Title: SU-F-J-217: Accurate Dose Volume Parameters Calculation for Revealing Rectum Dose-Toxicity Effect Using Deformable Registration in Cervical Cancer Brachytherapy: A Pilot Study

Abstract

Purpose: To study the feasibility of employing deformable registration methods for accurate rectum dose volume parameters calculation and their potentials in revealing rectum dose-toxicity between complication and non-complication cervical cancer patients with brachytherapy treatment. Method and Materials: Data from 60 patients treated with BT including planning images, treatment plans, and follow-up clinical exam were retrospectively collected. Among them, 12 patients complained about hematochezia were further examined with colonoscopy and scored as Grade 1–3 complication (CP). Meanwhile, another 12 non-complication (NCP) patients were selected as a reference group. To seek for potential gains in rectum toxicity prediction when fractional anatomical deformations are account for, the rectum dose volume parameters D0.1/1/2cc of the selected patients were retrospectively computed by three different approaches: the simple “worstcase scenario” (WS) addition method, an intensity-based deformable image registration (DIR) algorithm-Demons, and a more accurate, recent developed local topology preserved non-rigid point matching algorithm (TOP). Statistical significance of the differences between rectum doses of the CP group and the NCP group were tested by a two-tailed t-test and results were considered to be statistically significant if p < 0.05. Results: For the D0.1cc, no statistical differences are found between the CP and NCP group in all threemore » methods. For the D1cc, dose difference is not detected by the WS method, however, statistical differences between the two groups are observed by both Demons and TOP, and more evident in TOP. For the D2cc, the CP and NCP cases are statistically significance of the difference for all three methods but more pronounced with TOP. Conclusion: In this study, we calculated the rectum D0.1/1/2cc by simple WS addition and two DIR methods and seek for gains in rectum toxicity prediction. The results favor the claim that accurate dose deformation and summation tend to be more sensitive in unveiling the dose-toxicity relationship. This work is supported in part by grant from VARIAN MEDICAL SYSTEMS INC, the National Natural Science Foundation of China (no 81428019 and no 81301940), the Guangdong Natural Science Foundation (2015A030313302)and the 2015 Pearl River S&T Nova Program of Guangzhou (201506010096).« less

Authors:
; ; ;  [1]; ; ;  [2]
  1. Southern Medical University, Guangzhou, Guangdong (China)
  2. UT Southwestern Medical Center, Dallas, TX (United States)
Publication Date:
OSTI Identifier:
22642244
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 43; Journal Issue: 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; 61 RADIATION PROTECTION AND DOSIMETRY; ALGORITHMS; BRACHYTHERAPY; IMAGES; NEOPLASMS; PATIENTS; RADIATION DOSES; RECTUM; TOPOLOGY; TOXICITY

Citation Formats

Zhen, X, Chen, H, Liao, Y, Zhou, L, Hrycushko, B, Albuquerque, K, and Gu, X. SU-F-J-217: Accurate Dose Volume Parameters Calculation for Revealing Rectum Dose-Toxicity Effect Using Deformable Registration in Cervical Cancer Brachytherapy: A Pilot Study. United States: N. p., 2016. Web. doi:10.1118/1.4956125.
Zhen, X, Chen, H, Liao, Y, Zhou, L, Hrycushko, B, Albuquerque, K, & Gu, X. SU-F-J-217: Accurate Dose Volume Parameters Calculation for Revealing Rectum Dose-Toxicity Effect Using Deformable Registration in Cervical Cancer Brachytherapy: A Pilot Study. United States. doi:10.1118/1.4956125.
Zhen, X, Chen, H, Liao, Y, Zhou, L, Hrycushko, B, Albuquerque, K, and Gu, X. Wed . "SU-F-J-217: Accurate Dose Volume Parameters Calculation for Revealing Rectum Dose-Toxicity Effect Using Deformable Registration in Cervical Cancer Brachytherapy: A Pilot Study". United States. doi:10.1118/1.4956125.
