Title: MO-FG-CAMPUS-IeP1-03: Establishment of Provincial Diagnostic Reference Levels in Pediatric Imaging

Abstract

Purpose: To establish provincial diagnostic reference levels (DRLs) in pediatric general radiography and computed tomography (CT) as a tool for the optimization of exposure parameters. Methods: Patient dose survey was conducted in the only pediatric hospital in the province of Nova Scotia. The DRLs were established as the 75th percentile of patient dose distributions in different age groups. For routine radiography projections the DRLs were determined in terms of entrance surface dose (ESD) calculated from the radiation output measurements and the tube current-exposure time product (mAs) recorded for each examination. Patient thickness was measured by the technologist during the examination. The CR and DR systems, employing respectively a fixed technique and phototiming, were evaluated separately; a two-tailed Student’s t-test was used to determine the significance of differences between the means of dose distributions. The CT studies included routine head, chest, abdomen/pelvis, and chest/abdomen/pelvis. The volume CT dose index (CTDIvol) and dose-length product (DLP) values were extracted retrospectively from PACS. The correction factors based on the effective diameter of the patient were applied to the CT dosimetry metrics based on the standard phantoms. Results: The provincial DRLs were established in the following age groups: newborn, 1, 5, 10, and 15 yearmore » olds. In general radiography the DR systems demonstrated slightly lower dose than the CR for all views, however the differences were not statistically significant (p > 0.05) for all examinations. In CT the provincial DRLs were lower than the published data, except for head DLPs in all age categories. This might be due to the small patient sample size in the survey. Future work will include additional CT data collection over an extended period of time. Conclusion: Provincial DRLs were established in the dedicated children’s hospital to provide guidance for the other facilities in examinations of pediatric patients.« less

Authors:
 [1];  [2];  [1];  [2]
  1. Dalhousie University (Canada)
  2. (Canada)
Publication Date:
OSTI Identifier:
22653885
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 43; Journal Issue: 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; 61 RADIATION PROTECTION AND DOSIMETRY; BIOMEDICAL RADIOGRAPHY; COMPUTERIZED TOMOGRAPHY; HEAD; PATIENTS; PEDIATRICS; PERTURBED ANGULAR CORRELATION; RADIATION DOSE DISTRIBUTIONS

