Title: SU-F-I-59: Quality Assurance Phantom for PET/CT Alignment and Attenuation Correction

Abstract

Purpose: This study utilizes a commercial PET/CT phantom to investigate two specific properties of a PET/CT system: the alignment accuracy of PET images with those from CT used for attenuation correction and the accuracy of this correction in PET images. Methods: A commercial PET/CT phantom consisting of three aluminum rods, two long central cylinders containing uniform activity, and attenuating materials such as air, water, bone and iodine contrast was scanned using a standard PET/CT protocol. Images reconstructed with 2 mm slice thickness and a 512 by 512 matrix were obtained. The center of each aluminum rod in the PET and CT images was compared to evaluate alignment accuracy. ROIs were drawn on transaxial images of the central rods at each section of attenuating material to determine the corrected activity (in BQML). BQML values were graphed as a function of slice number to provide a visual representation of the attenuation-correction throughout the whole phantom. Results: Alignment accuracy is high between the PET and CT images. The maximum deviation between the two in the axial plane is less than 1.5 mm, which is less than the width of a single pixel. BQML values measured along different sections of the large central rodsmore » are similar among the different attenuating materials except iodine contrast. Deviation of BQML values in the air and bone sections from the water section is less than 1%. Conclusion: Accurate alignment of PET and CT images is critical to ensure proper calculation and application of CT-based attenuation correction. This study presents a simple and quick method to evaluate the two with a single acquisition. As the phantom also includes spheres of increasing diameter, this could serve as a straightforward means to annually evaluate the status of a modern PET/CT system.« less

Authors:
;  [1]
  1. Columbia University Medical Center, New York, NY (United States)
Publication Date:
OSTI Identifier:
22632124
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 43; Journal Issue: 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; 61 RADIATION PROTECTION AND DOSIMETRY; ACCURACY; ALIGNMENT; ATTENUATION; BIOMEDICAL RADIOGRAPHY; CORRECTIONS; IMAGE PROCESSING; IMAGES; PHANTOMS; POSITRON COMPUTED TOMOGRAPHY; QUALITY ASSURANCE; SKELETON

