Title: SU-F-T-504: Non-Divergent Planning Method for Craniospinal Irradiation

Abstract

Purpose: Traditional Craniospinal Irradiation (CSI) planning techniques require careful field placement to allow optimal divergence and field overlap at depth, and measurement of skin gap. The result of this is a necessary field overlap resulting in dose heterogeneity in the spinal canal. A novel, nondivergent field matching method has been developed to allow simple treatment planning and delivery without the need to measure skin gap. Methods: The CSI patient was simulated in the prone, and a plan was developed. Bilateral cranial fields were designed with couch and collimator rotation to eliminate divergence with the upper spine field and minimize anterior divergence into the lenses. Spinal posterior-to-anterior fields were designed with the couch rotated to 90 degrees to allow gantry rotation to eliminate divergence at the match line, and the collimator rotated to 90 degrees to allow appropriate field blocking with the MLCs. A match line for the two spinal fields was placed and the gantry rotated to equal angles in opposite directions about the match line. Jaw positions were then defined to allow 1mm overlap at the match line to avoid cold spots. A traditional CSI plan was generated using diverging spinal fields, and a comparison between the two techniquesmore » was generated. Results: The non-divergent treatment plan was able to deliver a highly uniform dose to the spinal cord with a cold spot of only 95% and maximum point dose of 115.8%, as compared to traditional plan cold spots of 87% and hot spots of 132% of the prescription dose. Conclusion: A non-divergent method for planning CSI patients has been developed and clinically implemented. Planning requires some geometric manipulation in order to achieve an adequate dose distribution, however, it can help to manage cold spots and simplify the shifts needed between spinal fields.« less

Authors:
; ;  [1]
  1. University of Toledo Medical Center, Toledo, OH (United States)
Publication Date:
OSTI Identifier:
22649091
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 43; Journal Issue: 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; HOT SPOTS; IRRADIATION; PLANNING; RADIATION DOSE DISTRIBUTIONS; SPINAL CORD; VERTEBRAE