Citation Formats

Tonkopi, E, Queen Elizabeth II Health Sciences Ctr, O’Brien, K, and IWK Health Centre, Halifax, NS. MO-FG-CAMPUS-IeP1-03: Establishment of Provincial Diagnostic Reference Levels in Pediatric Imaging. United States: N. p., 2016. Web. doi:10.1118/1.4957335.
Tonkopi, E, Queen Elizabeth II Health Sciences Ctr, O’Brien, K, & IWK Health Centre, Halifax, NS. MO-FG-CAMPUS-IeP1-03: Establishment of Provincial Diagnostic Reference Levels in Pediatric Imaging. United States. doi:10.1118/1.4957335.
Tonkopi, E, Queen Elizabeth II Health Sciences Ctr, O’Brien, K, and IWK Health Centre, Halifax, NS. 2016. "MO-FG-CAMPUS-IeP1-03: Establishment of Provincial Diagnostic Reference Levels in Pediatric Imaging". United States. doi:10.1118/1.4957335.
@article{osti_22653885,
title = {MO-FG-CAMPUS-IeP1-03: Establishment of Provincial Diagnostic Reference Levels in Pediatric Imaging},
author = {Tonkopi, E and Queen Elizabeth II Health Sciences Ctr and O’Brien, K and IWK Health Centre, Halifax, NS},
abstractNote = {Purpose: To establish provincial diagnostic reference levels (DRLs) in pediatric general radiography and computed tomography (CT) as a tool for the optimization of exposure parameters. Methods: Patient dose survey was conducted in the only pediatric hospital in the province of Nova Scotia. The DRLs were established as the 75th percentile of patient dose distributions in different age groups. For routine radiography projections the DRLs were determined in terms of entrance surface dose (ESD) calculated from the radiation output measurements and the tube current-exposure time product (mAs) recorded for each examination. Patient thickness was measured by the technologist during the examination. The CR and DR systems, employing respectively a fixed technique and phototiming, were evaluated separately; a two-tailed Student’s t-test was used to determine the significance of differences between the means of dose distributions. The CT studies included routine head, chest, abdomen/pelvis, and chest/abdomen/pelvis. The volume CT dose index (CTDIvol) and dose-length product (DLP) values were extracted retrospectively from PACS. The correction factors based on the effective diameter of the patient were applied to the CT dosimetry metrics based on the standard phantoms. Results: The provincial DRLs were established in the following age groups: newborn, 1, 5, 10, and 15 year olds. In general radiography the DR systems demonstrated slightly lower dose than the CR for all views, however the differences were not statistically significant (p > 0.05) for all examinations. In CT the provincial DRLs were lower than the published data, except for head DLPs in all age categories. This might be due to the small patient sample size in the survey. Future work will include additional CT data collection over an extended period of time. Conclusion: Provincial DRLs were established in the dedicated children’s hospital to provide guidance for the other facilities in examinations of pediatric patients.},
doi = {10.1118/1.4957335},
journal = {Medical Physics},
number = 6,
volume = 43,
place = {United States},
year = 2016,
month = 6
}
  • Purpose: Interventional cardiac procedures utilize frequent fluoroscopy and cineangiography, which impose considerable radiation risk to patients, especially pediatric patients. Accurate calculation of effective dose is important in order to estimate cancer risk over the rest of their lifetime. This study evaluates the difference in effective dose calculated by Monte Carlo simulation with those estimated by locally-derived conversion factors (CF-local) and by commonly quoted conversion factors from Karambatsakidou et al (CF-K). Methods: Effective dose (E),of 12 pediatric patients, age between 2.5–19 years old, who had undergone interventional cardiac procedures, were calculated using PCXMC-2.0 software. Tube spectrum, irradiation geometry, exposure parameters andmore » dose-area product (DAP) of each projection were included in the software calculation. Effective doses for each patient were also estimated by two Methods: 1) CF-local: conversion factor derived locally by generalizing results of 12 patients, multiplied by DAP of each patient gives E-local. 2) CF-K: selected factor from above-mentioned literature, multiplied by DAP of each patient gives E-K. Results: Mean of E, E-local and E-K were 16.01 mSv, 16.80 mSv and 22.25 mSv respectively. A deviation of −29.35% to +34.85% between E and E-local, while a greater deviation of −28.96% to +60.86% between E and EK were observed. E-K overestimated the effective dose for patients at age 7.5–19. Conclusion: Effective dose obtained by conversion factors is simple and quick to estimate radiation risk of pediatric patients. This study showed that estimation by CF-local may bear an error of 35% when compared with Monte Carlo calculation. If using conversion factors derived by other studies may result in an even greater error, of up to 60%, due to factors that are not catered for in the estimation, including patient size, projection angles, exposure parameters, tube filtration, etc. Users must be aware of these potential inaccuracies when simple conversion method is employed.« less
  • Purpose: To develop new ionization chamber dosimetry of absorbed dose to water in diagnostic kV x-ray beams, by using a beam quality conversion factor, kQ, for Co-60 to kV x-ray and an ionization conversion factor for a water-substitute plastic phantom. Methods: kQ was calculated for aluminum half value-layers (Al-HVLs) of 1.5 mm to 8 mm which were generated by kV x-ray beams of 50 to 120 kVp. Twenty-two energy spectra for ten effective energies (Eeff) were calculated by a SpecCalc program. Depth doses in water were calculated at 5 × 5 to 30 × 30 cm{sup 2} fields. Output factorsmore » were also obtained from the dose ratio for a 10 × 10 cm{sup 2} field. kQ was obtained for a PTW30013 Former ion chamber. In addition, an ionization conversion factor of the PWDT phantom to water was calculated. All calculations were performed with EGSnrc/cavity code and egs-chamber codes. Results: The x-ray beam energies for 1.5 mm to 8 mm Al-HVLs ranged in Eeff of 25.7 to 54.3 keV. kQ for 1.5 mm to 8 mm Al-HVLs were 0.831 to 0.897, at 1 and 2 cm depths for a 10 × 10 cm2 field. Similarly, output factors for 5 × 5 to 30 × 30 cm{sup 2} fields were 0.937 to 1.033 for 25.7 keV and 0.857 to 1.168 for 54.3 keV. The depth dose in a PWDT phantom decreased up to 5% compared to that in water at depth of ten percent of maximum dose for 1.5 mm Al-HVL. The ionization ratios of water/PWDT phantoms for the PTW30013 chamber were 1.012 to 1.007 for 1.5 mm to 8 mm Al-HVLs at 1 cm depth. Conclusion: It became possible to directly measure the absorbed dose to water with the ionization chamber in diagnostic kV x-ray beams, by using kQ and the PWDT phantom.« less
