Title: SU-G-IeP2-09: Iodine Imaging at Spectral CT with a Dual-Layer Detector

Abstract

Purpose: To evaluate the attenuation response of iodine and the accuracy of iodine quantification on a detector-based spectral CT scanner. Methods: A Gammex 461A phantom was scanned using a dual-layer detector (IQon, Philips) at 120 kVp using helical acquisition with a CDTIvol of 15 mGy to approximate the hospital’s clinical body protocol. No modifications to the standard protocol were necessary to enable spectral imaging. Iodine inserts at 6 concentrations (2, 5, 7.5, 10, 15, 20 mg/ml) were scanned individually at the center of the phantom and the 20 mg/ml insert was additionally scanned at the 3, 6, and 12 o’clock positions. Scans were repeated 10 times. Conventional, virtual monoenergetic (40–200 keV) and iodine-no-water images (with pixel values equal to iodine concentration of corresponding tissue) were reconstructed from acquired data. A circular ROI (diameter=30 pixels) was used in each conventional and monoenergetic image to measure the mean and standard deviation of the CT number in HU and in each iodine-no-water image to measure iodine concentration in mg/ml. Results: Mean CT number and contrast-to-noise ratio (CNR) measured from monoenergetic images increased with decreasing keV for all iodine concentrations and matched measurements from conventional images at 75 keV. Measurements from the 20 mlmore » insert showed the CT number is independent of location and CNR is a function only of noise, which was higher in the center. Measured concentration from iodine-no-water images matched phantom manufacturer suggested concentration to within 6% on average for inserts at the center of the phantom. Measured concentrations were systematically higher due to optimization of iodine quantification parameters for clinical mixtures of iodine and blood/tissue. Conclusion: Spectral acquisition and reconstruction with a dual-layer detector represents the physical behavior of iodine as expected and accurately quantifies the material concentration. This should permit a variety of clinical applications including lesion characterization, vessel patency, and myocardial perfusion. This study was performed as part of a research agreement among Philips Healthcare, University Hospitals of Cleveland, and Case Western Reserve University.« less

Authors:
 [1]; ; ;  [2];  [3];  [4]
  1. Case Western Reserve University, Cleveland, Ohio (United States)
  2. Philips Healthcare, Highland Heights, OH (United States)
  3. The University of Texas at Austin, Austin, TX (United States)
  4. University Hospitals Case Medical Center, Cleveland, OH (United States)
Publication Date:
OSTI Identifier:
22649361
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 43; Journal Issue: 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; 61 RADIATION PROTECTION AND DOSIMETRY; BIOMEDICAL RADIOGRAPHY; COMPUTERIZED TOMOGRAPHY; CONCENTRATION RATIO; ECOLOGICAL CONCENTRATION; EDUCATIONAL FACILITIES; IMAGES; IODINE; KEV RANGE 100-1000; KEV RANGE 10-100; PHANTOMS