Citation Formats

Sperling, N, Bogue, J, and Parsai, E. SU-F-T-504: Non-Divergent Planning Method for Craniospinal Irradiation. United States: N. p., 2016. Web. doi:10.1118/1.4956689.
Sperling, N, Bogue, J, & Parsai, E. SU-F-T-504: Non-Divergent Planning Method for Craniospinal Irradiation. United States. doi:10.1118/1.4956689.
Sperling, N, Bogue, J, and Parsai, E. 2016. "SU-F-T-504: Non-Divergent Planning Method for Craniospinal Irradiation". United States. doi:10.1118/1.4956689.
@article{osti_22649091,
title = {SU-F-T-504: Non-Divergent Planning Method for Craniospinal Irradiation},
author = {Sperling, N and Bogue, J and Parsai, E},
abstractNote = {Purpose: Traditional Craniospinal Irradiation (CSI) planning techniques require careful field placement to allow optimal divergence and field overlap at depth, and measurement of skin gap. The result of this is a necessary field overlap resulting in dose heterogeneity in the spinal canal. A novel, nondivergent field matching method has been developed to allow simple treatment planning and delivery without the need to measure skin gap. Methods: The CSI patient was simulated in the prone, and a plan was developed. Bilateral cranial fields were designed with couch and collimator rotation to eliminate divergence with the upper spine field and minimize anterior divergence into the lenses. Spinal posterior-to-anterior fields were designed with the couch rotated to 90 degrees to allow gantry rotation to eliminate divergence at the match line, and the collimator rotated to 90 degrees to allow appropriate field blocking with the MLCs. A match line for the two spinal fields was placed and the gantry rotated to equal angles in opposite directions about the match line. Jaw positions were then defined to allow 1mm overlap at the match line to avoid cold spots. A traditional CSI plan was generated using diverging spinal fields, and a comparison between the two techniques was generated. Results: The non-divergent treatment plan was able to deliver a highly uniform dose to the spinal cord with a cold spot of only 95% and maximum point dose of 115.8%, as compared to traditional plan cold spots of 87% and hot spots of 132% of the prescription dose. Conclusion: A non-divergent method for planning CSI patients has been developed and clinically implemented. Planning requires some geometric manipulation in order to achieve an adequate dose distribution, however, it can help to manage cold spots and simplify the shifts needed between spinal fields.},
doi = {10.1118/1.4956689},
journal = {Medical Physics},
number = 6,
volume = 43,
place = {United States},
year = 2016,
month = 6
}
  • Purpose: To develop a treatment planning technique that achieves optimal robustness against systematic position and range uncertainties, and interfield position errors for craniospinal irradiation (CSI) using spot scanning proton radiotherapy. Methods: Eighteen CSI patients who had previously been treated using photon radiation were used for this study. Eight patients were less than 10 years old. The prescription dose was 23.4Gy in 1.8Gy fractions. Two different field arrangement types were investigated: 1 posterior field per isocenter and 2 posterior oblique fields per isocenter. For each field type, two delivery configurations were used: 5cm bolus attached to the treatment table and amore » 4.5cm range shifter located inside the nozzle. The target for each plan was the whole brain and thecal sac. For children under the age of 10, all plan types were repeated with an additional dose of 21Gy prescribed to the vertebral bodies. Treatment fields were matched by stepping down the dose in 10% increments over 9cm. Robustness against 3% and 3mm uncertainties, as well as a 3mm inter-field error was analyzed. Dose coverage of the target and critical structure sparing for each plan type will be considered. Ease of planning and treatment delivery was also considered for each plan type. Results: The mean dose volume histograms show that the bolus plan with posterior beams gave the best overall plan, and all proton plans were comparable to or better than the photon plans. The plan type that was the most robust against the imposed uncertainties was also the bolus plan with posterior beams. This is also the plan configuration that is the easiest to deliver and plan. Conclusion: The bolus plan with posterior beams achieved optimal robustness against systematic position and range uncertainties, as well as inter-field position errors.« less
  • Purpose: The purpose of this study is to investigate a volumetric modulated arc therapy (VMAT) treatment planning technique for supine craniospinal irradiation (CSI). Evaluate the suitability of VMAT for CSI with dosimetric measurements and compare it to 3D conformal planning using specific plan metrics such as dose conformity, homogeneity, and dose of organs at risk (OAR). Methods: Ten CSI patients treated with conventional 3D technique were re-planned with VMAT. The PTV was contoured to include the whole contents of the brain and spinal canal with a uniform margin of 5 mm. VMAT plans were generated with two partial arcs coveringmore » the brain, two partial arcs for the superior portion of the spinal cord and two partial arcs covering the remaining inferior portion of the spinal cord. Conformity index (CI), heterogeneity indexes (HI) and max and mean doses of OAR were compared to 3D plans. VMAT plans were delivered onto an anthropomorphic phantom loaded with Gafchromic films and OSLDs placed at specific positions to evaluate the plan dose at the junctions and as well as the plan dose distributions. Results: This VMAT technique was validated with a clinical study of 10 patients. The average CI was 1.03±0.02 for VMAT plans and 1.96±0.32 for conformal plans. And the average HI was 1.15±0.01 for VMAT plans and 1.51±0.21 for conformal plans. The mean and max doses to the all OARs for VMAT plans were significantly lower than conformal plans. The measured dose in phantom for VAMT plans was comparable to the calculated dose in Eclipse and the doses at junctions were verified. Conclusion: VMAT CSI was able to achieve better dose conformity and heterogeneity as well as significantly reducing the dose to Heart, esophagus and larynx. VMAT CSI appears to be a dosimterically advantageous, faster delivery, has better reproducibility CSI treatment.« less
