Title: SU-G-BRB-03: Assessing the Sensitivity and False Positive Rate of the Integrated Quality Monitor (IQM) Large Area Ion Chamber to MLC Positioning Errors

Abstract

Purpose: To examine both the IQM’s sensitivity and false positive rate to varying MLC errors. By balancing these two characteristics, an optimal tolerance value can be derived. Methods: An un-modified SBRT Liver IMRT plan containing 7 fields was randomly selected as a representative clinical case. The active MLC positions for all fields were perturbed randomly from a square distribution of varying width (±1mm to ±5mm). These unmodified and modified plans were measured multiple times each by the IQM (a large area ion chamber mounted to a TrueBeam linac head). Measurements were analyzed relative to the initial, unmodified measurement. IQM readings are analyzed as a function of control points. In order to examine sensitivity to errors along a field’s delivery, each measured field was divided into 5 groups of control points, and the maximum error in each group was recorded. Since the plans have known errors, we compared how well the IQM is able to differentiate between unmodified and error plans. ROC curves and logistic regression were used to analyze this, independent of thresholds. Results: A likelihood-ratio Chi-square test showed that the IQM could significantly predict whether a plan had MLC errors, with the exception of the beginning and ending controlmore » points. Upon further examination, we determined there was ramp-up occurring at the beginning of delivery. Once the linac AFC was tuned, the subsequent measurements (relative to a new baseline) showed significant (p <0.005) abilities to predict MLC errors. Using the area under the curve, we show the IQM’s ability to detect errors increases with increasing MLC error (Spearman’s Rho=0.8056, p<0.0001). The optimal IQM count thresholds from the ROC curves are ±3%, ±2%, and ±7% for the beginning, middle 3, and end segments, respectively. Conclusion: The IQM has proven to be able to detect not only MLC errors, but also differences in beam tuning (ramp-up). Partially supported by the Susan Scott Foundation.« less

Authors:
; ; ;  [1]
  1. Cedars-Sinai Medical Center, Los Angeles, CA (United States)
Publication Date:
OSTI Identifier:
22649275
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 43; Journal Issue: 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; 61 RADIATION PROTECTION AND DOSIMETRY; ERRORS; IONIZATION CHAMBERS; LINEAR ACCELERATORS; POSITIONING; RADIOTHERAPY; SENSITIVITY

