Title: SU-F-J-17: Patient Localization Using MRI-Guided Soft Tissue for Head-And-Neck Radiotherapy: Indication for Margin Reduction and Its Feasibility

Abstract

Purpose: On-board MRI provides superior soft-tissue contrast, allowing patient alignment using tumor or nearby critical structures. This study aims to study H&N MRI-guided IGRT to analyze inter-fraction patient setup variations using soft-tissue targets and design appropriate CTV-to-PTV margin and clinical implication. Methods: 282 MR images for 10 H&N IMRT patients treated on a ViewRay system were retrospectively analyzed. Patients were immobilized using a thermoplastic mask on a customized headrest fitted in a radiofrequency coil and positioned to soft-tissue targets. The inter-fraction patient displacements were recorded to compute the PTV margins using the recipe: 2.5∑+0.7σ. New IMRT plans optimized on the revised PTVs were generated to evaluate the delivered dose distributions. An in-house dose deformation registration tool was used to assess the resulting dosimetric consequences when margin adaption is performed based on weekly MR images. The cumulative doses were compared to the reduced margin plans for targets and critical structures. Results: The inter-fraction displacements (and standard deviations), ∑ and σ were tabulated for MRI and compared to kVCBCT. The computed CTV-to-PTV margin was 3.5mm for soft-tissue based registration. There were minimal differences between the planned and delivered doses when comparing clinical and the PTV reduced margin plans: the paired t-tests yieldedmore » p=0.38 and 0.66 between the planned and delivered doses for the adapted margin plans for the maximum cord and mean parotid dose, respectively. Target V95 received comparable doses as planned for the reduced margin plans. Conclusion: The 0.35T MRI offers acceptable soft-tissue contrast and good spatial resolution for patient alignment and target visualization. Better tumor conspicuity from MRI allows soft-tissue based alignments with potentially improved accuracy, suggesting a benefit of margin reduction for H&N radiotherapy. The reduced margin plans (i.e., 2 mm) resulted in improved normal structure sparing and accurate dose delivery to achieve intended treatment goal under MR guidance.« less

Authors:
; ; ; ; ; ;  [1]
  1. UCLA, Los Angeles, CA (United States)
Publication Date:
OSTI Identifier:
22632152
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 43; Journal Issue: 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; 61 RADIATION PROTECTION AND DOSIMETRY; ACCURACY; ANIMAL TISSUES; DEFORMATION; HEAD; IMAGES; NECK; NEOPLASMS; NMR IMAGING; PATIENTS; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; RADIOTHERAPY; RADIOWAVE RADIATION; SPATIAL RESOLUTION; THERMOPLASTICS

