Title: SU-F-T-496: An Investigation of Two Novel Devices for Testing Linac Clearance During CT Simulation

Abstract

Purpose: This work’s objective is to determine the efficacy of two newly patented devices termed the Mor-O rings, Mark1 and Mark2, developed to predict collisions between the patient and a linac at the time of CT simulation. Methods: Mark1: A ring with an inner radius equal to the distance between the isocenter and the nearest portion of a linac head(diso) was made. This is mounted to a stand that allows vertical repositioning. The ring is placed around the patient on the CT table and aligned with isocenter. The patient is moved through the ring. If the ring touches the patient, a collision is predicted. To test this device, predicted collisions were marked on a phantom. The phantom was then repositioned on the linac table where the collisions were verified. Mark2: An arc with the radius diso was created with a re-locatable half-linac head wire-frame. The Mark2 is positioned in the same way as the Mark1 but can additionally mimic couch and gantry angulations. The Mark2 was tested with a volunteer using multiple couch, gantry and isocenter positions. The volunteer was then repositioned on the linac table to verify the angles of collisions. Results: Mark1: One isocenter out of ten showedmore » negative clearance (0.9mm) on the linac table. All other collisions were predicted with 1 to 36mm of additional clearance. Mark2: All collisions were prevented with an additional 1.3 to 14.8 degrees of clearance. Conclusion: The Mark1 prototype is able to predict all collisions when no couch angulation is used. This device takes less than a minute to setup and is simple to use. The Mark2, when testing beam geometries used for noncoplanar SBRT, was able to prevent all collisions with 1.3 to 14.8 degrees additional clearance. Improvements in construction for both devices could increase accuracy and usability. Andrew Morrow owns Morrow Physics, LLC and Brian Massingill owns Spur Physics, LLC. We are both listed on the patent for the devices investigated in this work (patent 9220922). Our current employer, Baylor Scott and White, is also on the patent.« less

Authors:
;  [1]
  1. Scott & White Hospital, Temple, TX (United States)
Publication Date:
OSTI Identifier:
22649083
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 43; Journal Issue: 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; 62 RADIOLOGY AND NUCLEAR MEDICINE; 61 RADIATION PROTECTION AND DOSIMETRY; CLEARANCE; COLLISIONS; COMPUTERIZED TOMOGRAPHY; EQUIPMENT; LINEAR ACCELERATORS; PATIENTS; RADIATION DOSES; SIMULATION; TESTING