Citation Formats

Boehnke, E McKenzie, DeMarco, J, Steers, J, and Fraass, B. SU-G-BRB-03: Assessing the Sensitivity and False Positive Rate of the Integrated Quality Monitor (IQM) Large Area Ion Chamber to MLC Positioning Errors. United States: N. p., 2016. Web. doi:10.1118/1.4956910.
Boehnke, E McKenzie, DeMarco, J, Steers, J, & Fraass, B. SU-G-BRB-03: Assessing the Sensitivity and False Positive Rate of the Integrated Quality Monitor (IQM) Large Area Ion Chamber to MLC Positioning Errors. United States. doi:10.1118/1.4956910.
Boehnke, E McKenzie, DeMarco, J, Steers, J, and Fraass, B. 2016. "SU-G-BRB-03: Assessing the Sensitivity and False Positive Rate of the Integrated Quality Monitor (IQM) Large Area Ion Chamber to MLC Positioning Errors". United States. doi:10.1118/1.4956910.
@article{osti_22649275,
title = {SU-G-BRB-03: Assessing the Sensitivity and False Positive Rate of the Integrated Quality Monitor (IQM) Large Area Ion Chamber to MLC Positioning Errors},
author = {Boehnke, E McKenzie and DeMarco, J and Steers, J and Fraass, B},
abstractNote = {Purpose: To examine both the IQM’s sensitivity and false positive rate to varying MLC errors. By balancing these two characteristics, an optimal tolerance value can be derived. Methods: An un-modified SBRT Liver IMRT plan containing 7 fields was randomly selected as a representative clinical case. The active MLC positions for all fields were perturbed randomly from a square distribution of varying width (±1mm to ±5mm). These unmodified and modified plans were measured multiple times each by the IQM (a large area ion chamber mounted to a TrueBeam linac head). Measurements were analyzed relative to the initial, unmodified measurement. IQM readings are analyzed as a function of control points. In order to examine sensitivity to errors along a field’s delivery, each measured field was divided into 5 groups of control points, and the maximum error in each group was recorded. Since the plans have known errors, we compared how well the IQM is able to differentiate between unmodified and error plans. ROC curves and logistic regression were used to analyze this, independent of thresholds. Results: A likelihood-ratio Chi-square test showed that the IQM could significantly predict whether a plan had MLC errors, with the exception of the beginning and ending control points. Upon further examination, we determined there was ramp-up occurring at the beginning of delivery. Once the linac AFC was tuned, the subsequent measurements (relative to a new baseline) showed significant (p <0.005) abilities to predict MLC errors. Using the area under the curve, we show the IQM’s ability to detect errors increases with increasing MLC error (Spearman’s Rho=0.8056, p<0.0001). The optimal IQM count thresholds from the ROC curves are ±3%, ±2%, and ±7% for the beginning, middle 3, and end segments, respectively. Conclusion: The IQM has proven to be able to detect not only MLC errors, but also differences in beam tuning (ramp-up). Partially supported by the Susan Scott Foundation.},
doi = {10.1118/1.4956910},
journal = {Medical Physics},
number = 6,
volume = 43,
place = {United States},
year = 2016,
month = 6
}
  • Purpose: The Integral Quality Monitor (IQM), developed by iRT Systems GmbH (Koblenz, Germany) is a large-area, linac-mounted ion chamber used to monitor photon fluence during patient treatment. Our previous work evaluated the change of the ion chamber’s response to deviations from static 1×1 cm2 and 10×10 cm2 photon beams and other characteristics integral to use in external beam detection. The aim of this work is to simulate two external beam radiation delivery errors, quantify the detection of simulated errors and evaluate the reduction in patient harm resulting from detection. Methods: Two well documented radiation oncology delivery errors were selected formore » simulation. The first error was recreated by modifying a wedged whole breast treatment, removing the physical wedge and calculating the planned dose with Pinnacle TPS (Philips Radiation Oncology Systems, Fitchburg, WI). The second error was recreated by modifying a static-gantry IMRT pharyngeal tonsil plan to be delivered in 3 unmodulated fractions. A radiation oncologist evaluated the dose for simulated errors and predicted morbidity and mortality commiserate with the original reported toxicity, indicating that reported errors were approximately simulated. The ion chamber signal of unmodified treatments was compared to the simulated error signal and evaluated in Pinnacle TPS again with radiation oncologist prediction of simulated patient harm. Results: Previous work established that transmission detector system measurements are stable within 0.5% standard deviation (SD). Errors causing signal change greater than 20 SD (10%) were considered detected. The whole breast and pharyngeal tonsil IMRT simulated error increased signal by 215% and 969%, respectively, indicating error detection after the first fraction and IMRT segment, respectively. Conclusion: The transmission detector system demonstrated utility in detecting clinically significant errors and reducing patient toxicity/harm in simulated external beam delivery. Future work will evaluate detection of other smaller magnitude delivery errors.« less
  • Intensity modulated radiotherapy (IMRT) irradiation technique uses steep dose gradients to protect organs at risk (OAR) during treatment course. Therefore even small mistakes in patient positioning during treatment course may lead to significant changes in dose distribution. The influence of small errors (order of millimeters) in patient's positioning on resulting dose absorbed in OARs and target volume (PTV, GTV) was simulated in this work.