Citation Formats

Qi, X, Yang, Y, Jack, N, Santhanam, A, Yang, L, Chen, A, and Low, D. SU-F-J-17: Patient Localization Using MRI-Guided Soft Tissue for Head-And-Neck Radiotherapy: Indication for Margin Reduction and Its Feasibility. United States: N. p., 2016. Web. doi:10.1118/1.4955925.
Qi, X, Yang, Y, Jack, N, Santhanam, A, Yang, L, Chen, A, & Low, D. SU-F-J-17: Patient Localization Using MRI-Guided Soft Tissue for Head-And-Neck Radiotherapy: Indication for Margin Reduction and Its Feasibility. United States. doi:10.1118/1.4955925.
Qi, X, Yang, Y, Jack, N, Santhanam, A, Yang, L, Chen, A, and Low, D. Wed . "SU-F-J-17: Patient Localization Using MRI-Guided Soft Tissue for Head-And-Neck Radiotherapy: Indication for Margin Reduction and Its Feasibility". United States. doi:10.1118/1.4955925.
@article{osti_22632152,
title = {SU-F-J-17: Patient Localization Using MRI-Guided Soft Tissue for Head-And-Neck Radiotherapy: Indication for Margin Reduction and Its Feasibility},
author = {Qi, X and Yang, Y and Jack, N and Santhanam, A and Yang, L and Chen, A and Low, D},
abstractNote = {Purpose: On-board MRI provides superior soft-tissue contrast, allowing patient alignment using tumor or nearby critical structures. This study aims to study H&N MRI-guided IGRT to analyze inter-fraction patient setup variations using soft-tissue targets and design appropriate CTV-to-PTV margin and clinical implication. Methods: 282 MR images for 10 H&N IMRT patients treated on a ViewRay system were retrospectively analyzed. Patients were immobilized using a thermoplastic mask on a customized headrest fitted in a radiofrequency coil and positioned to soft-tissue targets. The inter-fraction patient displacements were recorded to compute the PTV margins using the recipe: 2.5∑+0.7σ. New IMRT plans optimized on the revised PTVs were generated to evaluate the delivered dose distributions. An in-house dose deformation registration tool was used to assess the resulting dosimetric consequences when margin adaption is performed based on weekly MR images. The cumulative doses were compared to the reduced margin plans for targets and critical structures. Results: The inter-fraction displacements (and standard deviations), ∑ and σ were tabulated for MRI and compared to kVCBCT. The computed CTV-to-PTV margin was 3.5mm for soft-tissue based registration. There were minimal differences between the planned and delivered doses when comparing clinical and the PTV reduced margin plans: the paired t-tests yielded p=0.38 and 0.66 between the planned and delivered doses for the adapted margin plans for the maximum cord and mean parotid dose, respectively. Target V95 received comparable doses as planned for the reduced margin plans. Conclusion: The 0.35T MRI offers acceptable soft-tissue contrast and good spatial resolution for patient alignment and target visualization. Better tumor conspicuity from MRI allows soft-tissue based alignments with potentially improved accuracy, suggesting a benefit of margin reduction for H&N radiotherapy. The reduced margin plans (i.e., 2 mm) resulted in improved normal structure sparing and accurate dose delivery to achieve intended treatment goal under MR guidance.},
doi = {10.1118/1.4955925},
journal = {Medical Physics},
number = 6,
volume = 43,
place = {United States},
year = {Wed Jun 15 00:00:00 EDT 2016},
month = {Wed Jun 15 00:00:00 EDT 2016}
}
  • Purpose: We propose a surface-image-guided soft robotic patient positioning system for maskless head-and-neck radiotherapy. The ultimate goal of this project is to utilize a soft robot to realize non-rigid patient positioning and real-time motion compensation. In this proof-of-concept study, we design a position-based visual servoing control system for an air-bladder-based soft robot and investigate its performance in controlling the flexion/extension cranial motion on a mannequin head phantom. Methods: The current system consists of Microsoft Kinect depth camera, an inflatable air bladder (IAB), pressured air source, pneumatic valve actuators, custom-built current regulators, and a National Instruments myRIO microcontroller. The performance ofmore » the designed system was evaluated on a mannequin head, with a ball joint fixed below its neck to simulate torso-induced head motion along flexion/extension direction. The IAB is placed beneath the mannequin head. The Kinect camera captures images of the mannequin head, extracts the face, and measures the position of the head relative to the camera. This distance is sent to the myRIO, which runs control algorithms and sends actuation commands to the valves, inflating and deflating the IAB to induce head motion. Results: For a step input, i.e. regulation of the head to a constant displacement, the maximum error was a 6% overshoot, which the system then reduces to 0% steady-state error. In this initial investigation, the settling time to reach the regulated position was approximately 8 seconds, with 2 seconds of delay between the command start of motion due to capacitance of the pneumatics, for a total of 10 seconds to regulate the error. Conclusion: The surface image-guided soft robotic patient positioning system can achieve accurate mannequin head flexion/extension motion. Given this promising initial Result, the extension of the current one-dimensional soft robot control to multiple IABs for non-rigid positioning control will be pursued.« less
  • Purpose: To investigate the feasibility of improved normal tissue sparing for head-and-neck (H'N) image-guided radiotherapy (IGRT) by employing tighter CTV-to-PTV margins for target level II/III though a GPU-based deformable image registration and dose accumulation framework. Methods: Ten H'N simultaneous integrated boost cases treated on TomoTherapy were retrospectively analyzed. Weekly kVCT scans in addition to daily MVCT scans were acquired for each patient. Reduced margin plans were generated with 0- mm margin for level II and III PTV (while 3-5 mm margin for PTV1) and compared with the standard margin plan using 3-5mm margin to all CTV1-3 (reference plan). An in-housemore » developed GPU-based 3D image deformation tool was used to register and deform the weekly KVCTs with the planning CT and determine the delivered mean/minimum/maximum dose, dose volume histograms (DVHs), etc. Results: Compared with the reference plans, the averaged cord maximum, the right and left parotid doses reduced by 22.7 %, 16.5 %, and 9 % respectively in the reduced margin plans. The V95 for PTV2 and PTV3 were found within 2 and 5% between the reference and tighter margin plans. For the reduced margin plans, the averaged cumulative mean doses were consistent with the planned dose for PTV1, PTV2 and PTV3 within 1.5%, 1.7% and 1.4%. Similar dose variations of the delivered dose were seen for the reference and tighter margin plans. The delivered maximum and mean doses for the cord were 3.55 % and 2.37% higher than the planned doses; a 5 % higher cumulative mean dose for the parotids was also observed for the delivered dose than the planned doses in both plans. Conclusion: By imposing tighter CTV-to-PTV margins for level II and III targets for H'N irradiation, acceptable cumulative doses were achievable when coupled with weekly kVCT guidance while improving normal structure sparing.« less