@article{osti_22642244,
title = {SU-F-J-217: Accurate Dose Volume Parameters Calculation for Revealing Rectum Dose-Toxicity Effect Using Deformable Registration in Cervical Cancer Brachytherapy: A Pilot Study},
author = {Zhen, X and Chen, H and Liao, Y and Zhou, L and Hrycushko, B and Albuquerque, K and Gu, X},
abstractNote = {Purpose: To study the feasibility of employing deformable registration methods for accurate rectum dose volume parameters calculation and their potentials in revealing rectum dose-toxicity between complication and non-complication cervical cancer patients with brachytherapy treatment. Method and Materials: Data from 60 patients treated with BT including planning images, treatment plans, and follow-up clinical exam were retrospectively collected. Among them, 12 patients complained about hematochezia were further examined with colonoscopy and scored as Grade 1–3 complication (CP). Meanwhile, another 12 non-complication (NCP) patients were selected as a reference group. To seek for potential gains in rectum toxicity prediction when fractional anatomical deformations are account for, the rectum dose volume parameters D0.1/1/2cc of the selected patients were retrospectively computed by three different approaches: the simple “worstcase scenario” (WS) addition method, an intensity-based deformable image registration (DIR) algorithm-Demons, and a more accurate, recent developed local topology preserved non-rigid point matching algorithm (TOP). Statistical significance of the differences between rectum doses of the CP group and the NCP group were tested by a two-tailed t-test and results were considered to be statistically significant if p < 0.05. Results: For the D0.1cc, no statistical differences are found between the CP and NCP group in all three methods. For the D1cc, dose difference is not detected by the WS method, however, statistical differences between the two groups are observed by both Demons and TOP, and more evident in TOP. For the D2cc, the CP and NCP cases are statistically significance of the difference for all three methods but more pronounced with TOP. Conclusion: In this study, we calculated the rectum D0.1/1/2cc by simple WS addition and two DIR methods and seek for gains in rectum toxicity prediction. The results favor the claim that accurate dose deformation and summation tend to be more sensitive in unveiling the dose-toxicity relationship. This work is supported in part by grant from VARIAN MEDICAL SYSTEMS INC, the National Natural Science Foundation of China (no 81428019 and no 81301940), the Guangdong Natural Science Foundation (2015A030313302)and the 2015 Pearl River S&T Nova Program of Guangzhou (201506010096).},
doi = {10.1118/1.4956125},
journal = {Medical Physics},
number = 6,
volume = 43,
place = {United States},
year = {Wed Jun 15 00:00:00 EDT 2016},
month = {Wed Jun 15 00:00:00 EDT 2016}
}
  • Purpose: To compare the performance of two deformable image registration (DIR) algorithms for contour propagation and to evaluate the accuracy of DIR for use with high dose rate (HDR) brachytherapy planning for cervical cancer. Methods: Five patients undergoing HDR ring and tandem brachytherapy were included in this retrospective study. All patients underwent CT simulation and replanning prior to each fraction (3–5 fractions total). CT-to-CT DIR was performed using two commercially available software platforms: SmartAdapt, Varian Medical Systems (Demons) and Velocity AI, Velocity Medical Solutions (B-spline). Fraction 1 contours were deformed and propagated to each subsequent image set and compared tomore » contours manually drawn by an expert clinician. Dice similarity coefficients (DSC), defined as, DSC(A,B)=2(AandB)/(A+B) were calculated to quantify spatial overlap between manual (A) and deformed (B) contours. Additionally, clinician-assigned visual scores were used to describe and compare the performance of each DIR method and ultimately evaluate which was more clinically acceptable. Scoring was based on a 1–5 scale—with 1 meaning, “clinically acceptable with no contour changes” and 5 meaning, “clinically unacceptable”. Results: Statistically significant differences were not observed between the two DIR algorithms. The average DSC for the bladder, rectum and rectosigmoid