Title: SU-F-T-598: Robotic Radiosurgery System Versus Pencil Beam Scanning Proton Therapy for Definitive Intracranial Treatments

Abstract

Purpose: To characterize the dose distributions of Cyberknife and intensity-modulated-proton-therapy (IMPT). Methods: A total of 20 patients previously treated with Cyberknife were selected. The original planning-target-volume (PTV) was used in the ‘IMPT-ideal’ plan assuming a comparable image-guidance with Cyberknife. A 3mm expansion was made to create the proton-PTV for the ‘IMPT-3mm’ plan representing the current proton-therapy where a margin of 3mm is used to account for the inferior image-guidance. The proton range uncertainty was taken-care in beam-design by adding the proximal- and distal-margins (3%water-equivalent-depth+1mm) for both proton plans. The IMPT plans were generated to meet the same target coverage as the Cyberknife-plans. The plan quality of IMPT-ideal and IMPT-3mm were compared to the Cyberknife-plan. To characterize plan quality, we defined the ratio(R) of volumes encompassed by the selected isodose surfaces for Cyberknife and IMPT plans (VCK/VIMPT). Comparisons were made for both Cyberknife versus IMPT-ideal and Cyberknife versusIMPT-3mm to further discuss the impact of setup error margins used in proton therapy and the correlation with target size and location. Results: IMPT-ideal plans yield comparable plan quality as CK plans and slightly better OAR sparing while the IMPT-3mm plan results in a higher dose to the OARs, especially for centralized tumors. Comparingmore » to the IMPT-ideal plans, a slightly larger 80% (Ravg=1.05) dose cloud and significantly larger 50% (Ravg=1.3) and 20% (Ravg=1.60) dose clouds are seen in CK plans. However, the 3mm expansion results in a larger high and medium dose clouds in IMPT-3mm plans (Ravg=0.65 for 80%-isodose; Ravg=0.93 for 50%-isodose). The trend increases with the size of the target and the distance from the brainstem to the center of target. Conclusion: Cyberknife is more preferable for treating centralized targets and proton therapy is advantageous for the large and peripheral targets. Advanced image guidance would improve the efficacy of proton therapy for intracranial treatments.« less

Authors:
; ; ; ; ; ;  [1]
  1. UT Southwestern Medical Center, Dallas, TX (United States)
Publication Date:
OSTI Identifier:
22649169
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 43; Journal Issue: 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; 61 RADIATION PROTECTION AND DOSIMETRY; DESIGN; PLANNING; PRODUCTIVITY; PROTON BEAMS; RADIATION DOSE DISTRIBUTIONS; RADIOTHERAPY; WATER