Citation Formats

Morrow, A, and Massingill, B. SU-F-T-496: An Investigation of Two Novel Devices for Testing Linac Clearance During CT Simulation. United States: N. p., 2016. Web. doi:10.1118/1.4956681.
Morrow, A, & Massingill, B. SU-F-T-496: An Investigation of Two Novel Devices for Testing Linac Clearance During CT Simulation. United States. doi:10.1118/1.4956681.
Morrow, A, and Massingill, B. 2016. "SU-F-T-496: An Investigation of Two Novel Devices for Testing Linac Clearance During CT Simulation". United States. doi:10.1118/1.4956681.
@article{osti_22649083,
title = {SU-F-T-496: An Investigation of Two Novel Devices for Testing Linac Clearance During CT Simulation},
author = {Morrow, A and Massingill, B},
abstractNote = {Purpose: This work’s objective is to determine the efficacy of two newly patented devices termed the Mor-O rings, Mark1 and Mark2, developed to predict collisions between the patient and a linac at the time of CT simulation. Methods: Mark1: A ring with an inner radius equal to the distance between the isocenter and the nearest portion of a linac head(diso) was made. This is mounted to a stand that allows vertical repositioning. The ring is placed around the patient on the CT table and aligned with isocenter. The patient is moved through the ring. If the ring touches the patient, a collision is predicted. To test this device, predicted collisions were marked on a phantom. The phantom was then repositioned on the linac table where the collisions were verified. Mark2: An arc with the radius diso was created with a re-locatable half-linac head wire-frame. The Mark2 is positioned in the same way as the Mark1 but can additionally mimic couch and gantry angulations. The Mark2 was tested with a volunteer using multiple couch, gantry and isocenter positions. The volunteer was then repositioned on the linac table to verify the angles of collisions. Results: Mark1: One isocenter out of ten showed negative clearance (0.9mm) on the linac table. All other collisions were predicted with 1 to 36mm of additional clearance. Mark2: All collisions were prevented with an additional 1.3 to 14.8 degrees of clearance. Conclusion: The Mark1 prototype is able to predict all collisions when no couch angulation is used. This device takes less than a minute to setup and is simple to use. The Mark2, when testing beam geometries used for noncoplanar SBRT, was able to prevent all collisions with 1.3 to 14.8 degrees additional clearance. Improvements in construction for both devices could increase accuracy and usability. Andrew Morrow owns Morrow Physics, LLC and Brian Massingill owns Spur Physics, LLC. We are both listed on the patent for the devices investigated in this work (patent 9220922). Our current employer, Baylor Scott and White, is also on the patent.},
doi = {10.1118/1.4956681},
journal = {Medical Physics},
number = 6,
volume = 43,
place = {United States},
year = 2016,
month = 6
}
  • Purpose: MRI-guided radiation therapy systems (MRIgRT) are being developed to improve online imaging during treatment delivery. At present, the operation of single point dosimeters and an ionization chamber array have been characterized in such systems. This work investigates a novel 2D diode array, named “magic plate,” for both single point calibration and 2D positional performance, the latter being a key element of modern radiotherapy techniques that will be delivered by these systems. Methods: GEANT4 Monte Carlo methods have been employed to study the dose response of a silicon diode array to 6 MV photon beams, in the presence of in-linemore » and perpendicularly aligned uniform magnetic fields. The array consists of 121 silicon diodes (dimensions 1.5 × 1.5 × 0.38 mm{sup 3}) embedded in kapton substrate with 1 cm pitch, spanning a 10 × 10 cm{sup 2} area in total. A geometrically identical, water equivalent volume was simulated concurrently for comparison. The dose response of the silicon diode array was assessed for various photon beam field shapes and sizes, including an IMRT field, at 1 T. The dose response was further investigated at larger magnetic field strengths (1.5 and 3 T) for a 4 × 4 cm{sup 2} photon field size. Results: The magic plate diode array shows excellent correspondence (< ± 1%) to water dose in the in-line orientation, for all beam arrangements and magnetic field strengths investigated. The perpendicular orientation, however, exhibits a dose shift with respect to water at the high-dose-gradient beam edge of jaw-defined fields [maximum (4.3 ± 0.8)% over-response, maximum (1.8 ± 0.8)% under-response on opposing side for 1 T, uncertainty 1σ]. The trend is not evident in areas with in-field dose gradients typical of IMRT dose maps. Conclusions: A novel 121 pixel silicon diode array detector has been characterized by Monte Carlo simulation for its performance inside magnetic fields representative of current prototype and proposed MRI–linear accelerator systems. In the in-line orientation, the silicon dose is directly proportional to the water dose. In the perpendicular orientation, there is a shift in dose response relative to water in the highest dose gradient regions, at the edge of jaw-defined and single-segment MLC fields. The trend was not observed in-field for an IMRT beam. The array is expected to be a valuable tool in MRIgRT dosimetry.« less
  • An economical and reliable method for protecting steam conduit supports and suspension devices from overloads is discussed. The system provides increases in carrying capacity of the springs in the supports and suspension. (JMT)
  • A heavy ion beam diagnostic system installed at the Brookhaven Single Event Upset Test Facility is described. Calibration of the system with the help of {alpha}-particles, essential for linear energy transfer (LET) measurements, is discussed. Measured LET values for 20 different ions, including {sup 7}Li, {sup 9}B, {sup 12}C, {sup 16}O, {sup 19}F, {sup 28}Si, {sup 32}S, {sup 35}Cl, {sup 40}Ca, {sup 45}Sc, {sup 48}Ti, {sup 56}Fe, {sup 58}Ni, {sup 63}Cu, {sup 74}Ge, {sup 79}Br, {sup 107}Ag, {sup 127}I, {sup 197}Au, and {sup 235}U, with energies between 0.5 and 8.5 MeV/AMU but not exceeding 400 MeV for the heaviest ions,more » are presented in both graphical and numerical forms. Results are compared to predictions of the TRIM-90 simulation program, with an average difference between the measured and calculated values of 2 {+-} 6%.« less
  • A computer model in CST Studio Suite has been developed to evaluate several novel geometrically enhanced photocathode designs. This work was aimed at identifying a structure that would increase the total electron yield by a factor of two or greater in the 1–30 keV range. The modeling software was used to simulate the electric field and generate particle tracking for several potential structures. The final photocathode structure has been tailored to meet a set of detector performance requirements, namely, a spatial resolution of <40 μm and a temporal spread of 1–10 ps. As a result, we present the details ofmore » the geometrically enhanced photocathode model and resulting static field and electron emission characteristics.« less