  • Purpose: To present the results of a patient dose evaluation program in pediatric cardiology and propose local diagnostic reference levels (DRLs) for different types of procedure and age range, in addition to suggesting approaches to correlate patient dose values with patient weight. This study was the first conducted in Latin America for pediatric interventional cardiology under the auspices of the International Atomic Energy Agency. Methods: Over three years, the following data regarding demographic and patient dose values were collected: age, gender, weight, height, number of cine series, total number of cine frames, fluoroscopy time (FT), and two dosimetric quantities, dose-areamore » product (DAP) and cumulative dose (CD), at the patient entrance reference point. The third quartile values for FT, DAP, CD, number of cine series, and the DAP/body weight ratio were proposed as the set of quantities to use as local DRLs. Results: Five hundred and seventeen patients were divided into four age groups. Sample sizes by age group were 120 for <1 yr; 213 for 1 to <5 yr; 82 for 5 to <10 yr; and 102 for 10 to <16 yr. The third quartile values obtained for DAP by diagnostic and therapeutic procedures and age range were 1.17 and 1.11 Gy cm{sup 2} for <1 yr; 1.74 and 1.90 Gy cm{sup 2} for 1 to <5 yr; 2.83 and 3.22 Gy cm{sup 2} for 5 to <10 yr; and 7.34 and 8.68 Gy cm{sup 2} for 10 to <16 yr, respectively. The third quartile value obtained for the DAP/body weight ratio for the full sample of procedures was 0.17 (Gy cm{sup 2}/kg) for diagnostic and therapeutic procedures. Conclusions: The data presented in this paper are an initial attempt at establishing local DRLs in pediatric interventional cardiology, from a large sample of procedures for the standard age bands used in Europe, complemented with the values of the ratio between DAP and patient weight. This permits a rough estimate of DRLs for different patient weights and the refining of these values for the age bands when there may be large differences in child size. These DRLs were obtained at the largest pediatric hospital in Chile, with an active optimization program, and could be used by other hospitals in the Latin America region to compare their current patient dose values and determine whether corrective action is appropriate.« less
  • Purpose: Variance reduction techniques (VRTs) are employed in Monte Carlo simulations to obtain estimates with reduced statistical uncertainty for a given simulation time. In this work, we study the bias and efficiency of a VRT for estimating the response of imaging detectors. Methods: We implemented Directed Sampling (DS), preferentially directing a fraction of emitted optical photons directly towards the detector by altering the isotropic model. The weight of each optical photon is appropriately modified to maintain simulation estimates unbiased. We use a Monte Carlo tool called fastDETECT2 (part of the hybridMANTIS open-source package) for optical transport, modified for VRT. Themore » weight of each photon is calculated as the ratio of original probability (no VRT) and the new probability for a particular direction. For our analysis of bias and efficiency, we use pulse height spectra, point response functions, and Swank factors. We obtain results for a variety of cases including analog (no VRT, isotropic distribution), and DS with 0.2 and 0.8 optical photons directed towards the sensor plane. We used 10,000, 25-keV primaries. Results: The Swank factor for all cases in our simplified model converged fast (within the first 100 primaries) to a stable value of 0.9. The root mean square error per pixel for DS VRT for the point response function between analog and VRT cases was approximately 5e-4. Conclusion: Our preliminary results suggest that DS VRT does not affect the estimate of the mean for the Swank factor. Our findings indicate that it may be possible to design VRTs for imaging detector simulations to increase computational efficiency without introducing bias.« less
  • Purpose: The national diagnostic reference levels (NDRLs) is an efficient, concise and powerful standard for optimizing radiation protection of a patient. However, for each hospital the dose-reducing potential of focusing on establishment of local DRLs (LDRLs). A lot of study reported that Computed tomography exam contributed majority radiation dose in different medical modalities, therefore, routine abdomen CT exam was choose in initial pilot study in our study. Besides the mAs of routine abdomen CT exam was decided automatic exposure control by linear attenuation is relate to body shape of patient. In this study we would like to establish the localmore » diagnostic reference levels of routine abdomen exam in computed tomography according to body weight of patient. Methods and Materials: There are two clinical CT scanners (a Toshiba Aquilion and a Siemens Sensation) were performed in this study. For CT examinations the basic recommended dosimetric quantity is the Computed Tomography Dose Index (CTDI). The patient sample involved 82 adult patients of both sexes and divided into three groups by their body weight (50–60 kg, 60–70 kg, 70–80 kg).Carried out the routine abdomen examinations, and all exposure parameters have been collected and the corresponding CTDIv and DLP values have been determined. The average values were compared with the European DRLs. Results: The majority of patients (75%) were between 50–70 Kg of body weight, the numbers of patient in each group of weight were 40–50:7; 50–60:29; 60–70:33; 70–80:13. The LDRLs in each group were 10.81mGy, 14.46mGy, 20.27mGy and 21.04mGy, respectively. The DLP were 477mGy, 630mGy, 887mGy and 959mGy, respectively. No matter which group the LDRLs were lower than European DRLs. Conclusions: We would like to state that this was a pioneer work in local hospital in Chiayi. We hope that this may lead the way to further developments in Taiwan.« less