skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: SU-F-T-250: What Does It Take to Correctly Assess the High Failure Modes of an Advanced Radiotherapy Procedure Such as Stereotactic Body Radiation Therapy?

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4956390· OSTI ID:22648866
; ; ;  [1]
  1. Virginia Commonwealth University, Richmond, VA (United States)

Purpose: Assess the correct implementation of risk-based methodology of TG 100 to optimize quality management and patient safety procedures for Stereotactic Body Radiation Therapy. Methods: A detailed process map of SBRT treatment procedure was generated by a team of three physicists with varying clinical experience at our institution to assess the potential high-risk failure modes. The probabilities of occurrence (O), severity (S) and detectability (D) for potential failure mode in each step of the process map were assigned by these individuals independently on the scale from1 to 10. The risk priority numbers (RPN) were computed and analyzed. The highest 30 potential modes from each physicist’s analysis were then compared. Results: The RPN values assessed by the three physicists ranged from 30 to 300. The magnitudes of the RPN values from each physicist were different, and there was no concordance in the highest RPN values recorded by three physicists independently. The 10 highest RPN values belonged to sub steps of CT simulation, contouring and delivery in the SBRT process map. For these 10 highest RPN values, at least two physicists, irrespective of their length of experience had concordance but no general conclusions emerged. Conclusion: This study clearly shows that the risk-based assessment of a clinical process map requires great deal of preparation, group discussions, and participation by all stakeholders. One group albeit physicists cannot effectively implement risk-based methodology proposed by TG100. It should be a team effort in which the physicists can certainly play the leading role. This also corroborates TG100 recommendation that risk-based assessment of clinical processes is a multidisciplinary team effort.

OSTI ID:
22648866
Journal Information:
Medical Physics, Vol. 43, Issue 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English