  • Quality assurance (QA) procedures for intensity modulation radiation therapy (IMRT) usually involve an ion chamber measurement in a phantom using the beam configuration of the actual treatment plan. In our QA procedures it was observed that the degree of agreement between the measurement and the calculation could vary from plan to plan, from linac to linac, as well as over time, with a discrepancy up to 8%. In this paper we examine one aspect of the process which can contribute to such poor reproducibility, namely, the leaf end position accuracy. A series of measurements was designed to irradiate an ionmore » chamber using small beam segments where one multileaf collimator (MLC) edge covers half of the chamber. It was shown that the reproducibility varied up to 13%, which provides a possible explanation for the observed discrepancies above. A useful tool was also developed to measure ionization signals from individual segments of an IMRT sequence. In addition, an understanding of the leaf end position variations offers some insight into the overall quality of an IMRT dose distribution.« less
  • Geospatial technologies and digital data have developed and disseminated rapidly in conjunction with increasing computing performance and internet availability. The ability to store and transmit large datasets has encouraged the development of national datasets in geospatial format. National datasets are used by numerous agencies for analysis and modeling purposes because these datasets are standardized, and are considered to be of acceptable accuracy. At Oak Ridge National Laboratory, a national population model incorporating multiple ancillary variables was developed and one of the inputs required is a school database. This paper examines inaccuracies present within two national school datasets, TeleAtlas North Americamore » (TANA) and National Center of Education Statistics (NCES). Schools are an important component of the population model, because they serve as locations containing dense clusters of vulnerable populations. It is therefore essential to validate the quality of the school input data, which was made possible by increasing national coverage of high resolution imagery. Schools were also chosen since a 'real-world' representation of K-12 schools for the Philadelphia School District was produced; thereby enabling 'ground-truthing' of the national datasets. Analyses found the national datasets not standardized and incomplete, containing 76 to 90% of existing schools. The temporal accuracy of enrollment values of updating national datasets resulted in 89% inaccuracy to match 2003 data. Spatial rectification was required for 87% of the NCES points, of which 58% of the errors were attributed to the geocoding process. Lastly, it was found that by combining the two national datasets together, the resultant dataset provided a more useful and accurate solution. Acknowledgment Prepared by Oak Ridge National Laboratory, P.O. Box 2008, Oak Ridge, Tennessee 37831-6285, managed by UT-Battelle, LLC for the U. S. Department of Energy undercontract no. DEAC05-00OR22725. Copyright This manuscript has been authored by employees of UT-Battelle, LLC, under contract DE-AC05-00OR22725 with the U.S. Department of Energy. Accordingly, the United States Government retains and the publisher, by accepting the article for publication, acknowledges that the United States Government retains a non-exclusive, paid-up, irrevocable, world-wide license to publish or reproduce the published form of this manuscript, or allow others to do so, for United States Government purposes.« less
  • The Kepler mission is uniquely suited to study the frequencies of extrasolar planets. This goal requires knowledge of the incidence of false positives such as eclipsing binaries in the background of the targets, or physically bound to them, which can mimic the photometric signal of a transiting planet. We perform numerical simulations of the Kepler targets and of physical companions or stars in the background to predict the occurrence of astrophysical false positives detectable by the mission. Using real noise level estimates, we compute the number and characteristics of detectable eclipsing pairs involving main-sequence stars and non-main-sequence stars or planets,more » and we quantify the fraction of those that would pass the Kepler candidate vetting procedure. By comparing their distribution with that of the Kepler Objects of Interest (KOIs) detected during the first six quarters of operation of the spacecraft, we infer the false positive rate of Kepler and study its dependence on spectral type, candidate planet size, and orbital period. We find that the global false positive rate of Kepler is 9.4%, peaking for giant planets (6-22 R{sub Circled-Plus }) at 17.7%, reaching a low of 6.7% for small Neptunes (2-4 R{sub Circled-Plus }), and increasing again for Earth-size planets (0.8-1.25 R{sub Circled-Plus }) to 12.3%. Most importantly, we also quantify and characterize the distribution and rate of occurrence of planets down to Earth size with no prior assumptions on their frequency, by subtracting from the population of actual Kepler candidates our simulated population of astrophysical false positives. We find that 16.5% {+-} 3.6% of main-sequence FGK stars have at least one planet between 0.8 and 1.25 R{sub Circled-Plus} with orbital periods up to 85 days. This result is a significant step toward the determination of eta-earth, the occurrence of Earth-like planets in the habitable zone of their parent stars. There is no significant dependence of the rates of planet occurrence between 0.8 and 4 Earth radii with spectral type. In the process, we also derive a prescription for the signal recovery rate of Kepler that enables a good match to both the KOI size and orbital period distribution, as well as their signal-to-noise distribution.« less