Title: SU-F-I-76: Fluoroscopic X-Ray Beam Profiles for Spectra Incorporating Copper Filtration

Abstract

Purpose: The purpose of this investigation is to characterize and quantify X-ray beam profiles for fluoroscopic x-ray beam spectra incorporating spectral (copper) filtration. Methods: A PTW (Freiburg, Germany) type 60016 silicon diode detector and PTW MP3 water tank were used to measure X-ray beam profiles for 60, 80, 100 and 120 kVp x-ray beams at five different copper filtration thicknesses ranging from 0–0.9 mm at 22 and 42 cm fields of view and depths of 1, 5, and 10 cm in both the anode-cathode axis (inplane) and cross-plane directions. All measurements were acquired on a Siemens (Erlangen, Germany) Artis ZeeGo fluoroscope inverted from the typical orientation providing an x-ray beam originating from above the water surface with the water level set at 60 cm from the focal spot. Results: X-ray beam profiles for beam spectra without copper filtration compared well to previously published data by Fetterly et al. [Med Phys, 28, 205 (2001)]. Our data collection benefited from the geometric orientation of the fluoroscope, providing a beam perpendicular to the tank water surface, rather than through a thin side wall as did the previously mentioned study. Profiles for beams with copper filtration were obtained which have not been previously investigatedmore » and published. Beam profiles in the anode-cathode axis near the surface and at lower x-ray energy exhibited substantial heel effect, which became less pronounced at greater depth. At higher energy with copper filtration in the beam, the dose falloff out-of-field became less pronounced, as would be anticipated given higher scatter photon energy. Conclusion: The x-ray beam profile data for the fluoroscopic x-ray beams incorporating copper filtration are intended for use as reference data for estimating doses to organs or soft tissue, including fetal dose, involving similar beam qualities or for comparison with mathematical models.« less

Authors:
 [1];  [2]; ; ;  [1];  [3]
  1. Cleveland Clinic Foundation, Cleveland, OH (United States)
  2. (United States)
  3. Wayne State University School of Medicine, Detroit, MI (United States)
Publication Date:
OSTI Identifier:
22632136
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 43; Journal Issue: 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
61 RADIATION PROTECTION AND DOSIMETRY; 60 APPLIED LIFE SCIENCES; ANIMAL TISSUES; ANODES; BEAM PROFILES; CATHODES; FILTRATION; FLUOROSCOPY; ORGANS; PHOTON BEAMS; RADIATION DOSES; SILICON DIODES; THICKNESS; X RADIATION

