Title: SU-F-T-611: Critical Analysis and Efficacy of Linac Based (Beam Modulator) and Cyberknife Treatment Plans for Acoustic Neuroma/schwannoma

Abstract

Purpose: To study the critical analysis and efficacy of Linac and Cyberknife (CK) treatment plans for acoustic neuroma/schwannoma. Methods: Twelve of acoustic neuroma/schwannoma patients were taken for these study that. Treatment plans were generated in Multiplan treatment planning system (TPS) for CK using 5,7.5 and 10mm diameter collimators. Target volumes were in the range of 0.280 cc to 9.256 cc. Prescription dose (Rx) ranges from 1150cGy to 1950cGy delivered over 1 to 3 Fractions. For same patients stereotactic Volumetric modulated arc plans were generated using Elekta Linac with MLC thickness of 4mm in Monaco TPS. Appropriate calculation algorithms and grid size were used with same Rx and organ at risk (OAR) constrains for both Linac and CK plans. Treatment plans were developed to achieve at least 95% of the target volume to receive the Rx. The dosimetric indices such as conformity index (CI), coverage, OAR dose and volume receiving 50% of Rx (V50%) were used to evaluate the plans. Results: Target volumes ranges from 0.280 cc to 3.5cc shows the CI of 1.16±0.109 and 1.53±0.360 for cyberknife and Linac plans respectively. For small volume targets, the OARs were well spared in CK plans. There are no significant differences in CImore » and OAR doses were observed between CK and Linac plans that have the target volume >3.5 cc. Perhaps the V50% were lesser in CK plans, and found to be 12.8± 8.4 and 22.8 ± 15.0 for CK and Linac respectively. Conclusion: The analysis shows the importance of collimator size for small volume targets. The target volumes >3.5 cc can be treated in Linac as comparable with CK. For targets <3.5cc CK plans showed superior plan quality with better CI and OAR sparing than the Linac based plans. Further studies may require evaluating the clinical advantage of CK robotic system.« less

Authors:
; ; ; ;  [1]
  1. Medanta- The Medicity, Gurgaon, Haryana (Spain)
Publication Date:
OSTI Identifier:
22649176
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 43; Journal Issue: 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; 61 RADIATION PROTECTION AND DOSIMETRY; ACOUSTICS; LINEAR ACCELERATORS; PLANNING; RADIATION DOSES; RADIOTHERAPY

Citation Formats

KP, Karrthick, Kataria, T, Thiyagarajan, R, Selvan, T, and Abhishek, A. SU-F-T-611: Critical Analysis and Efficacy of Linac Based (Beam Modulator) and Cyberknife Treatment Plans for Acoustic Neuroma/schwannoma. United States: N. p., 2016. Web. doi:10.1118/1.4956796.
KP, Karrthick, Kataria, T, Thiyagarajan, R, Selvan, T, & Abhishek, A. SU-F-T-611: Critical Analysis and Efficacy of Linac Based (Beam Modulator) and Cyberknife Treatment Plans for Acoustic Neuroma/schwannoma. United States. doi:10.1118/1.4956796.
KP, Karrthick, Kataria, T, Thiyagarajan, R, Selvan, T, and Abhishek, A. Wed . "SU-F-T-611: Critical Analysis and Efficacy of Linac Based (Beam Modulator) and Cyberknife Treatment Plans for Acoustic Neuroma/schwannoma". United States. doi:10.1118/1.4956796.
@article{osti_22649176,
title = {SU-F-T-611: Critical Analysis and Efficacy of Linac Based (Beam Modulator) and Cyberknife Treatment Plans for Acoustic Neuroma/schwannoma},
author = {KP, Karrthick and Kataria, T and Thiyagarajan, R and Selvan, T and Abhishek, A},
abstractNote = {Purpose: To study the critical analysis and efficacy of Linac and Cyberknife (CK) treatment plans for acoustic neuroma/schwannoma. Methods: Twelve of acoustic neuroma/schwannoma patients were taken for these study that. Treatment plans were generated in Multiplan treatment planning system (TPS) for CK using 5,7.5 and 10mm diameter collimators. Target volumes were in the range of 0.280 cc to 9.256 cc. Prescription dose (Rx) ranges from 1150cGy to 1950cGy delivered over 1 to 3 Fractions. For same patients stereotactic Volumetric modulated arc plans were generated using Elekta Linac with MLC thickness of 4mm in Monaco TPS. Appropriate calculation algorithms and grid size were used with same Rx and organ at risk (OAR) constrains for both Linac and CK plans. Treatment plans were developed to achieve at least 95% of the target volume to receive the Rx. The dosimetric indices such as conformity index (CI), coverage, OAR dose and volume receiving 50% of Rx (V50%) were used to evaluate the plans. Results: Target volumes ranges from 0.280 cc to 3.5cc shows the CI of 1.16±0.109 and 1.53±0.360 for cyberknife and Linac plans respectively. For small volume targets, the OARs were well spared in CK plans. There are no significant differences in CI and OAR doses were observed between CK and Linac plans that have the target volume >3.5 cc. Perhaps the V50% were lesser in CK plans, and found to be 12.8± 8.4 and 22.8 ± 15.0 for CK and Linac respectively. Conclusion: The analysis shows the importance of collimator size for small volume targets. The target volumes >3.5 cc can be treated in Linac as comparable with CK. For targets <3.5cc CK plans showed superior plan quality with better CI and OAR sparing than the Linac based plans. Further studies may require evaluating the clinical advantage of CK robotic system.},
doi = {10.1118/1.4956796},
journal = {Medical Physics},
number = 6,
volume = 43,
place = {United States},
year = {Wed Jun 15 00:00:00 EDT 2016},
month = {Wed Jun 15 00:00:00 EDT 2016}
}