  • Purpose: To develop and compare a volumetric modulated arc therapy (VMAT) technique with conventional radiotherapy for craniospinal irradiation with respect to improved dose conformity and homogeneity in the planning target volume (PTV) and to reduced dose to organs at risk (OAR). Methods and Materials: Conventional craniospinal axis radiotherapy plans of 5 patients were acquired. The median (range) length of the PTV was 58.9 (48.1-83.7) cm. The 6-MV VMAT plans were inversely planned with one isocenter near the base of the brain and the minimum number of isocenters required for the specified lengths of spine. The plans were optimized with highmore » weighting for PTV coverage and low weighting for OAR sparing. Conformity and heterogeneity indices, dose-volume histograms, mean doses, and non-PTV integral doses from the two plans (prescription dose 23.4 Gy in 13 fractions) were compared. Results: The median (range) conformity index of VMAT was 1.22 (1.09-1.45), compared with 1.69 (1.44-2.67) for conventional plans (p = 0.04). The median (range) heterogeneity index was also lower for VMAT compared with conventional plans: 1.04 (1.03-1.07) vs. 1.12 (1.09-1.19), respectively (p = 0.04). A significant reduction of mean and maximum doses was observed in the heart, thyroid, esophagus, optic nerves, and eyes with VMAT when compared with conventional plans. A decrease in body V{sub 10Gy} was observed, but for 4 of 5 patients non-PTV integral dose was increased with VMAT when compared with the conventional plans. Conclusions: A VMAT technique to treat the craniospinal axis significantly reduces OAR dose, potentially leading to lower late organ toxicity. However, this is achieved at the expense of increased low-dose volumes, which is inherent to the technique, carrying a potentially increased risk of secondary malignancies.« less
  • Purpose: To describe how we moved to VMAT in the craniospinal radiotherapy planning process, the actual procedure details, and the results for the patients treated. Methods: Twelve patients underwent craniospinal irradiation with the new procedure, based on the paper by Lee et al. (IJROBP 82, 2012), with some additional modifications. Patients were treated in supine position in Varian Clinac iX linacs with 6 MV RapidArc; prescription doses ranged from 23.4 to 40 Gy (13 to 20 fractions); depending on the PTV length, 2 or 3 isocenters were used, all coordinates being equal except the longitudinal one, setting a few centimeter-longmore » overlapping region; 2 arcs (RA) sharing isocentre for the cranial region, RA1 encompassing cranium and superior spinal region, and RA2 intended to improve conformity, only for cranium; for spine, 1 or 2 isocenters were employed; optimization was performed with Eclipse (V 13.0) using AAA algorithm, establishing sets of optimization parameters to give high conformity while sparing OAR. In pediatric patients, homogeneous irradiation of the vertebrae was also required.Conformity (CI) and heterogeneity (HI) indices (same as Lee et al.), and mean and maximum doses for OAR were calculated. Several pre-treatment verification methods were used: Octavius4D (PTW) for each isocentre, point dose at the junction region, Portal Dosimetry (when possible), and independent MU verification software (Diamond, PTW). Results: CI median value was 1.02 (0.99–1.07) and HI, 1.07 (1.06–1.09); a great reduction was observed for CI and OAR mean doses with respect to Lee et al. data; median maximum eye lens dose was 7.3 Gy (4.0–12.0); mean LungV20Gy was 1.9%; in children, vertebrae were homogeneously irradiated (D95%=20.8 Gy, Dmean= 23.2 Gy).All pre-treatment verifications were found within our action levels except for Portal Dosimetry. Conclusion: A RapidArc planning process for craniospinal axis irradiation has been implemented with significant advantages on conformity, homogeneity, feasibility and efficiency. and increase brain tissue sparing. Variations in volume decrease may be related to shape or location of the tumor.« less
  • Purpose: Limiting the neurocognitive sequelae of radiotherapy (RT) has been an objective in the treatment of medulloblastoma. Conformal RT to less than the entire posterior fossa (PF) after craniospinal irradiation might reduce neurocognitive sequelae and requires evaluation. Methods and Materials: Between October 1996 and August 2003, 86 patients, 3-21 years of age, with newly diagnosed, average-risk medulloblastoma were treated in a prospective, institutional review board-approved, multi-institution trial of risk-adapted RT and dose-intensive chemotherapy. RT began within 28 days of definitive surgery and consisted of craniospinal irradiation (23.4 Gy), conformal PF RT (36.0 Gy), and primary site RT (55.8 Gy). Themore » planning target volume for the primary site included the postoperative tumor bed surrounded by an anatomically confined margin of 2 cm that was then expanded with a geometric margin of 0.3-0.5 cm. Chemotherapy was initiated 6 weeks after RT and included four cycles of high-dose cyclophosphamide, cisplatin, and vincristine. Results: At a median follow-up of 61.2 months (range, 5.2-115.0 months), the estimated 5-year event-free survival and cumulative incidence of PF failure rate was 83.0% {+-} 5.3% and 4.9% {+-} 2.4% ({+-} standard error), respectively. The targeting guidelines used in this study resulted in a mean reduction of 13% in the volume of the PF receiving doses >55 Gy compared with conventionally planned RT. The reductions in the dose to the temporal lobes, cochleae, and hypothalamus were statistically significant. Conclusion: This prospective trial has demonstrated that irradiation of less than the entire PF after 23.4 Gy craniospinal irradiation for average-risk medulloblastoma results in disease control comparable to that after treatment of the entire PF.« less