were 0.82±0.08, 0.67±0.13 and 0.48±0.18, respectively. The poorest contour agreement was observed for the rectosigmoid due to limited soft tissue contrast and drastic anatomical changes, i.e., organ shape/filling. Two clinicians gave nearly equivalent average scores of 2.75±0.91 for SmartAdapt and 2.75±0.94 for Velocity AI—indicating that for a majority of the cases, more than one of the three contours evaluated required major modifications. Conclusion: Limitations of both DIR algorithms resulted in inaccuracies in contour propagation in the pelvic region, thus hampering the clinical utility of this technology. Further work is required to optimize these algorithms and take advantage of the potential of DIR for HDR brachytherapy planning.« less
  • Purpose: We introduce a new method combined with the deformable image registration (DIR) and regions-of-interest mapping (ROIM) technique to accurately calculate dose on daily CBCT for esophageal cancer. Methods: Patients suffered from esophageal cancer were enrolled in the study. Prescription was set to 66 Gy/30 F and 54 Gy/30 F to the primary tumor (PTV66) and subclinical disease (PTV54) . Planning CT (pCT) were segmented into 8 substructures in terms of their differences in physical density, such as gross target volume (GTV), venae cava superior (SVC), aorta, heart, spinal cord, lung, muscle and bones. The pCT and its substructures weremore » transferred to the MIM software to readout their mean HU values. Afterwards, a deformable planning CT to daily KV-CBCT image registration method was then utilized to acquire a new structure set on CBCT. The newly generated structures on CBCT were then transferred back to the treatment planning system (TPS) and its HU information were overridden manually with mean HU values obtained from pCT. Finally, the treatment plan was projected onto the CBCT images with the same beam arrangements and monitor units (MUs) to accomplish dose calculation. Planning target volume (PTV) and organs at risk (OARs) from both of the pCT and CBCT were compared to evaluate the dose calculation accuracy. Results: It was found that the dose distribution in the CBCT showed little differences compared to the pCT, regardless of whether PTV or OARs were concerned. Specifically, dose variation in GTV, PTV54, PTV66, SVC, lung and heart were within 0.1%. The maximum dose variation was presented in the spinal cord, which was up to 2.7% dose difference. Conclusion: The proposed method combined with DIR and ROIM technique to accurately calculate dose distribution on CBCT for esophageal cancer is feasible.« less
  • Purpose: PET/CT provides important functional information for radiotherapy targeting of cervical cancer. However, repeated PET/CT procedures for external beam and subsequent brachytherapy expose patients to additional radiation and are not cost effective. Our goal is to investigate the possibility of propagating PET-active volumes for brachytherapy procedures through deformable image registration (DIR) of earlier PET/CT and ultimately to minimize the number of PET/CT image sessions required. Methods: Nine cervical cancer patients each received their brachytherapy preplanning PET/CT at the end of EBRT with a Syed template in place. The planning PET/CT was acquired on the day of brachytherapy treatment with themore » actual applicator (Syed or Tandem and Ring) and rigidly registered. The PET/CT images were then deformably registered creating a third (deformed) image set for target prediction. Regions of interest with standardized uptake values (SUV) greater than 65% of maximum SUV were contoured as target volumes in all three sets of PET images. The predictive value of the registered images was evaluated by comparing the preplanning and deformed PET volumes with the planning PET volume using Dice's coefficient (DC) and center-of-mass (COM) displacement. Results: The average DCs were 0.12±0.14 and 0.19±0.16 for rigid and deformable predicted target volumes, respectively. The average COM displacements were 1.9±0.9 cm and 1.7±0.7 cm for rigid and deformable registration, respectively. The DCs were improved by deformable registration, however, both were lower than published data for DIR in other modalities and clinical sites. Anatomical changes caused by different brachytherapy applicators could have posed a challenge to the DIR algorithm. The physiological change from interstitial needle placement may also contribute to lower DC. Conclusion: The clinical use of DIR in PET/CT for cervical cancer brachytherapy appears to be limited by applicator choice and requires further investigation.« less