Citation Formats

Ozguner, O, Dhanantwari, A, Halliburton, S, Utrup, S, Wen, G, and Jordan, D. SU-G-IeP2-09: Iodine Imaging at Spectral CT with a Dual-Layer Detector. United States: N. p., 2016. Web. doi:10.1118/1.4957014.
Ozguner, O, Dhanantwari, A, Halliburton, S, Utrup, S, Wen, G, & Jordan, D. SU-G-IeP2-09: Iodine Imaging at Spectral CT with a Dual-Layer Detector. United States. doi:10.1118/1.4957014.
Ozguner, O, Dhanantwari, A, Halliburton, S, Utrup, S, Wen, G, and Jordan, D. 2016. "SU-G-IeP2-09: Iodine Imaging at Spectral CT with a Dual-Layer Detector". United States. doi:10.1118/1.4957014.
@article{osti_22649361,
title = {SU-G-IeP2-09: Iodine Imaging at Spectral CT with a Dual-Layer Detector},
author = {Ozguner, O and Dhanantwari, A and Halliburton, S and Utrup, S and Wen, G and Jordan, D},
abstractNote = {Purpose: To evaluate the attenuation response of iodine and the accuracy of iodine quantification on a detector-based spectral CT scanner. Methods: A Gammex 461A phantom was scanned using a dual-layer detector (IQon, Philips) at 120 kVp using helical acquisition with a CDTIvol of 15 mGy to approximate the hospital’s clinical body protocol. No modifications to the standard protocol were necessary to enable spectral imaging. Iodine inserts at 6 concentrations (2, 5, 7.5, 10, 15, 20 mg/ml) were scanned individually at the center of the phantom and the 20 mg/ml insert was additionally scanned at the 3, 6, and 12 o’clock positions. Scans were repeated 10 times. Conventional, virtual monoenergetic (40–200 keV) and iodine-no-water images (with pixel values equal to iodine concentration of corresponding tissue) were reconstructed from acquired data. A circular ROI (diameter=30 pixels) was used in each conventional and monoenergetic image to measure the mean and standard deviation of the CT number in HU and in each iodine-no-water image to measure iodine concentration in mg/ml. Results: Mean CT number and contrast-to-noise ratio (CNR) measured from monoenergetic images increased with decreasing keV for all iodine concentrations and matched measurements from conventional images at 75 keV. Measurements from the 20 ml insert showed the CT number is independent of location and CNR is a function only of noise, which was higher in the center. Measured concentration from iodine-no-water images matched phantom manufacturer suggested concentration to within 6% on average for inserts at the center of the phantom. Measured concentrations were systematically higher due to optimization of iodine quantification parameters for clinical mixtures of iodine and blood/tissue. Conclusion: Spectral acquisition and reconstruction with a dual-layer detector represents the physical behavior of iodine as expected and accurately quantifies the material concentration. This should permit a variety of clinical applications including lesion characterization, vessel patency, and myocardial perfusion. This study was performed as part of a research agreement among Philips Healthcare, University Hospitals of Cleveland, and Case Western Reserve University.},
doi = {10.1118/1.4957014},
journal = {Medical Physics},
number = 6,
volume = 43,
place = {United States},
year = 2016,
month = 6
}
  • Purpose: In order to investigate novel methods to more accurately estimate the mineral composition of kidney stones using dual energy CT, it is desirable to be able to combine digital stones of known composition with actual phantom and patient scan data. In this work, we developed and validated a method to insert digital kidney stones into projection data acquired on a dual-source, dual-energy CT system. Methods: Attenuation properties of stones of different mineral composition were computed using tabulated mass attenuation coefficients, the chemical formula for each stone type, and the effective beam energy at each evaluated tube potential. A previouslymore » developed method to insert lesions into x-ray CT projection data was extended to include simultaneous dual-energy CT projections acquired on a dual-source gantry (Siemens Somatom Flash). Digital stones were forward projected onto both detectors and the resulting projections added to the physically acquired sinogram data. To validate the accuracy of the technique, digital stones were inserted into different locations in the ACR CT accreditation phantom; low and high contrast resolution, CT number accuracy and noise properties were compared before and after stone insertion. The procedure was repeated for two dual-energy tube potential pairs in clinical use on the scanner, 80/Sn140 kV and 100/Sn140 kV, respectively. Results: The images reconstructed after the insertion of digital kidney stones were consistent with the images reconstructed from the scanner. The largest average CT number difference for the 4 insert in the CT number accuracy module of the phantom was 3 HU. Conclusion: A framework was developed and validated for the creation of digital kidney stones of known mineral composition, and their projection-domain insertion into commercial dual-source, dual-energy CT projection data. This will allow a systematic investigation of the impact of scan and reconstruction parameters on stone attenuation and dual-energy behavior under rigorously controlled conditions. Dr. McCollough receives research support from Siemens Healthcare.« less
  • Purpose: To validate Computed Tomography Fractional Flow Reserve (CT-FFR) measurements with accurate 3D printed coronary phantoms. Methods: DICOM data from four phases in two patients imaged with a standard 320 × 0.5mm coronary CT acquisition (70–80% cardiac cycle) underwent semi-automated segmentation using a research workstation. Both patients had a >50% stenosis from the clinical image interpretation. Each volume was saved as a Stereo Lithographic (STL) file with 250 micron resolution. The 3D geometries were qualitatively assessed; the best of the four phases was 3D printed using a Stratasys Eden260V printer in Tango+, a rubber-like material that roughly emulates mechanical propertiesmore » of human vasculature. We connected the model to a programmable pump and measured the pressure drop using pressure sensors embedded proximal and distal to the arterial stenosis. Next, the STL files used for the 3D printed models were uploaded in the ANSYS meshing tool (ICEM CFD 16.1). A standard meshing process was applied and the meshed geometry was directly imported in the ANSYS Fluent for Computational Flow Dynamics simulations. The CFD simulations were used to calculate the CT-FFR and compared to the bench top FFR measured in the 3D printed phantoms. Results: FFR-CT measurements and phantoms were completed in within an hour after the segmentation. Patient 1 had a 60% stenosis that resulted in a CT-FFR of 0.68. The second case had a 50% stenosis and a CT-FFR of 0.75. The average bench top FFR measurements were 0.72 and 0.80, respectively. Conclusion: This pilot investigation demonstrated the use of a bench-top coronary model for CT-FFR validation. The measurements and the CFD simulations agreed within 6%. Project supported by Support: Toshiba America Medical Systems Corp.and NIH grant R01-EB002873. Project supported by Toshiba America Medical Systems Corp.and partial support from NIH grant R01-EB002873.« less