Citation Formats

Lin, M, Pompos, A, Gu, X, Yan, Y, Abdulrahman, R, Timmerman, R, and Jiang, S. SU-F-T-598: Robotic Radiosurgery System Versus Pencil Beam Scanning Proton Therapy for Definitive Intracranial Treatments. United States: N. p., 2016. Web. doi:10.1118/1.4956783.
Lin, M, Pompos, A, Gu, X, Yan, Y, Abdulrahman, R, Timmerman, R, & Jiang, S. SU-F-T-598: Robotic Radiosurgery System Versus Pencil Beam Scanning Proton Therapy for Definitive Intracranial Treatments. United States. doi:10.1118/1.4956783.
Lin, M, Pompos, A, Gu, X, Yan, Y, Abdulrahman, R, Timmerman, R, and Jiang, S. Wed . "SU-F-T-598: Robotic Radiosurgery System Versus Pencil Beam Scanning Proton Therapy for Definitive Intracranial Treatments". United States. doi:10.1118/1.4956783.
@article{osti_22649169,
title = {SU-F-T-598: Robotic Radiosurgery System Versus Pencil Beam Scanning Proton Therapy for Definitive Intracranial Treatments},
author = {Lin, M and Pompos, A and Gu, X and Yan, Y and Abdulrahman, R and Timmerman, R and Jiang, S},
abstractNote = {Purpose: To characterize the dose distributions of Cyberknife and intensity-modulated-proton-therapy (IMPT). Methods: A total of 20 patients previously treated with Cyberknife were selected. The original planning-target-volume (PTV) was used in the ‘IMPT-ideal’ plan assuming a comparable image-guidance with Cyberknife. A 3mm expansion was made to create the proton-PTV for the ‘IMPT-3mm’ plan representing the current proton-therapy where a margin of 3mm is used to account for the inferior image-guidance. The proton range uncertainty was taken-care in beam-design by adding the proximal- and distal-margins (3%water-equivalent-depth+1mm) for both proton plans. The IMPT plans were generated to meet the same target coverage as the Cyberknife-plans. The plan quality of IMPT-ideal and IMPT-3mm were compared to the Cyberknife-plan. To characterize plan quality, we defined the ratio(R) of volumes encompassed by the selected isodose surfaces for Cyberknife and IMPT plans (VCK/VIMPT). Comparisons were made for both Cyberknife versus IMPT-ideal and Cyberknife versusIMPT-3mm to further discuss the impact of setup error margins used in proton therapy and the correlation with target size and location. Results: IMPT-ideal plans yield comparable plan quality as CK plans and slightly better OAR sparing while the IMPT-3mm plan results in a higher dose to the OARs, especially for centralized tumors. Comparing to the IMPT-ideal plans, a slightly larger 80% (Ravg=1.05) dose cloud and significantly larger 50% (Ravg=1.3) and 20% (Ravg=1.60) dose clouds are seen in CK plans. However, the 3mm expansion results in a larger high and medium dose clouds in IMPT-3mm plans (Ravg=0.65 for 80%-isodose; Ravg=0.93 for 50%-isodose). The trend increases with the size of the target and the distance from the brainstem to the center of target. Conclusion: Cyberknife is more preferable for treating centralized targets and proton therapy is advantageous for the large and peripheral targets. Advanced image guidance would improve the efficacy of proton therapy for intracranial treatments.},
doi = {10.1118/1.4956783},
journal = {Medical Physics},
number = 6,
volume = 43,
place = {United States},
year = {Wed Jun 15 00:00:00 EDT 2016},
month = {Wed Jun 15 00:00:00 EDT 2016}
}
  • Purpose: To evaluate mass density effects of CT conversion table and its variation in current treatment planning system of spot scanning proton beam using an IROC proton lung phantom for this study. Methods: A proton lung phantom study was acquired to Imaging and Radiation Oncology Core Houston (IROC) Quality Assurance Center. Inside the lung phantom, GAF Chromic films and couples of thermal luminescent dosimeter (TLD) capsules embedded in specified PTV and adjacent structures to monitor delivered dosage and 3D dose distribution profiles. Various material such as cork (Lung), blue water (heart), Techron HPV (ribs) and organic material of balsa woodmore » and cork as dosimetry inserts within phantom of solid water (soft tissue). Relative stopping power (RLSP) values were provided. Our treatment planning system (TPS) doesn’t require SP instead relative density was converted relative to water. However lung phantom was irradiated by planning with density override and the results were compared with IROC measurements. The second attempt was conducted without density override and compared with IROC’s. Results: The higher passing rate of imaging and measurement results of the lung phantom irradiation met the criteria by IROC without density override. The film at coronal plane was found to be shift due to inclined cylinder insertion. The converted CT density worked as expected to correlate relative stopping power. Conclusion: The proton lung phantom provided by IROC is a useful tool to qualify our commissioned proton pencil beam delivery with TPS within reliable confidence. The relative mass stopping power ratios of materials were converted from the relative physical density relative to water and the results were satisfied.« less
  • Purpose: Development of a parametric equation suitable for a daily use in routine clinic to provide estimates of stray neutron doses in proton therapy. Methods: Monte Carlo (MC) calculations using the UF-NCI 1-year-old phantom were exercised to determine the variation of stray neutron doses as a function of irradiation parameters while performing intracranial treatments. This was done by individually changing the proton beam energy, modulation width, collimator aperture and thickness, compensator thickness and the air gap size while their impact on neutron doses were put into a single equation. The variation of neutron doses with distance from the target volumemore » was also included in it. Then, a first step consisted in establishing the fitting coefficients by using 221 learning data which were neutron absorbed doses obtained with MC simulations while a second step consisted in validating the final equation. Results: The variation of stray neutron doses with irradiation parameters were fitted with linear, polynomial, etc. model while a power-law model was used to fit the variation of stray neutron doses with the distance from the target volume. The parametric equation fitted well MC simulations while establishing fitting coefficients as the discrepancies on the estimate of neutron absorbed doses were within 10%. The discrepancy can reach ∼25% for the bladder, the farthest organ from the target volume. Finally, the validation showed results in compliance with MC calculations since the discrepancies were also within 10% for head-and-neck and thoracic organs while they can reach ∼25%, again for pelvic organs. Conclusion: The parametric equation presents promising results and will be validated for other target sites as well as other facilities to go towards a universal method.« less