  • Purpose: To investigate the impact of different clipbox volumes with automated registration techniques using commercially available software with on board volumetric imaging(OBI) for treatment verification in cervical cancer patients. Methods: Fifty cervical cancer patients received daily CBCT scans(on-board imaging v1.5 system, Varian Medical Systems) during the first treatment week and weekly thereafter were included this analysis. A total of 450 CBCT scans were registered to the planning CTscan using pelvic clipbox(clipbox-Pelvic) and around PTV clip box(clipbox- PTV). The translations(anterior-posterior, left-right, superior-inferior) and the rotations(yaw, pitch and roll) errors for each matches were recorded. The setup errors and the systematic andmore » random errors for both of the clip-boxes were calculated. Paired Samples t test was used to analysis the differences between clipbox-Pelvic and clipbox-PTV. Results: . The SD of systematic error(σ) was 1.0mm, 2.0mm,3.2mm and 1.9mm,2.3mm, 3.0mm in the AP, LR and SI directions for clipbox-Pelvic and clipbox-PTV, respectively. The average random error(Σ)was 1.7mm, 2.0mm,4.2mm and 1.7mm,3.4mm, 4.4mm in the AP, LR and SI directions for clipbox-Pelvic and clipbox-PTV, respectively. But, only the SI direction was acquired significantly differences between two image registration volumes(p=0.002,p=0.01 for mean and SD). For rotations, the yaw mean/SD and the pitch SD were acquired significantly differences between clipbox-Pelvic and clipbox-PTV. Conclusion: The defined volume for Image registration is important for cervical cancer when 3D/3D match was used. The alignment clipbox can effect the setup errors obtained. Further analysis is need to determine the optimal defined volume to use the image registration in cervical cancer. Conflict of interest: none.« less
  • Purpose: To propose and validate a deformable point matching scheme for surface deformation to facilitate accurate bladder dose summation for fractionated HDR cervical cancer treatment. Method: A deformable point matching scheme based on the thin plate spline robust point matching (TPSRPM) algorithm is proposed for bladder surface registration. The surface of bladders segmented from fractional CT images is extracted and discretized with triangular surface mesh. Deformation between the two bladder surfaces are obtained by matching the two meshes' vertices via the TPS-RPM algorithm, and the deformation vector fields (DVFs) characteristic of this deformation is estimated by B-spline approximation. Numerically, themore » algorithm is quantitatively compared with the Demons algorithm using five clinical cervical cancer cases by several metrics: vertex-to-vertex distance (VVD), Hausdorff distance (HD), percent error (PE), and conformity index (CI). Experimentally, the algorithm is validated on a balloon phantom with 12 surface fiducial markers. The balloon is inflated with different amount of water, and the displacement of fiducial markers is benchmarked as ground truth to study TPS-RPM calculated DVFs' accuracy. Results: In numerical evaluation, the mean VVD is 3.7(±2.0) mm after Demons, and 1.3(±0.9) mm after TPS-RPM. The mean HD is 14.4 mm after Demons, and 5.3mm after TPS-RPM. The mean PE is 101.7% after Demons and decreases to 18.7% after TPS-RPM. The mean CI is 0.63 after Demons, and increases to 0.90 after TPS-RPM. In the phantom study, the mean Euclidean distance of the fiducials is 7.4±3.0mm and 4.2±1.8mm after Demons and TPS-RPM, respectively. Conclusions: The bladder wall deformation is more accurate using the feature-based TPS-RPM algorithm than the intensity-based Demons algorithm, indicating that TPS-RPM has the potential for accurate bladder dose deformation and dose summation for multi-fractional cervical HDR brachytherapy. This work is supported in part by the National Natural ScienceFoundation of China (no 30970866 and no 81301940)« less