Citation Formats

Wunderle, K, Wayne State University School of Medicine, Detroit, MI, Godley, A, Shen, Z, Dong, F, and Rakowski, J. SU-F-I-76: Fluoroscopic X-Ray Beam Profiles for Spectra Incorporating Copper Filtration. United States: N. p., 2016. Web. doi:10.1118/1.4955904.
Wunderle, K, Wayne State University School of Medicine, Detroit, MI, Godley, A, Shen, Z, Dong, F, & Rakowski, J. SU-F-I-76: Fluoroscopic X-Ray Beam Profiles for Spectra Incorporating Copper Filtration. United States. doi:10.1118/1.4955904.
Wunderle, K, Wayne State University School of Medicine, Detroit, MI, Godley, A, Shen, Z, Dong, F, and Rakowski, J. 2016. "SU-F-I-76: Fluoroscopic X-Ray Beam Profiles for Spectra Incorporating Copper Filtration". United States. doi:10.1118/1.4955904.
@article{osti_22632136,
title = {SU-F-I-76: Fluoroscopic X-Ray Beam Profiles for Spectra Incorporating Copper Filtration},
author = {Wunderle, K and Wayne State University School of Medicine, Detroit, MI and Godley, A and Shen, Z and Dong, F and Rakowski, J},
abstractNote = {Purpose: The purpose of this investigation is to characterize and quantify X-ray beam profiles for fluoroscopic x-ray beam spectra incorporating spectral (copper) filtration. Methods: A PTW (Freiburg, Germany) type 60016 silicon diode detector and PTW MP3 water tank were used to measure X-ray beam profiles for 60, 80, 100 and 120 kVp x-ray beams at five different copper filtration thicknesses ranging from 0–0.9 mm at 22 and 42 cm fields of view and depths of 1, 5, and 10 cm in both the anode-cathode axis (inplane) and cross-plane directions. All measurements were acquired on a Siemens (Erlangen, Germany) Artis ZeeGo fluoroscope inverted from the typical orientation providing an x-ray beam originating from above the water surface with the water level set at 60 cm from the focal spot. Results: X-ray beam profiles for beam spectra without copper filtration compared well to previously published data by Fetterly et al. [Med Phys, 28, 205 (2001)]. Our data collection benefited from the geometric orientation of the fluoroscope, providing a beam perpendicular to the tank water surface, rather than through a thin side wall as did the previously mentioned study. Profiles for beams with copper filtration were obtained which have not been previously investigated and published. Beam profiles in the anode-cathode axis near the surface and at lower x-ray energy exhibited substantial heel effect, which became less pronounced at greater depth. At higher energy with copper filtration in the beam, the dose falloff out-of-field became less pronounced, as would be anticipated given higher scatter photon energy. Conclusion: The x-ray beam profile data for the fluoroscopic x-ray beams incorporating copper filtration are intended for use as reference data for estimating doses to organs or soft tissue, including fetal dose, involving similar beam qualities or for comparison with mathematical models.},
doi = {10.1118/1.4955904},
journal = {Medical Physics},
number = 6,
volume = 43,
place = {United States},
year = 2016,
month = 6
}
  • Purpose: The purpose of this investigation was to quantify percent depth dose (PDD) curves for fluoroscopic x-ray beam qualities incorporating added copper filtration. Methods: A PTW (Freiburg, Germany) MP3 water tank was used with a Standard Imaging (Middleton, WI) Exradin Model 11 Spokas Chamber to measure PDD curves for 60, 80, 100 and 120 kVp x-ray beams with copper filtration ranging from 0.0–0.9 mm at 22cm and 42cm fields of view from 0 to 150 mm of water. A free-in-air monitor chamber was used to normalize the water tank data to fluctuations in output from the fluoroscope. The measurements weremore » acquired on a Siemens (Erlangen, Germany) Artis ZeeGo fluoroscope. The fluoroscope was inverted from the typical orientation providing an x-ray beam originating from above the water tank. The water tank was positioned so that the water level was located at 60cm from the focal spot; which also represents the focal spot to interventional reference plane distance for that fluoroscope. Results: PDDs for 60, 80, 100, and 120 kVp with 0 mm of copper filtration compared well to previously published data by Fetterly et al. [Med Phys, 28, 205 (2001)] for those beam qualities given differences in fluoroscopes, geometric orientation, type of ionization chamber, and the water tank used for data collection. PDDs for 60, 80, 100, and 120 kVp with copper filtration were obtained and are presented, which have not been previously investigated and published. Conclusion: The equipment and processes used to acquire the reported data were sound and compared well with previously published data for PDDs without copper filtration. PDD data for the fluoroscopic x-ray beams incorporating copper filtration can be used as reference data for estimating organ or soft tissue dose at depth involving similar beam qualities or for comparison with mathematical models.« less
  • Purpose: The purpose of this investigation is to quantify various first half-value-layers (HVLs), second HVLs and homogeneity coefficients (HCs) for a state-of-the-art fluoroscope utilizing spectral (copper) filtration. Methods: A Radcal (Monrovia, Ca) AccuPro dosimeter with a 10×6-6 calibrated ionization chamber was used to measure air kerma for radiographic x-ray exposures made on a Siemens (Erlangen, Germany) Artis ZeeGo fluoroscope operated in the service mode. The ionization chamber was centered in the x-ray beam at 72 cm from the focal spot with a source-to-image-distance of 120 cm. The collimators were introduced to limit the x-ray field to approximately 5 cm ×more » 5 cm at the ionization chamber plane. Type-1100 aluminum filters, in 0.5 mm increments, were used to determine the HVL. Two HVL calculation methods were used, log-linear interpolation and Lambert-W interpolation as described by Mathieu [Med Phys, 38(8), 4546 (2011)]. Multiple measurements were made at 60, 80, 100, 120 kVp at spectral filtration thicknesses of 0, 0.1, 0.3, 0.6 and 0.9 mm. Results: First HVL, second HVL, and HCs are presented for the fluoroscopic x-ray beam spectra indicated above, with nearly identical results from the two interpolation methods. Accuracy of the set kVp was also determined and deviated less than 2%. First HVLs for fluoroscopic x-ray beam spectra without spectral filtration determined in our study were 7%–16% greater than previously published data by Fetterly et al. [Med Phys, 28, 205 (2001)]. However, the FDA minimum HVL requirements changed since that publication, requiring larger HVLs as of 2006. Additionally, x-ray tube and generator architecture have substantially changed over the last 15 years providing different beam spectra. Conclusion: X-ray beam quality characteristics for state-of-the-art fluoroscopes with spectral filtration have not been published. This study provides reference data which will be useful for defining beam qualities encountered on fluoroscopes using spectral filtration.« less
  • Purpose: Advanced dosimetry in CT (e.g. the Monte Carlo method) requires an accurate characterization of the shaped filter and radiation quality used during a scan. The purpose of this work was to develop a method where half value layer (HVL) profiles along shaped filters could be made. From the HVL profiles the beam shaping properties and effective photon spectrum for a particular scan can be inferred. Methods: A measurement rig was developed to allow determinations of the HVL under a scatter-free narrow-beam geometry and constant focal spot to ionization chamber distance for different fan angles. For each fan angle themore » HVL is obtained by fitting the transmission of radiation through different thicknesses of an Al absorber (type 1100) using an appropriate model. The effective Al thickness of shaped filters and effective photon spectra are estimated using a model of photon emission from a Tungsten anode. This method is used to obtain the effective photon spectra and effective Al thickness of shaped filters for a CT scanner recently introduced to the market. Results: This study resulted in a set of effective photon spectra (central ray) for each kVp along with effective Al thicknesses of the different shaped filters. The effective photon spectra and effective Al thicknesses of shaped filters were used to obtain numerically approximated HVL profiles and compared to measured HVL profiles (mean absolute percentage error = 0.02). The central axis HVL found in the vendor’s technical documentation were compared to approximated HVL values (mean absolute percentage error = 0.03). Conclusion: This work has resulted in a unique method of measuring HVL profiles along shaped filters in CT. Further the effective photon spectra and the effective Al thicknesses of shaped filters that were obtained can be incorporated into Monte Carlo simulations.« less
  • C-arm cone beam computed tomography is an advanced 3D imaging technology that is currently available on state-of-the-art flat-panel-based angiography systems. The overlay of cross-sectional imaging information can now be integrated with real-time fluoroscopy. This overlay technology was used to guide the placement of three percutaneous translumbar inferior vena cava catheters.