Title: SU-G-201-13: Investigation of Dose Variation Induced by HDR Ir-192 Source Global Shift Within the Varian Ring Applicator Using Monte Carlo Methods

Abstract

Purpose: The Varian tandem and ring applicators are used to deliver HDR Ir-192 brachytherapy for cervical cancer. The source path within the ring is hard to predict due to the larger interior ring lumen. Some studies showed the source could be several millimeters different from planned positions, while other studies demonstrated minimal dosimetric impact. A global shift can be applied to limit the effect of positioning offsets. The purpose of this study was to assess the necessities of implementing a global source shift using Monte Carlo (MC) simulations. Methods: The MCNP5 radiation transport code was used for all MC simulations. To accommodate TG-186 guidelines and eliminate inter-source attenuation, a BrachyVision plan with 10 dwell positions (0.5cm step sizes) was simulated as the summation of 10 individual sources with equal dwell times for simplification. To simplify the study, the tandem was also excluded from the MC model. Global shifts of ±0.1, ±0.3, ±0.5 cm were then simulated as distal and proximal from the reference positions. Dose was scored in water for all MC simulations and was normalized to 100% at the normalization point 0.5 cm from the cap in the ring plane. For dose comparison, Point A was 2 cm caudalmore » from the buildup cap and 2 cm lateral on either side of the ring axis. With seventy simulations, 108 photon histories gave a statistical uncertainties (k=1) <2% for (0.1 cm)3 voxels. Results: Compared to no global shift, average Point A doses were 0.0%, 0.4%, and 2.2% higher for distal global shifts, and 0.4%, 2.8%, and 5.1% higher for proximal global shifts, respectively. The MC Point A doses differed by < 1% when compared to BrachyVision. Conclusion: Dose variations were not substantial for ±0.3 cm global shifts, which is common in clinical practice.« less

Authors:
; ; ; ;  [1]
  1. Duke University Medical Center, Durham, NC (United States)
Publication Date:
OSTI Identifier:
22649255
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 43; Journal Issue: 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; 61 RADIATION PROTECTION AND DOSIMETRY; BRACHYTHERAPY; IRIDIUM 192; MONTE CARLO METHOD; POSITIONING; RADIATION DOSES; RINGS; SIMULATION; VARIATIONS

Citation Formats

Yang, Y, Cai, J, Meltsner, S, Chang, Z, and Craciunescu, O. SU-G-201-13: Investigation of Dose Variation Induced by HDR Ir-192 Source Global Shift Within the Varian Ring Applicator Using Monte Carlo Methods. United States: N. p., 2016. Web. doi:10.1118/1.4956886.
Yang, Y, Cai, J, Meltsner, S, Chang, Z, & Craciunescu, O. SU-G-201-13: Investigation of Dose Variation Induced by HDR Ir-192 Source Global Shift Within the Varian Ring Applicator Using Monte Carlo Methods. United States. doi:10.1118/1.4956886.
Yang, Y, Cai, J, Meltsner, S, Chang, Z, and Craciunescu, O. Wed . "SU-G-201-13: Investigation of Dose Variation Induced by HDR Ir-192 Source Global Shift Within the Varian Ring Applicator Using Monte Carlo Methods". United States. doi:10.1118/1.4956886.
@article{osti_22649255,
title = {SU-G-201-13: Investigation of Dose Variation Induced by HDR Ir-192 Source Global Shift Within the Varian Ring Applicator Using Monte Carlo Methods},
author = {Yang, Y and Cai, J and Meltsner, S and Chang, Z and Craciunescu, O},
abstractNote = {Purpose: The Varian tandem and ring applicators are used to deliver HDR Ir-192 brachytherapy for cervical cancer. The source path within the ring is hard to predict due to the larger interior ring lumen. Some studies showed the source could be several millimeters different from planned positions, while other studies demonstrated minimal dosimetric impact. A global shift can be applied to limit the effect of positioning offsets. The purpose of this study was to assess the necessities of implementing a global source shift using Monte Carlo (MC) simulations. Methods: The MCNP5 radiation transport code was used for all MC simulations. To accommodate TG-186 guidelines and eliminate inter-source attenuation, a BrachyVision plan with 10 dwell positions (0.5cm step sizes) was simulated as the summation of 10 individual sources with equal dwell times for simplification. To simplify the study, the tandem was also excluded from the MC model. Global shifts of ±0.1, ±0.3, ±0.5 cm were then simulated as distal and proximal from the reference positions. Dose was scored in water for all MC simulations and was normalized to 100% at the normalization point 0.5 cm from the cap in the ring plane. For dose comparison, Point A was 2 cm caudal from the buildup cap and 2 cm lateral on either side of the ring axis. With seventy simulations, 108 photon histories gave a statistical uncertainties (k=1) <2% for (0.1 cm)3 voxels. Results: Compared to no global shift, average Point A doses were 0.0%, 0.4%, and 2.2% higher for distal global shifts, and 0.4%, 2.8%, and 5.1% higher for proximal global shifts, respectively. The MC Point A doses differed by < 1% when compared to BrachyVision. Conclusion: Dose variations were not substantial for ±0.3 cm global shifts, which is common in clinical practice.},
doi = {10.1118/1.4956886},
journal = {Medical Physics},
number = 6,
volume = 43,
place = {United States},
year = {Wed Jun 15 00:00:00 EDT 2016},
month = {Wed Jun 15 00:00:00 EDT 2016}
}