Title: SU-F-J-112: Clinical Feasibility Test of An RF Pulse-Based MRI Method for the Quantitative Fat-Water Segmentation

Abstract

Purpose: Quantitative fat-water segmentation is important not only because of the clinical utility of fat-suppressed MRI images in better detecting lesions of clinical significance (in the midst of bright fat signal) but also because of the possible physical need, in which CT-like images based on the materials’ photon attenuation properties may have to be generated from MR images; particularly, as in the case of MR-only radiation oncology environment to obtain radiation dose calculation or as in the case of hybrid PET/MR modality to obtain attenuation correction map for the quantitative PET reconstruction. The majority of such fat-water quantitative segmentations have been performed by utilizing the Dixon’s method and its variations, which have to enforce the proper settings (often predefined) of echo time (TE) in the pulse sequences. Therefore, such methods have been unable to be directly combined with those ultrashort TE (UTE) sequences that, taking the advantage of very low TE values (∼ 10’s microsecond), might be beneficial to directly detect bones. Recently, an RF pulse-based method (http://dx.doi.org/10.1016/j.mri.2015.11.006), termed as PROD pulse method, was introduced as a method of quantitative fat-water segmentation that does not have to depend on predefined TE settings. Here, the clinical feasibility of this method ismore » verified in brain tumor patients by combining the PROD pulse with several sequences. Methods: In a clinical 3T MRI, the PROD pulse was combined with turbo spin echo (e.g. TR=1500, TE=16 or 60, ETL=15) or turbo field echo (e.g. TR=5.6, TE=2.8, ETL=12) sequences without specifying TE values. Results: The fat-water segmentation was possible without having to set specific TE values. Conclusion: The PROD pulse method is clinically feasible. Although not yet combined with UTE sequences in our laboratory, the method is potentially compatible with UTE sequences, and thus, might be useful to directly segment fat, water, bone and air.« less

Authors:
; ;  [1]
  1. William Beaumont Hospital, Royal Oak, MI (United States)
Publication Date:
OSTI Identifier:
22634719
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 43; Journal Issue: 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; 61 RADIATION PROTECTION AND DOSIMETRY; BRAIN; CORRECTIONS; FATS; NEOPLASMS; NMR IMAGING; PATIENTS; POSITRON COMPUTED TOMOGRAPHY; SKELETON; SPIN ECHO

