Title: TU-H-BRA-03: Performance of a Clinical Gridded Electron Gun in Magnetic Fields: Implications for MRI-Linac Therapy

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4957625· OSTI ID:22654026
;  [1];  [2];  [3];  [4];  [5];  [6]
  1. University of Sydney, Sydney (Australia)
  2. University of Victoria, VCH040, Victoria, BC (Australia)
  3. Illawarra Hospital, Wollongong, NSW (Australia)
  4. Varian Medical Systems, Palo Alto, California (United States)
  5. Liverpool Hospital and Ingham Institute, Liverpool, NSW (United Kingdom)
  6. Siemens Healthcare GmbH, Forchheim (Germany)

Purpose: Recent advances towards MRI Linac radiotherapy have motivated a wide range of studies characterizing electromagnetic interactions between the two devices. One of the most sensitive components is the linac electron gun. To data, only non gridded (diode) guns have been investigated however, most linac vendors utilize gridded (triode) guns, which enable efficient and robust beam gating. The purpose of this study was to develop a realistic model of a gridded gun used clinically, and to characterize its performance in magnetic fields. Methods: The gridded electron gun used on Varian high energy machines was measured using 3D laser scanning quoted as accurate to 0.1mm. Based on the scane, a detailed CAD mode was developed. From this, key geometry was extracted and a FEM model was developed (Opera/SCALA). Next, the high voltage (HV), grid voltage, and emission current were read from six dose matched TrueBeam linacs for the 6X, 10X and 15X photon modes (0 B-field). The mean values were used to represent each mode, which was simulated I constant magnetic fields from 0–200G in-line, and 0–35G perpendicular. Results: Experimentally measured HV, grid voltage, and emission current from 6X, 10X and 15X modes were respectively: 15±.03kV, 10±.08kV, 11±.03kV; 93±7V, 41±3V, and 70±6V; 327±27mA, 129±10mA, and 214±19mA. The error in simulated emission current of each mode was 3%,6%, and 3%. For in-line fields, 50% beam loss occurred at 114, 96, and 97G; for perpendicular; at 12, 13 and 14G. Sensitivity for a given geometry is primarily determined by HV setting. Conclusion: Future MRI-Linac systems will almost certainly use gridded guns. We present the first model of a clinical gridded gun, and match the experimental emission current to within 6% across three different operating modes. This clinical gun shows increased sensitivity to magnetic fields than previous work,and different modes show different sensitivity.

OSTI ID:
22654026
Journal Information:
Medical Physics, Vol. 43, Issue 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English