skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: SU-F-T-57: Delivered Activity Accuracy of Radium 223 Dichloride Injections, When Being Administrated for Castration Resistant Prostate Cancer, Symptomatic Bone Metastases. The Impact of Residual Activity in the Spent Syringe and Dispensing Accuracy of Ra 223

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4956192· OSTI ID:22642305
 [1]
  1. Marsden Medical Physics Associates, Denville, NJ (United States)

Purpose: To quantify the delivered activity accuracy of Radium 223 dichloride injections, when being administrated for castration – resistant prostate cancer, symptomatic bone metastases. The impact of residual activity in the spent syringe and dispensing accuracy of Ra 223. Methods: The administration is by slow intravenous injection over 1 minute followed by double flushing of the 10 mL syringe and IV with saline. Eighty (80) procedures was used to investigate variations in the activity from the amount prescribed (µCi) = 1.35 × Patient weight Kg. The Activity dispensed into a 10mL syringe using a NIST traceable Capintec CRC-25R Chamber and a cross calibrated capintec CRC-15R to measure activity in the syringe immediately before and after administration Results: The patients weight range from 121Ib to 235lb and doses ranging 74.25 µCi to 144.2 µCi. The deviation of dispensed dose vs Prescribed dose average +2.1% with a range of −1.1% to +5.7%. The Dose measured before administration ranges 79.3 µCi to 154.9 µCi. Deviation from the dispensed dose was show to average +2.9% with a range of −0.8% to +7.3%. The average residual dose post injection was 2.5 µCi or 2.2% of the pre injection activity. Ranging from 0.9 µCi to 6.2 µCi, 0.7% to 5.4% respectively. Subtracting the residual activity from that measured activity before injection and comparing it to prescription dose was shown to have an average variation of +2.7% with a range of −0.8% to 7.4%. Conclusion: The case resulted in the 6.2 µCi maximum residual dose had two syringes. A small, 82.8 µCi activity, case resulted in the 7.4% maximum variation in measures less residual verses prescription dose. The average +2.1 % dispenses activity of Ra 223 over the prescription dosage was seen to counteract the average 2.2% residual dosage found to remain in the syringe.

OSTI ID:
22642305
Journal Information:
Medical Physics, Vol. 43, Issue 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English