  • Purpose: The recent development of an MRI-linac allows adaptation of treatments to the anatomy of the moment. This anatomy, in turn, could be altered into a more favorable situation for radiotherapy purposes. The purpose of this study is to investigate the potential dosimetric benefits of manipulating rectal anatomy in MRI-guided interventional external-beam radiotherapy for rectal cancer. Methods: For this retrospective analysis, four patients (1M/3F) diagnosed with rectal cancer were included. These underwent MR-imaging using sonography transmission gel as endorectal contrast at time of diagnosis and standard, non-contrast, MR-imaging prior to radiotherapy planning. In the contrast scan, the rectum is inflatedmore » by the inserted contrast gel, thereby potentially increasing the distance between tumor and the organs-at-risk (OAR). Both anatomies were delineated and 7- beam IMRT-plans were calculated for both situations (RT-standard and RT-inflated), using in-house developed treatment planning software. Each plan was aimed to deliver 15Gy to the planning target volume (PTV; tumor+3mm margin) with a D99>95% and Dmax<120% of the planned dose. The D2cc dose to the OAR were then compared for both situations. Results: At equal (or better) target coverage, we found a mean reduction in D2cc of 4.1Gy/237% [range 2.6Gy–6.3Gy/70%–621%] for the bladder and of 2.0Gy/145% [range −0.7Gy–7.9Gy/−73%–442%] for the small-bowel, for the RT-inflated compared to the RT-standard plans. For the three female patients, a reduction in D2cc of 5.2Gy/191% [range 3.2Gy–9.2Gy/44%–475%] for the gynecological organs was found. We found all D2cc doses to be better for the RT-inflated plans, except for one patient for whom the bladder D2cc dose was slightly increased. Conclusion: Reduction of OAR dose by manipulation of anatomy is feasible. Inflation of the rectum results in more distance between OAR and PTV. This leads to a substantial reduction in dose to OAR at equal or better target coverage.« less
  • Purpose: To demonstrate real-time dose calculation of free-breathing MRI guided Co−60 treatments, using a motion model and Monte-Carlo dose calculation to accurately account for the interplay between irregular breathing motion and an IMRT delivery. Methods: ViewRay Co-60 dose distributions were optimized on ITVs contoured from free-breathing CT images of lung cancer patients. Each treatment plan was separated into 0.25s segments, accounting for the MLC positions and beam angles at each time point. A voxel-specific motion model derived from multiple fast-helical free-breathing CTs and deformable registration was calculated for each patient. 3D images for every 0.25s of a simulated treatment weremore » generated in real time, here using a bellows signal as a surrogate to accurately account for breathing irregularities. Monte-Carlo dose calculation was performed every 0.25s of the treatment, with the number of histories in each calculation scaled to give an overall 1% statistical uncertainty. Each dose calculation was deformed back to the reference image using the motion model and accumulated. The static and real-time dose calculations were compared. Results: Image generation was performed in real time at 4 frames per second (GPU). Monte-Carlo dose calculation was performed at approximately 1frame per second (CPU), giving a total calculation time of approximately 30 minutes per treatment. Results show both cold- and hot-spots in and around the ITV, and increased dose to contralateral lung as the tumor moves in and out of the beam during treatment. Conclusion: An accurate motion model combined with a fast Monte-Carlo dose calculation allows almost real-time dose calculation of a free-breathing treatment. When combined with sagittal 2D-cine-mode MRI during treatment to update the motion model in real time, this will allow the true delivered dose of a treatment to be calculated, providing a useful tool for adaptive planning and assessing the effectiveness of gated treatments.« less
  • Purpose: There is growing interests in applying MR-simulator(MR-sim) in radiotherapy but MR images subject to hardware, patient and pulse sequence dependent geometric distortion that may potentially influence target definition. This study aimed to evaluate the influence on head-and-neck tissue delineation, in terms of positional and volumetric variability, of two T1-weighted(T1w) MR sequences on a 1.5T MR-sim Methods: Four healthy volunteers were scanned (4 scans for each on different days) using both spin-echo (3DCUBE, TR/TE=500/14ms, TA=183s) and gradient-echo sequences (3DFSPGR, TE/TR=7/4ms, TA=173s) with identical coverage, voxel-size(0.8×0.8×1.0mm3), receiver-bandwidth(62.5kHz/pix) and geometric correction on a 1.5T MR-sim immobilized with personalized thermoplastic cast and head-rest.more » Under this setting, similar T1w contrast and signal-to-noise ratio were obtained, and factors other than sequence that might bias image distortion and tissue delineation were minimized. VOIs of parotid gland(PGR, PGL), pituitary gland(PIT) and eyeballs(EyeL, EyeR) were carefully drawn, and inter-scan coefficient-of-variation(CV) of VOI centroid position and volume were calculated for each subject. Mean and standard deviation(SD) of the CVs for four subjects were compared between sequences using Wilcoxon ranksum test. Results: The mean positional(<4%) and volumetric(<7%) CVs varied between tissues, majorly dependent on tissue inherent properties like volume, location, mobility and deformability. Smaller mean volumetric CV was found in 3DCUBE, probably due to its less proneness to tissue susceptibility, but only PGL showed significant difference(P<0.05). Positional CVs had no significant differences for all VOIs(P>0.05) between sequences, suggesting volumetric variation might be more sensitive to sequence-dependent delineation difference. Conclusion: Although 3DCUBE is considered less prone to tissue susceptibility-induced artifact and distortion, our preliminary data showed that both sequences had insignificant differences on positional and volumetric CV in most head-and-neck tissues except for PGL. This study is majorly limited in its small sample size. Influences of image contrasts(T1w v.s. T2w) and inter-observer difference have to be further investigated.« less