivered to the subclinical volume was 59.4 Gy in the dose painting group and 54 Gy in the concomitant boost group. Percutaneous endoscopic gastrostomy feeding tubes (PEGs) were placed before the beginning of treatment in 84% of the patients. Acute and late toxicity were graded according to the Radiation Therapy Oncology Group (RTOG) radiation morbidity scoring criteria. Toxicity was also evaluated using subjective criteria such as the presence of esophageal stricture, and the need for PEG usage. The local progression-free, regional progression-free, and distant metastases-free rates, and overall survival were calculated using the Kaplan-Meier method. Results: Three patients had persistent locoregional disease after treatment. The 2-year estimates of local progression-free, regional progression-free, distant metastases-free, and overall survival were 98%, 88%, 84%, and 98%, respectively. The worst acute mucositis experienced was Grade 1 in 4 patients (8%), Grade 2 in 27 (54%), and Grade 3 in 19 (38%). Xerostomia decreased with increasing time interval from the end of radiotherapy, and among the patients with at least 9 months of follow-up there was 67% Grade 0-1 and 33% Grade 2 toxicity. Of the 42 patients who required upfront PEG placement, 6 were still using PEG for nutrition at the time of this analysis. Three patients had cervical esophageal strictures, and of these, 1 was still PEG dependent 1 year after treatment. Two of these patients were treated with the IMRT concomitant boost am and pm approach, whereas the other was treated with the dose painting technique. Conclusions: Intensity-modulated radiotherapy achieved encouraging local control rates in patients with oropharyngeal carcinoma. Treatment toxicity was acceptable even in the setting of concurrent chemotherapy. Long-term follow-up is needed to confirm these preliminary findings.« less
  • Purpose: To evaluate the outcomes of patients with Ewing's sarcoma family of tumors (ESFT) treated with modern radiotherapy techniques with MRI along with optimal chemotherapy. Methods and Materials: The records of all 60 patients with ESFT who received radiation to the primary site between 1990 and 2004 were reviewed. All patients received chemotherapy, including vincristine, doxorubicin, cyclophosphamide, ifosfamide, and etoposide. Radiation was used as the sole modality for local control in 31 patients and was given either before (n = 3) or after surgical resection (n = 26) in the remainder. All patients had MRI and CT scan-based treatment planning,more » and 43% received intensity-modulated radiation therapy. Radiation doses ranged from 30 Gy to 60 Gy (median, 51 Gy), and 35% received hyperfractionated radiotherapy. Results: Median age was 16 years (range, 2-40 years). Because of selection bias for radiotherapy, the majority of primary tumors were centrally located (72%): spine (n = 18), pelvis (n = 15), extremities (n 12), chest wall (n = 5), head and neck (n = 5), and other (n = 5). Thirty-eight percent of patients presented with metastatic disease, and 52% of primary tumors were {>=}8 cm. Actuarial 3-year local control was 77%. The presence of metastases at diagnosis was an adverse prognostic factor for local control (84% vs. 61%, p = 0.036). No other predictive factors for local failure were identified. In patients without metastatic disease, 3-year disease-free and overall survival rates were 70% and 86%, respectively, whereas in patients with metastases they were both 21%. Follow-up of surviving patients was 6-178 months (median, 41 months). Conclusion: In this unfavorable cohort of ESFT patients, radiation therapy was an effective modality for local control, especially for patients without metastases. The presence of metastases at diagnosis is a predictive factor not only for death but also for local failure.« less
  • Portal verification images were generated from the photon contamination in electron beams produced by a linear accelerator during treatment of patients receiving high-energy electron radiation therapy (8-14 MeV). An experimental storage phosphor system was used to record the images and display them on laser-printed film. Images were obtained from four or more treatment fractions from 21 cases of head and neck cancer. Precision in field placement was estimated by determining the position of a selected anatomic landmark relative to the center of the field for each series of images. The average standard deviation in the field-position measurements was 3.8 mm.more » Several procedural problems were also detected and corrected after review of the verification images. The results indicate that the emphasis placed on monitoring and control of field-positioning error in high-energy electron treatments should be similar to the emphasis placed on this aspect of error in photon treatment.« less