Citation Formats

Yee, S, Wloch, J, and Pirkola, M. SU-F-J-112: Clinical Feasibility Test of An RF Pulse-Based MRI Method for the Quantitative Fat-Water Segmentation. United States: N. p., 2016. Web. doi:10.1118/1.4956020.
Yee, S, Wloch, J, & Pirkola, M. SU-F-J-112: Clinical Feasibility Test of An RF Pulse-Based MRI Method for the Quantitative Fat-Water Segmentation. United States. doi:10.1118/1.4956020.
Yee, S, Wloch, J, and Pirkola, M. 2016. "SU-F-J-112: Clinical Feasibility Test of An RF Pulse-Based MRI Method for the Quantitative Fat-Water Segmentation". United States. doi:10.1118/1.4956020.
@article{osti_22634719,
title = {SU-F-J-112: Clinical Feasibility Test of An RF Pulse-Based MRI Method for the Quantitative Fat-Water Segmentation},
author = {Yee, S and Wloch, J and Pirkola, M},
abstractNote = {Purpose: Quantitative fat-water segmentation is important not only because of the clinical utility of fat-suppressed MRI images in better detecting lesions of clinical significance (in the midst of bright fat signal) but also because of the possible physical need, in which CT-like images based on the materials’ photon attenuation properties may have to be generated from MR images; particularly, as in the case of MR-only radiation oncology environment to obtain radiation dose calculation or as in the case of hybrid PET/MR modality to obtain attenuation correction map for the quantitative PET reconstruction. The majority of such fat-water quantitative segmentations have been performed by utilizing the Dixon’s method and its variations, which have to enforce the proper settings (often predefined) of echo time (TE) in the pulse sequences. Therefore, such methods have been unable to be directly combined with those ultrashort TE (UTE) sequences that, taking the advantage of very low TE values (∼ 10’s microsecond), might be beneficial to directly detect bones. Recently, an RF pulse-based method (http://dx.doi.org/10.1016/j.mri.2015.11.006), termed as PROD pulse method, was introduced as a method of quantitative fat-water segmentation that does not have to depend on predefined TE settings. Here, the clinical feasibility of this method is verified in brain tumor patients by combining the PROD pulse with several sequences. Methods: In a clinical 3T MRI, the PROD pulse was combined with turbo spin echo (e.g. TR=1500, TE=16 or 60, ETL=15) or turbo field echo (e.g. TR=5.6, TE=2.8, ETL=12) sequences without specifying TE values. Results: The fat-water segmentation was possible without having to set specific TE values. Conclusion: The PROD pulse method is clinically feasible. Although not yet combined with UTE sequences in our laboratory, the method is potentially compatible with UTE sequences, and thus, might be useful to directly segment fat, water, bone and air.},
doi = {10.1118/1.4956020},
journal = {Medical Physics},
number = 6,
volume = 43,
place = {United States},
year = 2016,
month = 6
}
  • Quantitative assessment of metastatic disease in bone is often considered immeasurable and, as such, patients with skeletal metastases are often excluded from clinical trials. In order to effectively quantify the impact of metastatic tumor involvement in the spine, accurate segmentation of the vertebra is required. Manual segmentation can be accurate but involves extensive and time-consuming user interaction. Potential solutions to automating segmentation of metastatically involved vertebrae are demons deformable image registration and level set methods. The purpose of this study was to develop a semiautomated method to accurately segment tumor-bearing vertebrae using the aforementioned techniques. By maintaining morphology of anmore » atlas, the demons-level set composite algorithm was able to accurately differentiate between trans-cortical tumors and surrounding soft tissue of identical intensity. The algorithm successfully segmented both the vertebral body and trabecular centrum of tumor-involved and healthy vertebrae. This work validates our approach as equivalent in accuracy to an experienced user.« less
  • Purpose: To validate and clinically evaluate autocontouring using atlas-based autosegmentation (ABAS) of computed tomography images. Methods and Materials: The data from 10 head-and-neck patients were selected as input for ABAS, and neck levels I-V and 20 organs at risk were manually contoured according to published guidelines. The total contouring times were recorded. Two different ABAS strategies, multiple and single subject, were evaluated, and the similarity of the autocontours with the atlas contours was assessed using Dice coefficients and the mean distances, using the leave-one-out method. For 12 clinically treated patients, 5 experienced observers edited the autosegmented contours. The editing timesmore » were recorded. The Dice coefficients and mean distances were calculated among the clinically used contours, autocontours, and edited autocontours. Finally, an expert panel scored all autocontours and the edited autocontours regarding their adequacy relative to the published atlas. Results: The time to autosegment all the structures using ABAS was 7 min/patient. No significant differences were observed in the autosegmentation accuracy for stage N0 and N+ patients. The multisubject atlas performed best, with a Dice coefficient and mean distance of 0.74 and 2 mm, 0.67 and 3 mm, 0.71 and 2 mm, 0.50 and 2 mm, and 0.78 and 2 mm for the salivary glands, neck levels, chewing muscles, swallowing muscles, and spinal cord-brainstem, respectively. The mean Dice coefficient and mean distance of the autocontours vs. the clinical contours was 0.8 and 2.4 mm for the neck levels and salivary glands, respectively. For the autocontours vs. the edited autocontours, the mean Dice coefficient and mean distance was 0.9 and 1.6 mm, respectively. The expert panel scored 100% of the autocontours as a 'minor deviation, editable' or better. The expert panel scored 88% of the edited contours as good compared with 83% of the clinical contours. The total editing time was 66 min. Conclusion: Multiple-subject ABAS of computed tomography images proved to be a useful novel tool in the rapid delineation of target and normal tissues. Although editing of the autocontours is inevitable, a substantial time reduction was achieved using editing, instead of manual contouring (180 vs. 66 min).« less
  • Purpose: The aim of this study was to assess whether clinically acceptable segmentations of organs at risk (OARs) in head and neck cancer can be obtained automatically and efficiently using the novel “similarity and truth estimation for propagated segmentations” (STEPS) compared to the traditional “simultaneous truth and performance level estimation” (STAPLE) algorithm. Methods: First, 6 OARs were contoured by 2 radiation oncologists in a dataset of 100 patients with head and neck cancer on planning computed tomography images. Each image in the dataset was then automatically segmented with STAPLE and STEPS using those manual contours. Dice similarity coefficient (DSC) wasmore » then used to compare the accuracy of these automatic methods. Second, in a blind experiment, three separate and distinct trained physicians graded manual and automatic segmentations into one of the following three grades: clinically acceptable as determined by universal delineation guidelines (grade A), reasonably acceptable for clinical practice upon manual editing (grade B), and not acceptable (grade C). Finally, STEPS segmentations graded B were selected and one of the physicians manually edited them to grade A. Editing time was recorded. Results: Significant improvements in DSC can be seen when using the STEPS algorithm on large structures such as the brainstem, spinal canal, and left/right parotid compared to the STAPLE algorithm (all p < 0.001). In addition, across all three trained physicians, manual and STEPS segmentation grades were not significantly different for the brainstem, spinal canal, parotid (right/left), and optic chiasm (all p > 0.100). In contrast, STEPS segmentation grades were lower for the eyes (p < 0.001). Across all OARs and all physicians, STEPS produced segmentations graded as well as manual contouring at a rate of 83%, giving a lower bound on this rate of 80% with 95% confidence. Reduction in manual interaction time was on average 61% and 93% when automatic segmentations did and did not, respectively, require manual editing. Conclusions: The STEPS algorithm showed better performance than the STAPLE algorithm in segmenting OARs for radiotherapy of the head and neck. It can automatically produce clinically acceptable segmentation of OARs, with results as relevant as manual contouring for the brainstem, spinal canal, the parotids (left/right), and optic chiasm. A substantial reduction in manual labor was achieved when using STEPS even when manual editing was necessary.« less
  • To calculate the transient accumulation of soluble organic matter in activated sludge, an equation based on COD and respirometry is described. In this approach the difference between the actual utilized organic matter and the metabolized matter is expressed as substrate transient accumulation. The amount of substrate generated by the lysis of dead microorganisms was taken into account, and the metabolized organic matter was expressed in two main endergonic functions of O/sub 2/ consumption, that is, assimilation and maintenance. The equation thus derived was simplified to contain parameters such as COD and O/sub 2/ consumption and tested on experimental results. Themore » results show that the values obtained using this equation compared well to substrate removal in the absence of O/sub 2/ in short-term batch cultures. 44 references.« less