  • Purpose: The study has the goal to demonstrate that breast and soft tissue-equivalent phantoms for dosimetry applications in the diagnostic energy range can be fabricated using common 3D printing methods. Methods: 3D printing provides the opportunity to rapidly prototype uniquely designed objects from a variety of materials. Common 3D printers are usually limited to printing objects based on thermoplastic materials such as PLA, or ABS. The most commonly available plastic is PLA, which has a density significantly greater than soft tissue. We utilized a popular 3D printer to demonstrate that tissue specific phantom materials can be generated through the carefulmore » selection of 3D printing parameters. A series of stepwedges were designed and printed using a Makerbot Replicator2 3D printing system. The print file provides custom adjustment of the infill density, orientation and position of the object on the printer stage, selection of infill patterns, and other control parameters. The x-ray attenuation and uniformity of fabricated phantoms were evaluated and compared to common tissue-equivalent phantom materials, acrylic and BR12. X-ray exposure measurements were made using narrow beam geometry on a clinical mammography unit at 28 kVp on the series of phantoms. The 3D printed phantoms were imaged at 28 kVp to visualize the internal structure and uniformity in different planes of the phantoms. Results: By utilizing specific in-fill density and patterns we are able to produce a phantom closely matching the attenuation characteristics of BR12 at 28 kVp. The in-fill patterns used are heterogeneous, so a judicious selection of fill pattern and the orientation of the fill pattern must be made in order to obtain homogenous attenuation along the intended direction of beam propagation. Conclusions: By careful manipulation of the printing parameters, breast and soft tissue-equivalent phantoms appropriate for use at imaging energies can be fabricated using 3D printing techniques.« less
  • Purpose: To study the performance of different dual energy computed tomography (DECT) techniques, which are available today, and future multi energy CT (MECT) employing novel photon counting detectors in an image-based material decomposition task. Methods: The material decomposition performance of different energy-resolved CT acquisition techniques is assessed and compared in a simulation study of virtual non-contrast imaging and iodine quantification. The material-specific images are obtained via a statistically optimal image-based material decomposition. A projection-based maximum likelihood approach was used for comparison with the authors’ image-based method. The different dedicated dual energy CT techniques are simulated employing realistic noise models andmore » x-ray spectra. The authors compare dual source DECT with fast kV switching DECT and the dual layer sandwich detector DECT approach. Subsequent scanning and a subtraction method are studied as well. Further, the authors benchmark future MECT with novel photon counting detectors in a dedicated DECT application against the performance of today’s DECT using a realistic model. Additionally, possible dual source concepts employing photon counting detectors are studied. Results: The DECT comparison study shows that dual source DECT has the best performance, followed by the fast kV switching technique and the sandwich detector approach. Comparing DECT with future MECT, the authors found noticeable material image quality improvements for an ideal photon counting detector; however, a realistic detector model with multiple energy bins predicts a performance on the level of dual source DECT at 100 kV/Sn 140 kV. Employing photon counting detectors in dual source concepts can improve the performance again above the level of a single realistic photon counting detector and also above the level of dual source DECT. Conclusions: Substantial differences in the performance of today’s DECT approaches were found for the application of virtual non-contrast and iodine imaging. Future MECT with realistic photon counting detectors currently can only perform comparably to dual source DECT at 100 kV/Sn 140 kV. Dual source concepts with photon counting detectors could be a solution to this problem, promising a better performance.« less
  • A dual wavelength imaging system has been developed and installed on C-2U to capture 2D images of a He jet in the Scrape-Off Layer (SOL) of an advanced beam-driven Field-Reversed Configuration (FRC) plasma. The system was designed to optically split two identical images and pass them through 1 nm FWHM filters. Dual wavelength images are focused adjacent on a large format CCD chip and recorded simultaneously with a time resolution down to 10 μs using a gated micro-channel plate. The relatively compact optical system images a 10 cm plasma region with a spatial resolution of 0.2 cm and can bemore » used in a harsh environment with high electro-magnetic noise and high magnetic field. The dual wavelength imaging system provides 2D images of either electron density or temperature by observing spectral line pairs emitted by He jet atoms in the SOL. A large field of view, combined with good space and time resolution of the imaging system, allows visualization of macro-flows in the SOL. First 2D images of the electron density and temperature observed in the SOL of the C-2U FRC are presented.« less