  • Purpose: Eclipse proton Monte Carlo AcurosPT 13.7 was commissioned and experimentally validated for an IBA dedicated PBS nozzle in water. Topas 1.3 was used to isolate the cause of differences in output and penumbra between simulation and experiment. Methods: The spot profiles were measured in air at five locations using Lynx. PTW-34070 Bragg peak chamber (Freiburg, Germany) was used to collect the relative integral Bragg peak for 15 proton energies from 100 MeV to 225 MeV. The phase space parameters (σx, σθ, ρxθ) number of protons per MU, energy spread and calculated mean energy provided by AcurosPT were identically implementedmore » into Topas. The absolute dose, profiles and field size factors measured using ionization chamber arrays were compared with both AcurosPT and Topas. Results: The beam spot size, σx, and the angular spread, σθ, in air were both energy-dependent: in particular, the spot size in air at isocentre ranged from 2.8 to 5.3 mm, and the angular spread ranged from 2.7 mrad to 6 mrad. The number of protons per MU increased from ∼9E7 at 100 MeV to ∼1.5E8 at 225 MeV. Both AcurosPT and TOPAS agree with experiment within 2 mm penumbra difference or 3% dose difference for scenarios including central axis depth dose and profiles at two depths in multi-spot square fields, from 40 to 200 mm, for all the investigated single-energy and multi-energy beams, indicating clinically acceptable source model and radiation transport algorithm in water. Conclusion: By comparing measured data and TOPAS simulation using the same source model, the AcurosPT 13.7 was validated in water within 2 mm penumbra difference or 3% dose difference. Benchmarks versus an independent Monte Carlo code are recommended to study the agreement in output, filed size factors and penumbra differences. This project is partially supported by the Varian grant under the master agreement between University of Pennsylvania and Varian.« less
  • Purpose: The main purpose of this study is to investigate the optimum oblique-beam arrangement for bilateral metallic prosthesis prostate cancer treatment in pencil beam scanning (PBS) proton therapy. Methods: A computed tomography dataset of bilateral metallic prosthesis prostate cancer case was selected for this retrospective study. A total of four beams (rightanterior- oblique [RAO], left-anterior-oblique [LAO], left-posterior-oblique [LPO], and right-posterior-oblique [RPO]) were selected for treatment planning. PBS plans were generated using multi-field-optimization technique for a total dose of 79.2 Gy[RBE] to be delivered in 44 fractions. Specifically, five different PBS plans were generated based on 2.5% ± 2 mm rangemore » uncertainty using five different beam arrangements (i)LAO+RAO+LPO+RPO, (ii)LAO+RAO, (iii)LPO+RPO, (iv)RAO+LPO, and (v)LAO+RPO. Each PBS plan was optimized by applying identical dose-volume constraints to the PTV, rectum, and bladder. Treatment plans were then compared based on the dose-volume histograms results. Results: The PTV coverage was found to be greater than 99% in all five plans. The homogeneity index (HI) was found to be almost identical (range, 0.03–0.04). The PTV mean dose was found to be comparable (range, 81.0–81.1 Gy[RBE]). For the rectum, the lowest mean dose (8.0 Gy[RBE]) and highest mean dose (31.1 Gy[RBE]) were found in RAO+LAO plan and LPO+RPO plan, respectively. LAO+RAO plan produced the most favorable dosimetric results of the rectum in the medium-dose region (V50) and high-dose region (V70). For the bladder, the lowest (5.0 Gy[RBE]) and highest mean dose (10.3 Gy[RBE]) were found in LPO+RPO plan and RAO+LAO plan, respectively. Other dosimetric results (V50 and V70) of the bladder were slightly better in LPO+RPO plan than in other plans. Conclusion: Dosimetric findings from this study suggest that two anterior-oblique proton beams arrangement (LAO+RAO) is a more favorable option with the possibility of reducing rectal dose significantly while maintaining comparable target coverage and acceptable bladder dose.« less
  • Purpose: To study if abdominal compression can reduce breathing motion and mitigate interplay effect in pencil beam scanning proton therapy (PBSPT) treatment of liver tumors in order to better spare healthy liver volumes compared with photon therapy. Methods: Ten patients, six having large tumors initially treated with IMRT and four having small tumors treated with SBRT, were replanned for PBSPT. ITV and beam-specific PTVs based on 4D-CT were used to ensure target coverage in PBSPT. The use of an abdominal compression belt and volumetric repainting was investigated to mitigate the interplay effect between breathing motion and PBSPT dynamic delivery. Anmore » in-house Matlab script has been developed to simulate this interplay effect. The dose is computed on each phase individually by sorting all spots according to their simulated delivery timing. The final dose distribution is then obtained by accumulating all dose maps to a reference phase. Results: For equivalent target coverage PBSPT reduced average healthy liver dose by 9.5% of the prescription dose compared with IMRT/SBRT. Abdominal compression of 113.2±42.2 mmHg was effective for all 10 patients and reduced average motion by 2.25 mm. As a result, the average ITV volume decreased from 128.2% to 123.1% of CTV volume. Similarly, the average beam-specific PTV volume decreased from 193.2% to 183.3%. For 8 of the 10 patients, the average motion was reduced below 5 mm, and up to 3 repainting were sufficient to mitigate interplay. For the other two patients with larger residual motion, 4–5 repainting were needed. Conclusion: We recommend evaluation of the 4DCT motion histogram following simulation and the interplay effect following treatment planning in order to personalize the use of compression and volumetric repainting for each patient. Abdominal compression enables safe and more effective PBS treatment of liver tumors by reduction of motion and interplay effect. Kevin Souris is supported by IBA and Televie Grant from F.R.S.-FNRS. Liyong Lin is partially supported by Varian.« less