Citation Formats

Lin, T, and Hamacher, K. SU-F-I-59: Quality Assurance Phantom for PET/CT Alignment and Attenuation Correction. United States: N. p., 2016. Web. doi:10.1118/1.4955887.
Lin, T, & Hamacher, K. SU-F-I-59: Quality Assurance Phantom for PET/CT Alignment and Attenuation Correction. United States. doi:10.1118/1.4955887.
Lin, T, and Hamacher, K. 2016. "SU-F-I-59: Quality Assurance Phantom for PET/CT Alignment and Attenuation Correction". United States. doi:10.1118/1.4955887.
@article{osti_22632124,
title = {SU-F-I-59: Quality Assurance Phantom for PET/CT Alignment and Attenuation Correction},
author = {Lin, T and Hamacher, K},
abstractNote = {Purpose: This study utilizes a commercial PET/CT phantom to investigate two specific properties of a PET/CT system: the alignment accuracy of PET images with those from CT used for attenuation correction and the accuracy of this correction in PET images. Methods: A commercial PET/CT phantom consisting of three aluminum rods, two long central cylinders containing uniform activity, and attenuating materials such as air, water, bone and iodine contrast was scanned using a standard PET/CT protocol. Images reconstructed with 2 mm slice thickness and a 512 by 512 matrix were obtained. The center of each aluminum rod in the PET and CT images was compared to evaluate alignment accuracy. ROIs were drawn on transaxial images of the central rods at each section of attenuating material to determine the corrected activity (in BQML). BQML values were graphed as a function of slice number to provide a visual representation of the attenuation-correction throughout the whole phantom. Results: Alignment accuracy is high between the PET and CT images. The maximum deviation between the two in the axial plane is less than 1.5 mm, which is less than the width of a single pixel. BQML values measured along different sections of the large central rods are similar among the different attenuating materials except iodine contrast. Deviation of BQML values in the air and bone sections from the water section is less than 1%. Conclusion: Accurate alignment of PET and CT images is critical to ensure proper calculation and application of CT-based attenuation correction. This study presents a simple and quick method to evaluate the two with a single acquisition. As the phantom also includes spheres of increasing diameter, this could serve as a straightforward means to annually evaluate the status of a modern PET/CT system.},
doi = {10.1118/1.4955887},
journal = {Medical Physics},
number = 6,
volume = 43,
place = {United States},
year = 2016,
month = 6
}
  • Purpose: To discuss experiences and results for a 3D printed QA phantom used for daily alignment purposes for a six degrees-of-freedom (6DoF) table Methods: A 3D model was created using a fused deposition modeling (FDM) printer using free online computer-aided design (CAD) software. The model has been under use for daily QA alignment tests for a 6DoF couch. An aligned and angled baseplate were also printed in order to introduce known angles for 6DoF corrections during image-guidance. Unique registration contours were created on the faces of the phantom in order to achieve a better cone-beam computed tomography (CBCT) match usingmore » an auto-registration algorithm. A BB was also introduced at the center of the phantom in order to deliver an integrated daily Winston-Lutz (WL) test. Translational, rotational, and WL results were tabulated over one month. Results: The ’honeycomb’ structure of the print was apparent in the EPID images for the WL test, which affected the results of the analysis software. This was fixed by inserting a cube made of polyoxymethylene within the 3D phantom that encompass the BB. Auto-registration results for the three translational and three rotational from a known offset to the BB isocenter consistently fell within 1 mm and 0.2°, respectively. WL tests resulted in an average of 0.71 ± 0.14 mm. Conclusion: 3D printed models allow for accurate builds that can be customized to a variety of clinical needs. Results from translational, rotational, and WL show consistent results over a month’s time. Given its relatively cheap and streamlined workflow, 3D printing could be implemented into any clinic looking to create customized phantoms.« less
  • Purpose: Chang’s mathematical formulation is a common method of attenuation correction applied on reconstructed Jaszczak phantom images. Though Chang’s attenuation correction method has been used for 360° angle acquisition, its applicability for 180° angle acquisition remains a question with one vendor’s camera software producing artifacts. The objective of this work is to ensure that Chang’s attenuation correction technique can be applied for reconstructed Jaszczak phantom images acquired in both 360° and 180° mode. Methods: The Jaszczak phantom filled with 20 mCi of diluted Tc-99m was placed on the patient table of Siemens e.cam™ (n = 2) and Siemens Symbia™ (nmore » = 1) dual head gamma cameras centered both in lateral and axial directions. A total of 3 scans were done at 180° and 2 scans at 360° orbit acquisition modes. Thirty two million counts were acquired for both modes. Reconstruction of the projection data was performed using filtered back projection smoothed with pre reconstruction Butterworth filter (order: 6, cutoff: 0.55). Reconstructed transaxial slices were attenuation corrected by Chang’s attenuation correction technique as implemented in the camera software. Corrections were also done using a modified technique where photon path lengths for all possible attenuation paths through a pixel in the image space were added to estimate the corresponding attenuation factor. The inverse of the attenuation factor was utilized to correct the attenuated pixel counts. Results: Comparable uniformity and noise were observed for 360° acquired phantom images attenuation corrected by the vendor technique (28.3% and 7.9%) and the proposed technique (26.8% and 8.4%). The difference in uniformity for 180° acquisition between the proposed technique (22.6% and 6.8%) and the vendor technique (57.6% and 30.1%) was more substantial. Conclusion: Assessment of attenuation correction performance by phantom uniformity analysis illustrated improved uniformity with the proposed algorithm compared to the camera software.« less
  • A simple, easy to use, quality assurance and performance test phantom was developed for the xenon/computed tomography (CT) cerebral blood flow method. The phantom combines an inhalation system which allows for the simulation of xenon buildup or washout in the arterial blood as well as a multisection translatable cylinder in which several sections can be scanned during a preselected protocol to simulate the CT enhancement in brain tissue during a study. The phantom and scanning protocol are described and their use is demonstrated. The results compare favorably to the theoretically expected fast, intermediate, and slow flow values designed into themore » phantom.« less
  • Purpose: The evolution of three-dimensional conformal radiation treatment has led to the use of multileaf collimators (MLCs) in intensity-modulated radiation therapy (IMRT) and other treatment techniques to increase the conformity of the dose distribution. A new quality assurance (QA) phantom has been designed to check the handling of MLC settings in treatment planning and delivery. Methods and materials: The phantom consists of a Perspex block with stepped edges that can be rotated in all planes. The design allows for the assessment of several MLC and micro-MLC types from various manufacturers, and is therefore applicable to most radiation therapy institutions employingmore » MLCs. The phantom is computed tomography (CT) scanned as is a patient, and QA assessments can be made of field edge display for a variety of shapes and orientations on both radiation treatment planning systems (RTPS) and computed tomography simulators. Results: The dimensions of the phantom were verified to be physically correct within an uncertainty range of 0-0.7 mm. Errors in leaf position larger than 1 mm were easily identified by multiple observers. Conclusions: The MLC geometry phantom is a useful tool in the QA of radiation therapy with application to RTPS, CT simulators, and virtual simulation packages with MLC display capabilities.« less
  • The objective of this work is to develop a dosimetric phantom quality assurance (QA) of linear accelerators capable of cone-beam CT (CBCT) image guided and intensity-modulated radiotherapy (IG-IMRT). This phantom is to be used in an integral test to quantify in real-time both the performance of the image guidance and the dose delivery systems in terms of dose localization. The prototype IG-IMRT QA phantom consisted of a cylindrical imaging phantom (CatPhan) combined with an array of 11 radiation diodes mounted on a 10 cm diameter disk, oriented perpendicular to the phantom axis. Basic diode response characterization was performed for 6more » and 18 MV photons. The diode response was compared to planning system calculations in the open and penumbrae regions of simple and complex beam arrangements. The clinical use of the QA phantom was illustrated in an integral test of an IG-IMRT treatment designed for a clinical spinal radiosurgery case. The sensitivity of the phantom to multileaf collimator (MLC) calibration and setup errors in the clinical setting was assessed by introducing errors in the IMRT plan or by displacing the phantom. The diodes offered good response linearity and long-term reproducibility for both 6 and 18 MV. Axial dosimetry of coplanar beams (in a plane containing the beam axes) was made possible with the nearly isoplanatic response of the diodes over 360 deg. of gantry (usually within {+-}1%). For single beam geometry, errors in phantom placement as small as 0.5 mm could be accurately detected (in gradient {>=}1%/mm). In clinical setting, MLC systematic errors of 1 mm on a single MLC bank introduced in the IMRT plan were easily detectable with the QA phantom. The QA phantom demonstrated also sufficient sensitivity for the detection of setup errors as small as 1 mm for the IMRT delivery. These results demonstrated that the prototype can accurately and efficiently verify the entire IG-IMRT process. This tool, in conjunction with image guidance capabilities has the potential to streamline this QA process and improve the level of performance of image guided and intensity modulated radiotherapy.« less